Us er ’s a nd S erv ice G uid e Agi lent T echno log ies 85 032B/ E 50 Ω Type- N Calibr at ion Kit s Agilent P art Num ber : 8503 2-90 020 Pri nted in US A Print D a te: July 2 002 Supersed es: J une 2 000 © Copyright 1993, 2000, 2002 Agi lent T echnol ogies , I nc.
ii 85032B/ E Docu me ntat ion W arra nty THE MATERIAL CONTAINED IN THIS D OCUM ENT IS PROVIDED "AS IS ," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS .
Contents 85032B/E iii 1. Gen era l I n f or mat io n Calibr atio n Ki t O verv iew . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Ki t Cont ents . . . . . . . . . . . . . . . . . . . . .
iv 85032B/E Contents 4. P erformance V erifi cation Int rod uctio n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4- 2 H o w A g i l e n t V e r i f i es th e De vi c es in Th is K i t .
85032B/ E 1- 1 1 Gen e ral In forma tio n.
1- 2 850 32B/E General Inf ormation Calibrati on Kit Overview Calibr ati on Kit Overview The Agil ent 85032B a nd 85032E type- N calibr ation kits are used to cal ibrate Agi lent network analyzers up to 6 G Hz for measureme nts of c omponents wit h 50 Ω type-N connector s .
85032B/ E 1-3 General Information Calibrat ion Kit Overvie w Adap ters Like the other devic es in the kit, t he adapter s are b uilt t o very ti ght tole rances to p rovide good broadb and performanc e. T he adapter s utili ze a dual -beaded connector struc ture to ensure st able , r ep ea tabl e conn ections .
1- 4 850 32B/E General Inf ormation Incoming Inspection Inc o mi ng Ins p e c tio n V er ify that the shi pment is c omplete by ref erring to Figu re 6- 1 or Figu re 6-2 . Check for damage . The f oam-lined storag e case provi des prot ection during shi pping.
85032B/ E 1-5 General Information Recording t he Devi ce Serial Numbers Recor ding the Devic e Seri al Numbers In additi on to the kit ser ial number , the devic es i n this kit are ind ividuall y serial ized (serial numbers are l abeled onto the b ody of e ach dev ice).
1- 6 850 32B/E General Inf ormation Clarifyin g the Sex of a Connector Clarify ing the Sex of a C onn e ctor In thi s manual, the s ex of c a li bra tion de vices a nd ada pt ers ar e r ef erred to in t erms of thei r connector inter face . F or example , a male open ha s a male connector .
85032B/ E 2-1 2 Spec ifica tion s.
2-2 85032B/ E Specificati ons En vir onmental Requi rement s Environment al Requirement s T emperature —What to W atch Out F or Changes i n temperatur e can aff ect el ectric al charac teristi cs. Theref ore , the operating temperatur e is a critica l fact or in performance .
85032B/ E 2-3 Specifications Mechanical Characterist ics Mechani cal Ch ara cteri sti cs Mechanical charact eristics such a s center conduct or protrus ion and pi n depth a re not performance specific ations . They are , howev er , important s upplemental c harac teristi cs related to electrica l perfor mance .
2-4 85032B/ E Specificati ons Mechanical Character istics NO TE The gage s for measuri ng type-N c onnectors compensate f or the d esigned offs et of 5.2 6 mm (0.2 07 inc h), theref ore, pro trusion and recession r eadings are in relatio n to a zero re ference pl ane (as i f the i nner and o uter conductor planes w ere int ended to be f lush).
85032B/ E 2-5 Specifications Electrical Specifi cations Elec tr ical Spec ifica tio ns The electr ical specific ations i n T able 2- 3 apply to the devices in y our calib ration kit when connected with an Agil ent pr ecision inte rface .
2-6 85032B/ E Specificati ons Electrical Specifi cat ions.
85032B/ E 3-1 3 Us e, Main ten anc e, an d Ca re of the De vic es.
3-2 85032B/ E Use, Maintenance, and Care of the D evi ces Electr ostatic Disc harge Elect ros tati c Discha rg e Protect ion agai nst ESD ( electrosta tic di scharge) i s es sential whi le c onnecting , ins pecting , or cl eaning connectors attached to a static -sensitiv e circuit (such as t hose found in test sets) .
85032B/ E 3-3 Use, Maintenance, and Care of t he Devices Visual Inspection Visual Ins pect ion Visual inspe ction and, i f nec essary , cl eaning s hould be done every time a c onnection is made. Metal parti cl es from the c onnector t hreads may fal l into the connector when i t is disconnect ed.
3-4 85032B/ E Use, Maintenance, and Care of the D evi ces Cleani ng Connect or s Inspect F emale Connectors P ay specia l attent ion to the contact fingers in the female c enter cond uctor . These can be bent o r broken, a nd damage to t hem is not alw ays e asy to see .
85032B/ E 3-5 Use, Maintenance, and Care of t he Devices Cleani ng Connect ors Use a l int-fr ee swa b or cl eaning c loth mo istened wit h isopr opyl alc ohol to r emove any dirt or stubbor n conta minants on a connec tor t hat cannot b e removed w ith compress ed air or nitrogen.
3-6 85032B/ E Use, Maintenance, and Care of the D evi ces Gagi ng Connect ors Gagi ng Connec to rs The g ages avail able f rom Agilent T echnol ogies a re int ended fo r preventiv e maintenance and tr oubleshooti ng purposes only . (See T a bl e 6-3 o n pag e 6-5 fo r par t num ber inf ormation.
85032B/ E 3-7 Use, Maintenance, and Care of t he Devices Gagi ng Connect ors When to Gage C onnectors Gage a connec tor at the f ollowing tim es: • Prior to using a device for the f irst ti me: record the pin d epth measur ement so tha t it can be compared with future r eadings .
3-8 85032B/ E Use, Maintenance, and Care of the D evi ces Gagi ng Connect ors Gaging Procedures Gaging Male Type-N Connectors NO TE Alwa ys ho ld a connec tor gage by the gag e barrel, below the dial i ndicat or . This g ives the best stabilit y , and improves measurement ac curacy .
85032B/ E 3-9 Use, Maintenance, and Care of t he Devices Gagi ng Connect ors Fi gure 3-3 Gaging Male Type- N Connectors.
3-10 85032B/ E Use, Maintenance, and Care of the D evi ces Gagi ng Connect ors Gaging F emale Type-N C onnectors NO TE Alwa ys ho ld a connec tor gage by the gag e barrel, below the dial i ndicat or . This g ives the best stabilit y , and improves measurement ac curacy .
85032B/ E 3-11 Use, Maintenance, and Care of t he Devices Gagi ng Connect ors Fi gure 3-4 Gaging F emale Type-N Connectors.
3-12 85032B/ E Use, Maintenance, and Care of the D evi ces Connect ions Co nne c ti on s Good connecti ons requi re a ski lled op erator . The mo st com m o n caus e of me asu re me n t erro r is bad connect ions . The f ollowing procedure s illustr ate how to make good c onnecti ons .
85032B/ E 3-13 Use, Maintenance, and Care of t he Devices Connections Using a torque wrenc h guara ntees that the c onnection is not too tight, preventing possib le connect or damage . It also gua rantees that all connections a re eq ually ti ght each ti me.
3-14 85032B/ E Use, Maintenance, and Care of the D evi ces Connect ions 5. Ap pl y do wnwa rd force perpendic ul ar t o the wrench han dle . See Fi gu re 3- 6 .
85032B/ E 3-15 Use, Maintenance, and Care of t he Devices Handli ng and Storage How to Separate a Connection T o a void l ateral (bending) f orce on the connecto r mating plane s urf aces , alw ays supp ort the devices and connectio ns. CAUTI ON Turn the connect or nut, not the device bod y .
3-16 85032B/ E Use, Maintenance, and Care of the D evi ces Handli ng and Storage.
85032B/ E 4-1 4 P erform an ce V er ificat ion.
4-2 85032B/ E P erf ormance V erification Int roduct ion In troduct ion The perfor mance of your cali bra tio n kit c an on ly be verif ied b y retur ning the kit to Agilen t T echnolog ies for r ecert ificati on.
85032B/ E 4-3 P erf ormance V erification Recer ti fication Rec ert ification The foll owing will be prov ided with a recer tified kit: • a new calibrati on sti cker affixed to t he c ase • a cert.
4-4 85032B/ E P erf ormance V erification Recertification.
85032B/ E 5-1 5 T roub le shoot ing.
5-2 85032B/ E T r oubleshoot ing T ro ubl eshoot ing Pr ocess Tr oub l esh ooting Pr ocess If you suspec t a bad c alibration, or if your network analy zer does not pass per formance verif ication, follow the ste ps in Figure 5-1 .
85032B/ E 5-3 T roubleshooting Returning a Kit o r Device to Agilent Ret u rning a K it or Device t o Agilent If your ki t or device requi res service , conta ct the Agile nt T ec hnologies office nearest yo u for informati on on wher e to send i t. See T ab le 5-1 .
5-4 85032B/ E T r oubleshoot ing Returni ng a Kit or Devi ce to A gilent.
85032B/ E 6-1 6 R eplacea ble P arts.
6-2 85032B/ E Replaceab le P ar ts Int roduct ion In troduct ion T able 6-1 lists the r eplac ement part number s for items incl uded in the 8503 2B calibrat ion kit and Figu r e 6 -1 i llustrates each of t hese items .
85032B/ E 6-3 Replaceable Parts Intr oducti on Figur e 6-1 R epl acea b le P arts f or the 85032B C a libration Kit.
6-4 85032B/ E Replaceab le P ar ts Int roduct ion Figure 6-2 Repla cea ble P a rts for the 8 5032 E Calib ration Kit T able 6-2 Re plac eable P arts for the 8 503 2E Calib ration Kit Ite m No.
85032B/ E 6-5 Replaceable Parts Intr oducti on Ta b l e 6 - 3 R epl acea b le P arts— Item s Not I nclud ed in the C alib ration K it Item N o. Desc ri pt ion Qt y Agi l ent Part N um ber Connector Gages a (Type-N) a. T o ensure you choose th e corr ect ga ge, ref er to , “Clar ify in g th e Sex of a C onne cto r” on p ag e 1-6 .
6-6 85032B/ E Replaceab le P ar ts Int roduct ion.
85032B/ E A-1 A S tand ard D efini tio ns.
A-2 85032B/ E Standard De finitions Standar d Class Assignments Stand ard C lass A ssignm ents Class a ssignment orga nizes cal ibration s tandards i nto a for mat compatib le with t he err or models us ed i n the measu rement c ali brat ion.
85032B/ E A-3 Standard Definitions Standar d Class Assignments Blank F orm The standard c lass a ssignments l isted i n T able A-1 may be changed to m eet your specif ic requirements . T able A-2 i s provided to re cord the modified standard cl ass as signments .
A-4 85032B/ E Standard De finitions Nominal Standar d Definit ions Nomina l Stand ard De fi nit ions Standard definiti ons p rovide the c onstants needed to m athematicall y model t he elect rical charac teristi cs (del ay , atten uation, and impedance) of ea ch calib ration sta ndard.
85032B/ E A-5 Standard Definitions Nominal Standar d Definiti ons T able A -3 Standard Definitions Sy st em Z 0 a = 50 Ω a. Ensur e syste m impe d ance (Z 0 ) of networ k anal y z er is set to this value . Ca libr ati on K it L abe l: Stand ard b b.
A-6 85032B/ E Standard De finitions Nominal Standar d Definit ions Blank F orm The sta ndard defi nitions l isted in T a ble A-3 may be c hanged to meet your sp ecific requir ements . T able A-4 is pr ovided to r ecord t he modified s tandard definitions .
Index 85032B/ E Index-1 A adapt ers , 1- 3 part numbers , 6- 5 Agilent T e ch nologie s, cont act ing , 5-3 alcohol isopropyl as cl eanin g solvent , 3- 4 altitud e , 2-2 B blank fo rm standa rd cla s.
Index-2 85032B/ E Index connec ting , 3-14 disc onne ctin g , 3-14 freq uency spe cificat ions , 2-5 freq uency ra nge , 2- 5 , 4-3 G gage connec tor di al , 3- 7 handling , 3-8 , 3- 10 mas ter , 3-8 .
Index 85032B/ E Index-3 tempe rature , 2-2 supplem ent al char acteri st ics , 2-3 supplem ent al elec tri cal charac ter istic s , 2- 5 supplies cleaning , 1-3 sys te m im pe dan c e setti ng , A-4 T.
Een belangrijk punt na aankoop van elk apparaat Agilent Technologies 85032B/E 50 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Agilent Technologies 85032B/E 50 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Agilent Technologies 85032B/E 50 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Agilent Technologies 85032B/E 50 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Agilent Technologies 85032B/E 50 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Agilent Technologies 85032B/E 50 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Agilent Technologies 85032B/E 50 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Agilent Technologies 85032B/E 50 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.