Agilent PN 8510-16 Controlling T est Port Output Power Flatness Product Note UNCORRECTED POWER CORRECTED POWER OUTPUT POWER Agilent 8510C Network Analyzer.
2 Introduction Designers and manufacturers of active devices oft en need to control the pow er lev el at the t est port 1 of t heir pow er -sensitive devices, but f ind dif- ficulty in o vercoming inser tion losses. The insertion losses occur as a result of connecting components in the measurement pat h between the source and the DUT .
3 System Configuration The basic 8510C measurement system conf igura- tion needed for test por t f latness correction is sho wn in Figure 1. A sing le channel pow er meter such as the A gilent 437B, E4418A, or E4418B is required, along with a compatible sensor , for per- forming t he po wer f latness calibration.
4 Flatness-Correction Operation The basic setup procedure t o obtain f lat t est port output po wer is illustrated in Table 1. To simplify the execution of t he procedure, t he front panel hardk eys are enclosed in [brackets]. The sof tkeys are enclosed in {braces}.
5 T able 1. Flatness-correction calibration procedure 8510C Keystrokes Description Set up the power meter (see T able 2) For proper operation, the E4418B must be set up before initiating the flatness-correction routine. See T able 2 for specific instructions.
6 T able 2. Agilent E4418B power meter setup E4418B Keystrokes Description Preset and zero the power meter [PRESET/LOCAL] Return the power meter to a known state. {CONFIRM} [ZERO/CAL] {ZERO} Set up the 437B emulation on the power meter [SYSTEM/INPUTS] The 437B command set must be activated in order to perform the power flatness calibration.
7 The follo wing er ror messages may occur while attempting t o enable the f latness cor rection and/or reducing the test por t pow er level: • If f latness cor rection is enabled before reducing the test por t pow er level, IF Overload or Source 1 W arning—RF Unleveled ma y be display ed on the analyzer as the source becomes unlev eled.
8 V erifying a flatness-correction calibration T o verify t he f latness-correction calibration, t he po wer sensor should be reconnected t o the t est port t o measure the t est por t po wer at individual frequencies. Since the measurement system is calibrated in 50W , inaccuracies will occur when a device is not well mat ched.
9 Using test port 1 calibrations on test port 2 In most cases, port 1 will be t he input por t of t he DUT . When por t 2 must be used as the input por t to t he device, t he user may choose to use a por t 1 f latness-cor rection calibration on por t 2 since the port 1 and por t 2 signal pat hs are symmetrical.
10 Practical Application Examples Absolute output power measurements with flatness correction One benefit of t he f latness-correction capability is the ability t o measure the absolute output pow er of active devices.
11 Amplifier measurement example Step-by-step instr uctions for setting up and apply- ing f latness cor rections for t he measurement of an amplifier are sho wn in Table 7 . Gain, absolute output po wer , and gain compression measurements procedures are cov ered.
Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. W e strive to ensure that you get the test and measurement capabilities you paid for and obt ain the suppor t you need.
Een belangrijk punt na aankoop van elk apparaat Agilent Technologies PN 8510-16 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Agilent Technologies PN 8510-16 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Agilent Technologies PN 8510-16 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Agilent Technologies PN 8510-16 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Agilent Technologies PN 8510-16 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Agilent Technologies PN 8510-16 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Agilent Technologies PN 8510-16 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Agilent Technologies PN 8510-16 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.