Gebruiksaanwijzing /service van het product 1WMPD4002313 van de fabrikant A&D
Ga naar pagina of 24
Weighing Indicator for Check Weighing Scales 1WMPD4002313.
This Manual and Marks All safety messages are identified by the following, “W ARNI NG” or “CAUTION”, of ANSI Z535.4 (American National S tandar d Institute: Product Safety Signs and Labels). The meanings are as follows: WA R N I N G A potentially hazardous situatio n which, if not avoided, could result in death or serious injury .
1 Content s 1. COMPLI ANCE ................................................................................................. 2 2. INTRO DUCTION .............................................................................................. 3 3. UNP ACKING .
2 1. COMPLIANCE Compliance with FCC rules Please note that this equipment generat es, uses and can radiate radio frequency energy . This equipment has been tested and has been found to comply with the limits of a Class A computing de vice pursuant to Subpart J of Part 15 of FCC rules.
3 2. INTRODUCTION This manual describes how this product wor ks and how to get the most out of it in terms of performance. The FS - D weighing indicator is used to bu ild high performance check weighing scales. It has similar specifications and functions as an indicator of the FS - i series.
4 3. UNP ACKING Unpack the indicator carefully and keep the p acking material if you are likely to transport the indicator again in the future. When unpacking, check whether all of the following items.
5 4. INST ALLA TION and PRECAUTIONS 4-1. Precautions for Installing the Indicator Ground the indicator so that the user will not be subjected an electric shock. Do not handle the main power cord with wet hands. The AC plug is not water-resist ant.
6 Setscrew B Setscrew A 4. Solder the wires to the pin contact s of the connector . A 5 m or less load cell cable is recommended as the FS - D uses a 4 wire system (no remote sensing). 5. Reassemble the connector plug and tighten the setscrew A. 6.
7 4-4. Setting Up the Indicator Connect the main power cord to an outlet that has an earth ground terminal. Y ou may use the earth terminal on the rear si de of the display to ground the indicator . If necessary , adjust the viewing angle of the display by loosening the 2 clamps, changing the angle and re-tightening the clamps.
8 5. NAMES AND FUNCTIONS No. NAMES FUNCTIONS / OPERA TION WEIGHT DA T A DISPLA Y Indicates the weight data. WEIGHING UNIT INDICA T OR Indicates the weighing unit in use. ANALOG WEIGHT DISPLA Y Indicates zero to full scale or comparison limit s and results.
9 No. NAMES FUNCTIONS / OPERA TION COMP ARISON INDICA T ORS Indicates the results of the weight comparison LO (RED), OK (GREEN) and HI (YELLOW). ST ABLE ANNUNCIA TOR T urns ON when the weight reading is ST ABLE. ZERO ANNUNCIA TOR T urns ON when the weight is in the center of ZERO range.
10 6. FUNCTION P ARAMETER SETTINGS T o design a platform scale, se t the cap acity , weighing unit and other functions according to the specifications of the scale. Normally , end users do not have to set these function parameters and these setting can be sealed together with the calibration switch.
11 6-2. Function list Item Function number Description C01- 0 CAL switch enabled, 0 key enabled C01- 1 CAL switch disabled, 0 key disabled Calibration C01- 2 CAL switch enabled, 0 key disabled Do .
12 7. CALIBRA TION This function adjusts the scale (i ndicator) for accurate weighing. Calibration must be done when the FS - D is initially connected to a load cell (base unit). Calibration may also be required acco rding the changes in the environment.
13 If you need ZERO calibration but do not need SP AN calibration, turn the power off to exit from the calibration procedure. 7. T o calibrate with a different weight, change the displayed value using the 10-key pad. If you en ter the wrong number , press the C key .
14 8. AUT O-T ARE FUNCTION The FS - D has an auto-tare function to be used with the comp arator function enabled. If the weight value is within the OK r ange and stable for a preset time, then the indicator will automatically tare the weight and show zero.
15 9. ADDITIONAL INFORMA T ION This section describes the additional functions and corrections to the instruction manual of FS - i series (WM+PD4001332 and 1WMPD4001368) which are common to the FS - D. 9-1. F-Function list The following function settings are added (“ 1 1-2.
16 The frame ground and the shie ld are connected internally and there is no difference between them in use. Data form at Example of data Out of range “kg” (+) O L , +99999 9 9 9 _ k g C R L F Out of range “kg” (+) O L , +9999 . 9 9 9 _ k g C R L F The position of decimal point is diffe rent according to the model and/or “ f02 ”.
17 Command H I , + 0 0 1 0 0 C R L F A s su m e 2 d e cimal place number . Reply H I , + 0 0 1 0 0 C R L F 1.00% will be set as LO limit %. Command L O , + 0 0 1 0 0 C R L F A s su m e 2 d e cimal place number . Reply L O , + 0 0 1 0 0 C R L F 1.00% will be set as LO limit %.
18 10. ERROR MESSAGES Overload error W arning to indicate that an object beyond the weighing capacity has been placed on the weighing p an. Remov e the object from the weighing pan. Power-on error W arning to indicate that an object beyond the power-on zero range (50% of the weighing capacity or 10%) i s placed on the weighing platfo rm.
19 1 1. SPECIFICA TIONS 1 1-1. Capacity and minimum display C02 - setting 0 * 1 2 3 4 5 6 7 Capacity 6 15 30 60 150 300 600 1500 0.002 * 0.005 0.01 0.02 0.05 0.1 0.2 0.5 0.001 0.002 0.005 0.01 0.02 0.05 0.1 0.2 kg Min. display 0.0005 0.001 0.002 0.005 0.
20 1 1-3. Options OP - 02 (HC - 02 i ) SLA Battery (Y uasa Battery NP4-6 recommended.) OP - 03 (FS - 03 i ) RS - 232C + Comp arator Rela y Output (See note.) OP - 04 (FS - 04 i ) RS - 422/485 + Comp arator Re lay Output (See note.) Note OP - 03 and OP - 04 cannot coexist.
.
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95 131 U.
Een belangrijk punt na aankoop van elk apparaat A&D 1WMPD4002313 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen A&D 1WMPD4002313 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens A&D 1WMPD4002313 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding A&D 1WMPD4002313 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over A&D 1WMPD4002313 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van A&D 1WMPD4002313 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de A&D 1WMPD4002313 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met A&D 1WMPD4002313 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.