USER MANUAL H ELI O 2 HELIUM-OxygEn BLEndEr (DISS and NIST Connections) Model No. PM5400 Series (shown) PM5500 Series SA VE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA Fax: (+001) 610-262-6080 www .
1 H ELI O 2 BLENDEr HELIUM-OXYGEN RECEIVING / INSPECTION Remove the Precision Medical, Inc. H ELI O 2 (Helium-Oxygen) Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider . INTENDED USE The Precision Medical, Inc.
2 H ELI O 2 BLENDEr HELIUM-OXYGEN READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the H ELI O 2 Blender . This is provided for your safety and to prevent damage to the H ELI O 2 Blender . If you do not understand this manual, DO NOT USE the H ELI O 2 Blender and contact your Provider .
3 H ELI O 2 BLENDEr HELIUM-OXYGEN W ARNING • Only trained, qualied medical personnel under the direct supervision of a licensed physician should operate the H ELI O 2 Blender . • Use this H ELI O 2 Blender only for its Intended Use as described in this manual.
4 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS High Flow Low Flow Model PM5580: 80/20 heliox PM5570: 70/30 heliox PM5480: 80/20 heliox PM5470: 70/30 heliox Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i th b ot h s u p p l y p re s s u r e s a t 50 p s i ( 3.
5 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS continued Dimensions: (without ttings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.
6 H ELI O 2 BLENDEr HELIUM-OXYGEN DIAGRAMS CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. PM5400 and PM5500 Models A B F E D C H G.
7 H ELI O 2 BLENDEr HELIUM-OXYGEN COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A dia l used for sele cti ng oxyg en con cen tra tion s betw een 20% -100% or 30% -100%. The F I O 2 scale is used for reference only . The actual F I O 2 must be veried with an Alarmed Oxygen Monitor / Analyzer .
8 H ELI O 2 BLENDEr HELIUM-OXYGEN PRE-USE TESTING W ARNING • Read this User Manual before installing or operating the H ELI O 2 Blender . • Conrm the concentration of heliox with an alarmed Oxygen Monitor / Analyzer . • Conrm contents of heliox cylinder prior to use.
9 H ELI O 2 BLENDEr HELIUM-OXYGEN REVERSE GAS FLOW PROCEDURE ( CE Requirements ONL Y) 1. Assure bleed ow valve is not engaged. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the H ELI O 2 Blender to ensure that there is no outlet ow .
10 H ELI O 2 BLENDEr HELIUM-OXYGEN 7. Conrm the concentration of heliox with an Oxygen Monitor / Analyzer . If necessary activate the bleed ow valve to maintain F I O 2 accuracy . 8. T o activate the bleed, turn and rotate the knurled collar back until it contacts the cover .
1 1 H ELI O 2 BLENDEr HELIUM-OXYGEN OXYGEN FLOWMETER CONVERSIONS Oxygen Flowmeter Setting Corrected Heliox Flow (LPM) at V arious F I O 2 Settings 20% 30% 40% 50% 60% 70% 80% 90% 100% 1 1.8 1.6 1.4 1.3 1.2 1.18 1.15 1.02 1.0 2 3.6 3.2 2.8 2.6 2.4 2.4 2.
12 H ELI O 2 BLENDEr HELIUM-OXYGEN TECHNICAL DESCRIPTION For a complete T echnical Description of the H ELI O 2 Blender and list of Replacement Parts, reference the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .precisionmedical.
13 H ELI O 2 BLENDEr HELIUM-OXYGEN TROUBLESHOOTING If the H ELI O 2 Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide, refer to the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .
14 H ELI O 2 BLENDEr HELIUM-OXYGEN LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precis ion Medical, Inc. warran ts that the HeliO 2 Blende r , (the Product), will be free of defects in workmanship .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo Europe (European Ofce) Molenstraat 15 2513 BH, The Hague The Netherlands Phone: +31 (0) 70.
Een belangrijk punt na aankoop van elk apparaat Helio PM5400 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Helio PM5400 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Helio PM5400 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Helio PM5400 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Helio PM5400 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Helio PM5400 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Helio PM5400 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Helio PM5400 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.