Gebruiksaanwijzing /service van het product 4909 CPVC van de fabrikant Honeywell
Ga naar pagina of 34
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99.
ii 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual iii A bout This Document Abstract The purpose of this document is to support the installation operation and m aintenance of the 4909 CPVC Conductivity Cell Remov al/Insertion Assem bly.
iv 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Organization Telephone Address Honeywell TAC 1-800-243-9883 Voice 1100 Virginia Drive Fort Washington, PA 19034.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual v Contents 1. INTRODUCTION ....................................................................................................1 1.1 Overview ............................
vi 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4. MA INTENA NCE A ND REPLAC EMENT PA RTS .................................................17 4.1 To Clean The Cell ...................................................
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual vii Tables Table 4-1 Replacem ent Parts __________________________________________________________ 18 Table 4-2 Voltage.
.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 1 1. Introduction 1.1 Overview The 4909 Conduct ivit y C ell Ins ertion/R emoval A ssembly is design ed fo.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 2 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 1.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 3 1/ 4" NP T Plug Cabl e 1 - 1/ 2" NPT Nipple Conducti vity Cell P urg e Po rt C ompr es s ion H.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 4 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Cable C onductivity Cell Ad ap t er * * Au t oma t ic Tempera t ur e Com pensato r , w hen specifie d by cat a log nu mber, is m ounted insid e ad apter.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 5 A D C B Ter mina l Boar d As s e m b l y #6- 32 Bi ndi ng He ad Sc r e w & Wa s h e r Vi e w of B a .
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 6 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99.
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7 2. Specifications and Model Selection Guide 2.1 Specifications Cell Constants 0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm -1 . Electrode Material Nickel or Platinum , as specified.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 8 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Integral Automatic Temperature Compensation Refer to Section 2.2 – Model Selection Guid e Leadwire Tefzel covered, 18-g age cable (7 feet, 20 f eet, or Universal Head, as specified).
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 9 049 TABLE IV Selection 09 08 Leadwire Length ’ 7 ft.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 10 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 049 TABLE VII - OPTIONS Tagging None 0_ Linen L_ Stainless Steel S_ Certificate of Calibration No _ 0 Yes _ 1 Notes: 1.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 11 3. Installation 3.1 Requirements To insur e that a re presenta tive sample is being measured at all times, the solutio n must m o ve thr ough and completely purge out the cell channels or guard tube.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 12 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Allow at least 1/2 inch clearance bey ond the end of th e cell and 1/8 to 3/16 inch radius clearance surrounding the cell to p e rmit circulation of the solutio n.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 13 3. Using a strap wrench on the m ounting nipple, tighten the assembly an additional 1-1/2 to 2 turns. Do not use stillson or chain type wrenches as they m ay damage and weaken the CPVC plastic.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 14 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 3.5 Removal 1. REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Sh ut off purge line, if used. Dis connect wi ring conn ections if X1 is selected for Table IV in the MSG.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 15 2.5" (64mm) App r ox 4.1" (105mm) App r ox 12.5" (317mm) X " See Bel ow App r ox 19.7" max . (502mm) Allow a minim um clearance o f 36" (914 m m) beyond this poin t for removal of electrode.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 16 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 2. 5" (64mm) Ap prox 4 .1" (105mm) Approx 12. 5" (317mm) X " See Below A pprox 19.
Maintenance and Replacement Parts 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 17 4. Maintenance and Replacement Parts If a series of below normal condu ctivity readings or abov e normal resistiv ity readings occu r, this m ay indicate that the cell is not filled with process solution.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 18 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4.3 Replacement Parts Table 4-1 Replacement Parts Description Part Number Complete Ass embly See Figs.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 19 5. Platinization and Platinum Black Only the electrodes havi ng cons tants from 5 t.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 A D C B View o f J unct ion Box Head w ith Cap Re moved WH I T E BLA CK RED GREEN Cell Temp. Com p.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 21 A A D C B View o f J unct i on Box He ad with Cap Removed WH I T E BLA CK RED GREEN Cell Temp.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 22 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 WH I T E BLACK RED GREE N Cell Temp. Comp. WK S H G R Five Point Termin al Board. Each Terminal W i ll Accept #16 G age Max .
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 23 WH I T E BLA CK RED GREEN Cell Temp.
.
.
Industrial Auto mation and Co ntrol Honey well, Inc. 1100 Virginia D rive Fort Wa shington, Pennsy lvania 19034.
Een belangrijk punt na aankoop van elk apparaat Honeywell 4909 CPVC (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Honeywell 4909 CPVC heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Honeywell 4909 CPVC vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Honeywell 4909 CPVC leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Honeywell 4909 CPVC krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Honeywell 4909 CPVC bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Honeywell 4909 CPVC kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Honeywell 4909 CPVC . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.