NI 5412 Specifications NI PXI/PCI-5412 14-Bit 100 MS/s Arbitrary Waveform Generator Unless otherwise noted, the followi ng conditions were used for each specification: • Interpola tion set to maximum allo wed factor f or a gi v en sampl e rate. • Signals terminated with 50 Ω .
NI 5412 Specifications 2 ni.com CH 0 (Channel 0 Analog Output , Front Panel Connector) Table 1. Specif ication Va l u e Comments Number of Channels 1 — Connector SMB (jack) — Output Voltage Characteristics Output Paths The software-selectable Main Ou tput Path setting provides full-scale vo ltages from 12.
© National Instruments Corpor ation 3 NI 5412 Specifications Maximum Output Voltage Maximum Output Vo l t a g e Pat h Load Maximum Output V oltage (V pk -pk ) The Maxi mum Output V oltage of the NI 5412 is determined by the Amplitude Range and the Of fset Range.
NI 5412 Specifications 4 ni.com Frequency and Transient Response Bandwidth 20 MHz –3 dB Digital Interpolation Filter Software-selectable Finite Im pulse Response (FIR) filter . A vailable interpolation f actors are 2, 4, or 8. — Passband Flatness Low-Gain and High-Gain Amplifiers Path — ±1.
© National Instruments Corpor ation 5 NI 5412 Specifications Figure 1. Pulse Response, Low-Gai n Amplifier Path 50 Ω Load –2.0 –1.6 –0.4 –1.
NI 5412 Specifications 6 ni.com Table 1. (Continued) Specification Va l u e Comments Suggested Maximum Frequencies for Common F unctions Function Path — Lo w-Gain Amplif ier High-Gain Amplifier Sine.
© National Instruments Corpor ation 7 NI 5412 Specifications Sample Clock Spectral Characteristics (Continued) A verage Noise Density Pat h Amplitude Range A verage Noise Density A verage Noise Density at small amplitudes is limited by a –148 dBm/Hz noise floor .
NI 5412 Specifications 8 ni.com Sample Rate Rang e and Resolution Sample Clock Source Sample Rate Range Sample Rate Resolution — Divide-b y- N 23.8 4 S/s to 100 MS/s Settable to ( 100 MS/s) / N (1 ≤ N ≤ 4,194,304) High- Resolution 10 S/s to 100 MS/s 1.
© National Instruments Corpor ation 9 NI 5412 Specifications System Phase Noise and Jitter (10 MHz Carrier) Sample Clock Source System Phase Noise Density (dBc/Hz) Offset System Output Jitter (Integrated from 100 Hz to 100 kHz) 1. High- Resolution specifications va r y wi t h Sample Rate.
NI 5412 Specification s 10 ni.com Onboard Clock (Internal VCXO) Phase-Locked Loop (PLL) Reference Clock Table 3. Specif ication Va l u e Comments Clock Source Internal sample clocks can either be locked to a Reference Clock using a phase-locked loop or be deriv ed from th e onboard VCXO frequency reference.
© National Instruments Corpor ation 11 NI 5412 Specifications CLK IN (Sample Clock and Reference Cloc k Input, Front Panel Connector) Table 5. Specif ication Va l u e Comments Connector SMB (jack) — Direction Input — Destinations 1. Sample Clock 2.
NI 5412 Specification s 12 ni.com PFI 0 and PFI 1 (Programmable Function Interfac e, Front Panel Connectors) Table 6. Specif ication Va l u e Comments Connectors T wo SMB (jack) — Direction Bi-directional — Frequenc y Range DC to 105 MHz — As an Input (Trigger) Destinations Start T rigger — Maximum Input Overload –2 V to +7 V — V IH 2.
© National Instruments Corpor ation 13 NI 5412 Specifications Start T rigger V OH Minimum: 2.9 V (open load), 1.4 V (50 Ω load) Output dri vers are +3.3 V TTL compatible. Measured with a 1 m cable. V OL Maximum: 0.2 V (open load), 0.2 V (50 Ω load) Rise/F all T ime (20% to 80%) ≤ 2.
NI 5412 Specification s 14 ni.com Delay from Start T rigger to CH 0 Analog Output Interpolation Factor T ypical Delay Refer to t s2 at NI Signal Generators Help»Devices» NI 5412» NI <bus>-5412» Triggering» Trigger Timing . Digital Inter polation F ilter disabled.
© National Instruments Corpor ation 15 NI 5412 Specifications Markers Table 8. Specif ication Va l u e Comments Destinations 1. PFI<0..1> (SMB front panel connectors) 2. NI PXI-5412 —PXI_T rig<0..6> (backplane connector) NI PCI-5412 —R TSI<0.
NI 5412 Specification s 16 ni.com W aveform and Instruction Memor y Utilization Table 9. Specif ication Va l u e Comments Memory Usage The NI 5412 uses the Synchronization and Memory Core (SMC) technology in which wav eforms and instructions share onboard memory .
© National Instruments Corpor ation 17 NI 5412 Specifications Loop Count 1 to 16,777,215. Burst trigger: Unlimited — Quantum W aveform size must be an integer multiple of four samp les. — Memory Limits 8 MB Standard 32 MB Option 256 MB Option All trigger modes except where noted.
NI 5412 Specification s 18 ni.com Calibration Power Table 10. Specif ication Va l u e Comments Self-Calibrati on An onboard, 24-bit ADC and precision voltage reference are used to calibrate the DC gain and offset. The self-calibration is initiated by the user through the software and takes approximately 75 seconds to complete.
© National Instruments Corpor ation 19 NI 5412 Specifications Software Table 12. Specif ication Va l u e Comments Dri ver Software NI-FGEN 2.3 or later ve rsion. NI-FGEN is an IVI-compliant driv er that all ows you to conf igure, control , and calibrate the NI 5412.
NI 5412 Specification s 20 ni.com Environment NI PXI-5412 Environment Note T o ensure that the NI PXI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced-Air Cooling Note to Users incl uded in the NI 5412 kit. The NI PXI-5412 is intended for indoor use only .
© National Instruments Corpor ation 21 NI 5412 Specifications NI PCI-5412 Environment Note T o ensure that the NI PCI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced -Air Cooling Note to Users inclu ded in the NI 5412 kit. Also, to maxim ize airflo w and extend the life of the device, lea ve any adjacent PCI slots empty .
NI 5412 Specification s 22 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Table 15. Specification Va l u e Comments Safety The NI 5412 meets the requirements of the following standards for safety and electrical equipment for measurement, control, and laboratory use: • IEC 61010 -1, EN 61010-1 • UL 61010-1 • CAN/CSA C22.
© National Instruments Corpor ation 23 NI 5412 Specifications EMC/EMI CE, C-T ick, and FCC Part 15 (Class A) Compliant Notes : 1. This device is not intended for , and is restricted from, use in residential areas. 2. For EMC compliance, op erate this device with shielded cabling.
NI 5412 Specification s 24 ni.com Physical Table 16. Specif ication Va l u e Comments Dimensions NI PXI-5412 NI PCI-5412 — Single 3U PXI slot. CompactPCI compatib le. 2.0 × 13.0 × 21.6 cm (0.8 × 5.1 × 8.5 inches) 34.07 × 10.67 × 2.03 cm (13.4 × 4.
© National Instruments Corpor ation 25 NI 5412 Specifications Where to Go for Support The National Instrum ents W eb site is y our complete resource for technical support.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more information a bout National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s .
Een belangrijk punt na aankoop van elk apparaat National Instruments NI 5412 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen National Instruments NI 5412 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens National Instruments NI 5412 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding National Instruments NI 5412 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over National Instruments NI 5412 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van National Instruments NI 5412 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de National Instruments NI 5412 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met National Instruments NI 5412 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.