Gebruiksaanwijzing /service van het product NI PXI/PCI-5114 van de fabrikant National Instruments
Ga naar pagina of 20
NI PXI/PCI-5114 Specifications 8-Bit 250 MS/s Digitizer Unless otherwise noted, the followi ng conditions were used for each specification: • All filter settings • All impedance selections • Sample clock set to 250 MS/s T ypical values are representati ve of an av erage unit operating at room temperature.
NI PXI/PCI-5114 Specifications 2 ni.com Power .............. .................... ................ .................... .................... .............14 Software .................. .................... .................... .................... ...
© National Instruments Corporat ion 3 NI PXI/PCI-5 114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Of fset 50 Ω 1 M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offset Range (V) 0.
NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (T ypical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/F all T ime, T ypical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.
© National Instruments Corporat ion 5 NI PXI/PCI-5 114 Specifications Specif ication Va l u e Commen ts Spectral Characteristics Spurious Free Dynamic Range with Harmonics (SFDR), T ypical Range (V pk-pk ) 10 MHz, –1 dBFS input signal. Includes the 2 nd through the 5 th harmonics.
NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Va l u e Comments Sources Internal, Onboard Clock (i nternal VCXO)* External, CLK IN (front panel SMB connector) * Internal Sample Clock is locked to the Reference Clock or deri ved from the onboard VCXO.
© National Instruments Corporat ion 7 NI PXI/PCI-5 114 Specifications Phase-Locked Loo p (PLL) Reference Clock External Sample Clock Sources CLK IN (front panel SMB connector) — Frequenc y Range 50 MHz to 250 MHz Di vide by n decimation av ailable where 1 ≤ n ≤ 65,535.
NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) T r igger Reference (Stop) T rigger Specification Va l u e Comments Input V oltage Range Sine wav e: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wav e: 0.
© National Instruments Corporat ion 9 NI PXI/PCI-5 114 Specifications Analog Trigger (Edge, Window, and Hysteresis Trigger Types) Sources CH 0 (front panel BNC connector) CH 1 (front panel BNC connec.
NI PXI/PCI-5114 Specifications 10 ni.com TRIG (External T rigger , Front Panel Connector) PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) T ypes Specific Line Any Line Specific Field — Standards SDTV : M-NTSC, B/G-P AL, SECAM, M-P AL EDTV : 480i/59.
© National Instruments Corporat ion 11 NI PXI/PCI-5 114 Specifications As an Input (Trigger) Destinations Start T rigger (Acquisition Arm) Reference (Stop) T rigger Arm Referen ce Advance T rigger — Input Impedance 150 k Ω — V IH 2.0 V — V IL 0.
NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any number o f SMC-based modules in a chassis.
© National Instruments Corporat ion 13 NI PXI/PCI-5 114 Specifications W aveform Specifications Specif ication Va l u e Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channe.
NI PXI/PCI-5114 Specifications 14 ni.com Calibration Power Specification Va l u e Comments Self-Calibrati on Self-calibration i s done on soft ware command. The calibration corrects for gain, of fset, compensated 1 M Ω attenuat or , triggering, and timing adjustment errors fo r all input ranges.
© National Instruments Corporat ion 15 NI PXI/PCI-5 114 Specifications Software Specification Va l u e Comm ents Dri ver Software NI PXI-5114 : NI-SCOPE 2.9 or later . NI PCI-5114 : NI-SCOPE 3.1 or later . NI-SCOPE is an IVI-complian t dri ver that allo ws you to configure, control, and calib rate the NI 5114.
NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note T o ensure that the NI PXI-5114 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only .
© National Instruments Corporat ion 17 NI PXI/PCI-5 114 Specifications NI PCI-5114 Note T o ensure that the NI PCI-5114 cools ef fectiv ely , follo w the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit. Also , to maximize airflo w and extend the life of the de vice, leave an y adjacent PCI slots empty .
NI PXI/PCI-5114 Specifications 18 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Specification Va l u e Comm ents Safety The NI 5114 meets the requirements of the following standards of safety for electri cal equipment for measurement, control, and laboratory use: • IEC 61010 -1, EN 6101 0-1 • UL 61010-1, CAN/CSA-C22.
© National Instruments Corporat ion 19 NI PXI/PCI-5 114 Specifications Physical Specification Va l u e Comments Dimensions NI PXI-5114 NI PCI-5114 — 3U, One slot, PXI/cPCI Module 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) 35.5 × 2.0 × 11.3 cm (13.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and company names mentioned herein are trademarks or trad e names of their respective companies.
Een belangrijk punt na aankoop van elk apparaat National Instruments NI PXI/PCI-5114 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen National Instruments NI PXI/PCI-5114 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens National Instruments NI PXI/PCI-5114 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding National Instruments NI PXI/PCI-5114 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over National Instruments NI PXI/PCI-5114 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van National Instruments NI PXI/PCI-5114 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de National Instruments NI PXI/PCI-5114 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met National Instruments NI PXI/PCI-5114 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.