ManMal Operator% I - 03 a Tipping Bucket Rain Gauge 8 &E RG-2500 Series.
MSSO395FVRlB 3591RUSH I Call for Your FREE Handbook Request Form Today: (203) r/ Environmental Monitoring and Control 1/ Electric Heaters V Data Acquisition Systems pH and Conductivit y v P Flow and L.
Y i.
Unpacking Instructions Remove the Packing List and verify that you have received all equipment, including the following (quantities in parentheses): RG-2500 Tipping Bucket Rain Gauge (1) Operator ’ s Manual (1) If you have any questions about the shipment, please call the OMEGA Customer Service Department.
Y.
. 6 . . . . . . . . , . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . 6 Chapter 6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ......................................... 4 Chapter 5 Maintenance .
lbs/18 lbs Optional Accessories: Wind Screen, Alter Type Event Recorder, Drum Chart Calibrator Bottle Digital Event Counter Model RG-952 Model RG-6113-A Model RG-2595 Model RG-2596 1 lbs/12 lbs RG-2500-12: 14 8” x 17 ” RG-2500-12: 12 ” x 20 ” Weight/Shipping: RG-2500: 7 + 3% for 0 to 6 inches per hour Switch: Magnetic reed, SPST N.
3/8 inches. The holes in the feet should be matched to anchor bolts located in the foundation or on the platforms. for severely tilted foundations, use a set of three hex nuts on each bolt to provide leveling and locking of the rain gauge onto the foundation.
portions.of the gauge causing some error in the counts. 3 As rain is collected by the eight-inch diameter orifice, the rain drops are directed down through the funnel and into the bucket assembly. As soon as the bucket has collected enough rain to represent 0.
t 8.24 = 114.8. fof a quart of water should be 946.3 45% since the control of the methods is not as good as when the testing is done in the laboratory. The total number of tips should equal the amount of water poured into the gauge, divided by the calibrated number of tips per the number of milliliters of water shown in the table.
4.4 Adjustments Should the calibration of the rain gauge appear to be incorrect and in need of adjustment, change the heights of the two calibration posts adjustment screws. To change the screw heights, loosen the nut on the top of the post that locks the screw into place.
1 normally connected to the rain gauge. Manually move the bucket assembly and look for he switch closure. Upon completion of testing and cleaning of the ga ge, replace the outer cover and funnel and recheck the level of the gauge using a carpenter ’ s level across the top of the gauge.
/=l”“ i’ SbOO346 11 - 25 - w A x Y RGN 1 1 OF I SHEE T UYCE OOEL ?HEMATIC, SWITCH, MAGNETIC, TIPPING BUCKET RAIN GAUGE t 1. TYPICAL OF BOTH HEATED AND UNHEATED RAIN GAUGES . @ 175 VD C MAXIMU M TYPICAL CLOSURE TIME 0.1 SECON D NOTES: 0.25A -0 BLK MAGNET N.
, Nc+JE DATE 11-25-97 DW $ . o ; 0347 BY RGN SCALE 1 1 DF us*GE SHEET UODE‘ ?XHEMATIC, HEATER POWER, TIPPING BUCKET RAIN GAUGE & SOCKE T C +60’ F 2-PRON G P ‘ LUG +40-F OPENS AT T!iE RAI N GAUGE.
NONE 93011 6 l-07-93 0.7t: owe. No. SCKE RGN BY cs 1 IslifEr 1 _. 2500 USAm_ _ L(oo. @ 120 DEG MOUNTING FEET PATTERN 8 INCH RAIN GAGE PLCS DIA 3 WTH LEVELER 0.
. . < 1 - 93011 8 NONE l-07-93 OAIE om.No. ME RGN By w 1 lsafm I 250+12 us*cE uoo. @ 120 DE G MOUNTING FEET PATTERN I 12 INCH RAIN GAGE PLCS .- 3 -.
a I= a.
1, NONE OAIE t-o & LEVELER 12 INCH RAIN GAGE BASE ASSEMBL Y I MTH FOOJI~JE & ADJUSTMENT SCREW TIPPING BUCKET SIGNAL OUTPUT CONTACT CLOSURE COLLECTION TUBE f- 2 PIN SIGNAL TERMINAL BLOCK CALIB.
L t USAef REVISED AUG 1994 UOD. - 1. 4 PIN TERMINAL BLOCK IS POSITIONED BEHIND HEATER . 2. OUTER COVER NOT SHOWN. ’ 12 INCH HEATED RAIN GAGE nn E WIRING DIAGRAM NOTES W lT H LE M LER RAIN GAGE BASE .
1 NONE DATE l-07-93 l”“I”” 93022 2 RGN By I OF METI 2500 USAcf Yoo. nRE RAIN GAGE INSTALLATION KIT ASSEMBLY , IC ANCMT FOR INSTALLATIONS USING A 24” ALTER STYLE WIND SCREEN THIS HEIGHT MUST BE ADJUSTED SO THAT THE TOP EDGE OF THE RAIN GAGE IS WITHIN 1 INCH OF THE TOP EDGE OF THE WIND SCREEN LEAVES.
, NONE DATE l-07-93 930223 “O. RCN [“ ‘ I” BY 1 (x l S”EE ll -2500 UOD. USAGE "RE RAIN GAGE INSTALLATION KIT ASSEMBLY ’ C 3/8 WASHER. ,NUTS FOR ALL THREE OF THE RAIN GAGE MOUNTING FEET UNTIL THE RAIN GAGE IS LEVEL ACCORDING TO THE SPIRIT BUBBLE LEVEL LOCATED ON THE RAIN GAGE BASE.
I ’ , I I I / -- I .
j=lIo*ci8000323 10-6-97 DAK RCN 61 I DRAWING NOT TO SCALE Df t ISMCEI WJ-ISCCU USACE WDEL -+d!s POWER COR D W ARN I NG LABEL F??E 5A. 3A G STRAIN RELIEF CLAMP & HEX NUT & FUSE HOLDER BRACKET .
7-S/8, LARGE FUNNEL ............. . 1 15 FUNNELL ............ . 1 SCREEN, ALLEN1/8 .................. . 1 SCREEN, 2" DIA, SMALL 8-32X2SS.. ................. . 2 STUD, 10-32X1-1/2 S S ................ . 3 WRENCH, 6-32X2SS.. ................. . 1 STUD, .
Y 16.
‘ I995 OMEGA ENGINEERING, INC. All rights reserved. This documentation may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
M2234/0395 & Dissolved Oxygen Instruments pH, Conductivity m & Wastewater Treatment @’ Industrial Water & Water Monitors @ Air, Soil & Tubing @ pumps m Refractometers D Metering &.
Een belangrijk punt na aankoop van elk apparaat Omega Engineering RG-2500 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Omega Engineering RG-2500 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Omega Engineering RG-2500 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Omega Engineering RG-2500 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Omega Engineering RG-2500 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Omega Engineering RG-2500 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Omega Engineering RG-2500 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Omega Engineering RG-2500 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.