USER MANUAL AIR-O XYGEN BLENDER (DIS S and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CAUTION F ederal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA F ax: (+001) 610-262-6080 www .
1 Blender Air-Oxygen CONTENTS RECEIVING / INSPECTION ......................................................... 2 INTENDED USE .......................................................................... 2 READ ALL INSTRUCTIONS BEFORE USING .............
2 Blender Air-Oxygen RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air -Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider .
3 Blender Air-Oxygen SAFETY INFORMA TION - W ARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury . WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury .
4 Blender Air-Oxygen WARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender , and 3 lpm for the Low Flow Blender .
5 Blender Air-Oxygen SPECIFICA TIONS Model PM5200 High Flow PM5300 Low Flow Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) w i t h N O B L E E D Auxiliary Outlet Flow Range 2 - 100 lpm 0 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 .
6 Blender Air-Oxygen SPECIFICA TIONS continued Dimensions: (without fittings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.4 cm) W eight: 2.
7 Blender Air-Oxygen DIAGRAMS CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Depending on model, your fittings may differ from these diagrams.
8 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A di al use d fo r sel ect in g ox yg en con ce ntr ati on s be tw een 21%-100%. The FiO 2 scale is used for reference only . This Dial does not rotate 360°. The dial starts at 21% and ends at 100%.
9 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION F Air Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose . G Alarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply .
10 Blender Air-Oxygen ALARM TEST 1. The Air -Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air -Oxygen Blender . 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly .
11 Blender Air-Oxygen 3. T o activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover . 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor . 5. Confirm the flow of air and/or oxygen mixture to the patient.
12 Blender Air-Oxygen 1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water .
13 Blender Air-Oxygen TROUBLESHOOTING If the Air -Oxygen Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide refer to the Air -Oxygen Blender Service Manual (P/N 504827) available on the Internet; www .
14 Blender Air-Oxygen LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precision Medical, Inc. warrants that the Blender , (the Product), will be free of defects in workmanship and/or material for the .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo E.
Een belangrijk punt na aankoop van elk apparaat Precision Trading PM5200 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Precision Trading PM5200 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Precision Trading PM5200 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Precision Trading PM5200 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Precision Trading PM5200 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Precision Trading PM5200 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Precision Trading PM5200 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Precision Trading PM5200 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.