MBR-1404G MBR-1404GR Blood Bank Refrigerator INSTRUCTION MANUAL MBR-1404GR.
1 Note: 1. No part of this manual may be reprodu ced in any form without the expressed w ritten permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies.
2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user compli es with this manual as it contains important safety advice. Items and procedures are described so that y ou can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or serv ice personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire.
4 PRECAUTIONS FOR SAFE OPERATION If this unit is to be used for storing poisons, radioactive material or other harmful products, ensure that it is in a safe area. Failure to do so may lead to an adverse effect on the health of personnel in the area and the local environment.
5 PRECAUTIONS FOR SAFE OPERATION Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. Improper installation may result in water sp illage or injury from the unit tipping over. Do not install the unit at a place w ith the possibility of something falling on it.
6 PRECAUTIONS FOR SAFE OPERATION Dispose of any water in the evaporation tray completely before moving the unit. Spilled or splashed water may cause current leakage or electric shock. Be careful not to tip the unit over during movement to prevent damage or injury.
7 CAUTIONS FOR USAGE 1. Adequate space should be provided between the items inside the unit to allow air circulation. 2. Never store corrosive materials such as acid or alkali unless the container is completely sealed up. Corrosion may lead to failure of the unit in time.
8 ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1): 1. Indoor use; 2. Altitude up to 2000 m; 3. Ambient temperature 5 o C to 40 o C 4. Maximum relative humidity 80% for temperature up to 31 o C decreasing linearly to 50% relative humidity at 40 o C; 5.
9 REFRIGERATOR COMPONENTS MBR-1404GR 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17.
10 REFRIGERATOR COMPONENTS 1. Front cover: Open this cover to connect to a remote al arm or when cleaning the condenser filter. Refer to page 19. 2. Control panel: Panel opens when the right top of the lowe r part of the cover is pushed. Refer to page 12.
11 REFRIGERATOR COMPONENTS Inside of front cover 1. Condenser filter: This filter is to prevent dust from plugging the condenser. Please clean the filter once a month. 2. Glow starter: This is for fluorescent light. Use “FG-4P” for replacement. 3.
12 REFRIGERATOR COMPONENTS Control panel 1. Alarm buzzer stop key (BUZZER): Press this key to temporarily s ilence the buzzer in the event that the alarm operates and buzzer sounds.
13 INSTALLATION Installation site T o operate this unit properly and to obtain maximum performance, inst all the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to dire ct sunlight may lead to inadequate cooling.
14 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, cl ean them with a neutral detergent and wipe it up with a wet cloth.
15 BEFORE COMMENCING OPERATION Preparation of monitor bottle Prior to the operation of the unit, fill the monitor bo ttles (upper and lower) with 10% glycerol (or other solution such as ethylene glycol) by the following procedure: 1. Remove the top and bottom drawer (or shelf).
16 START-UP OF UNIT Follow the procedures for the initia l and consequent operations of the unit. 1. Connect unit to dedicated power supply. Do not put any product in the unit at this time. 2. When the power is initially applied, the audible alarm may sound.
17 OPERATING INSTRUCTIONS Temperature display The unit is set at the factory to obtai n the proper temperature automatically (4 o C±1.5 o C). The high temperature alarm is preset at 6 o C.
18 OPERATING INSTRUCTIONS Alarm, safety function, and self diagnostic function This unit has the alarm and safety functions shown in Table 2, and also self diagnostic function. Table 2 Alarm and safety functions Kind of alarm or safety Situation Indication Buzzer Safety operation High temperature alarm If the upper or lower temp.
19 ROUTINE MAINTENANCE WARNING Alw ay s disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electr ic shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
20 ROUTINE MAINTENANCE Cleaning of evaporating tray 1. Remove the evaporating tray installed on the back of the frame by lifting the tray. 2. Wash away any accumulated dirt in the evaporating tray. 3. After cleaning of the evaporating tray, return it to the original position.
21 ROUTINE MAINTENANCE Calibration of temperature sensor The temperature sensors are attached to the monitor bottles. Follow the following instructions for proper calibration (zero adjustment): 1. Remove the monitor bottle cover as shown in the figure.
22 TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when sw itched on ■ There is a power failure. ■ The fuse is blown or the ci rcuit breaker is inactivated. ■ The unit is not connected to the power supply.
23 SPECIFICATIONS Name Blood Bank Refrigerator Model MBR-1404G MBR-1404GR External dimensions W1440 x D830 x H1950 (mm) Internal dimensions W1320 x D697 x H1500 (mm) Effective capacity 1287 L Exterior.
24 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: (List all potentially hazardous materials that have been stored in this unit.
7FB6P101236004 (5 April 2005) Recycl ed paper SANYO Electric Biomedical Co., Ltd. 10-15, Hongo 3-Chome Bunkyoku, T o kyo 1 13-8434 Jap an Printed in Japan.
Een belangrijk punt na aankoop van elk apparaat Sanyo MBR-1404G (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Sanyo MBR-1404G heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Sanyo MBR-1404G vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Sanyo MBR-1404G leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Sanyo MBR-1404G krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Sanyo MBR-1404G bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Sanyo MBR-1404G kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Sanyo MBR-1404G . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.