Gebruiksaanwijzing /service van het product MGD 521 M van de fabrikant Barco
Ga naar pagina of 36
Inst allation & User Manual MGD 521 M MGD 521 MKII Medical Greyscale Displays V599326/Rev . 02 June 2003 BarcoView Theodoor Sevenslaan 106 8500 Kortrijk, Belgium Phone: +32(0)56 23 32 44 Fax: +32(0)56 23 33 74 E-mail: sales.medical@barco.com http://www .
MGD 521 M - MKII User's Guide 3 © 2003 BARCO nv. All rights reserved. Trademark information TrueGrey, MediCal, BarcoMed, and MeDis are registered trademarks of BARCO NV.
MGD 521 M - MKII User's Guide 4.
MGD 521 M - MKII User's Guide 5 Regulations This apparatus conforms to: IEC601-1, UL2601-1, cUL2601-1, EN 60601-1 This apparatus is classified as Type B without Applied Part.
MGD 521 M - MKII User's Guide 6.
MGD 521 M - MKII User's Guide 7 CONTENTS 1. Introduction ..................................................................................................... 9 1.1 Overview ........................................................................
MGD 521 M - MKII User's Guide 8.
Introduction 9 1. INTRODUCTION 1.1 Overview Resolution and bandwidths The BARCO MGD 521 M and MGD 521 MKII displays are ultra-high resolution, grayscale portrait displays.
Introduction 10 adjust and check BARCOs medical displays. A lot of adjustments can also be done on the monitor itself, by means of an extensive on-screen menu system, accessible from the rear panel controls. Power saving system The MGD 521 displays are equipped with a power saving system.
Installation 11 1.3 About the manuals This guide is meant for people who want to install and use the MGD 521 displays, as well as people who need to install and adjust the displays. Chapter 4, Adjustments, is meant for this second category only, because it describes actions and procedures that require a technical skill to be performed properly.
Installation 12 2. INSTALLATION 2.1 Precautions Keep your original packaging. It is designed for this display and is the ideal protection during transport. Do not lift the display all by yourself to avoid injury. Avoid reflections in the picture tube to reduce eye strain.
Installation 13 2.3 Controls and connectors 1 REMOTE OUT REMOTE IN VIDEO HS / CS VS 2 4 5 6 7 8 9 10 11 3 {1} Optical sensor plug {2} Ambient Light Compensation (ALC) sensor (optional) {3} Green power LED {4} Control knob, combination of a push button and a turning knob.
Installation 14 2.4 Signal connection a) Connection of video signals The connected equipment must comply to all relevant safety demands. Important: The use of low-quality video cables can distort the video signal and influence diagnosis. 1. Check the impedance of the imaging board that produces the video signals you want to connect.
Installation 15 b) Connection of data signals The display can be controlled remotely by a computer through the serial data bus. A typical example of this, is the MediCal software that controls the display. MediCal runs on a PC that is connected through the serial data bus.
Installation 16 d) Connection of power 1. Power cord: Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 300 V min., provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application.
Installation 17 Important considerations The best environment for diagnostic imaging is one with controlled and dimmed ambient light. The human eye's sensitivity depends on the ambient light strength. It is most sensitive to small contrast changes (or subtle image details) at limited ambient light levels.
Operation 18 3. OPERATION: USER CONTROLS Important : The best way to adjust and control the display is by using MediCal, which displays the ideal test patterns for correct adjustment. However, in case you do not dispose of MediCal, you can control and adjust the display by means of the built-in On-Screen Display (OSD).
Operation 19 3.2 Contrast and brightness control Contrast and brightness can be controlled through the OSD menus. To activate the OSD Turn the control knob {4} from left to right or from right to left. As a result, the OSD main menu appears. Note: The control knob should turn over at least 30 degrees before the OSD becomes visible.
Operation 20 To control contrast manually 1 Turn the control knob to select Contrast in the OSD main menu. 2 Press the control knob shortly. The contrast control menu appears. 3 In the contrast control menu, select Manual contrast adjustment and press the control knob shortly.
Operation 21 STATUS INFORMATION Display Name: MGD 521MKII Serial Number: 5114147 Mode Name: New mode Mode H-frequency: 175.32 kHz Mode V-frequency: 79.94 Hz Software Version: V 1.01 Run time: 50 hours White level: 300 cd/m2 Activity timeout 30s SH compensation OFF Actual H-frequency:125.
Operation 22 Software Version: The internal software version. Run time: The total operating time since production, including burn-in time. White level: The displays white level or light output after calibration, in calibrated position.
Advanced 23 4. ADVANCED SETTINGS AND ADJUSTMENTS (FOR SERVICE STAFF ONLY) Important: The functions and controls described in this chapter can have a serious impact on the performance of the display. They should be touched by trained service staff only! 4.
Advanced 24 Explanation of the Settings menu items LUC: LUC is short for Luminance Uniformity Control. When switched on, the internal waveform processor will modulate the gain of the video amplifier so that the luminance is equal all over the screen (within a tolerance of +/- 5%).
Advanced 25 ADJUSTMENTS - Geometry - Focus - ALC - Luminance uniformity - Factory reset - Exit Adjustments menu 4.2 Adjustments a) How to make the adjustments 1 Turn the control knob at the rear from left to right or from right to left to activate the On-Screen Display (OSD) main menu.
Advanced 26 Width Horizontal position Height Vertical position Blanking - Blanking left Blanking - Blanking right Blanking - Blanking top.
Advanced 27 Interlacing - Pre correction In most cases you do not need to change this adjustment. You may need to increase this adjustment only when the connected video and sync signal contains an abn.
Advanced 28 SH Compensation: SH Position With SH Compensation on, we advise to use the SH Position control instead of the Horizontal Position control in the Geometry menu.
Advanced 29 Advanced geometry - Corner top right Advanced geometry - Corner bottom left Advanced geometry - Corner bottom right Advanced geometry - Linearity & S-correction - Horizontal linearity .
Advanced 30 Advanced geometry - Advanced bowing & skewing - Horizontal bowing left Advanced geometry - Advanced bowing & skewing - Horizontal bowing right Advanced geometry - Advanced bowing &.
Advanced 31 Focus Static focus Adjust Static focus to obtain an image as sharp as possible in the center. Focus zone You can adjust the focus in 25 independent zones. Turn the control knob to obtain an image as sharp as possible in the selected zone. Press the control knob to switch to the next zone.
Advanced 32 Luminance uniformity This menu allows you to calibrate the display, including the Luminance Uniformity circuit. Proceed as follows: 1 Connect the X-Rite DTP92 sensor and apply the black shield (see chapter "Installation"). 2 Follow the instructions on the screen.
Maintenance 33 5. MAINTENANCE Picture tube The glass panel of the picture tube is handled with a special coating. Take care not to damage or scratch the coating. Clean the picture tube with a soft woolen or cotton cloth. The cloth should be moist, not wet! Use a watery solution or a mild commercial glass cleaning solution.
Troubleshooting 34 6. TROUBLESHOOTING There appears no image on the picture tube, the green LED at the front is out Check if the power cord is properly connected to the power outlet and to the display. Check if the power button is switched on.
Background information 35 APPENDIX A: BACKGROUND INFORMATION Luminance Uniformity Correction A characteristic (or limitation) of every picture tube (CRT) is that the luminance decreases towards the edge of the screen surface. The decrease is normally 20 to 30 %.
Background information 36 At last, the imaging board delivers no sync signals at all, and the micro-controller switches off all but one power supplies in the display. Only the micro-controller's own supply keeps on running, resulting in a very low power consumption of 5 W.
Technical specifications 37 Picture tube: Faceplate transmission: +/- 32% Faceplate type: ARC panel Image representation: portrait Phosphor: P45 (standard) or P104 (option for MGD 521 MKII only) Light output: (P45 phosphor) MGD 521: Calibrated: 300 Cd/m2 MGD 521M: Calibrated: 400 Cd/m2 Peak: 600 Cd/m2 Resolution: Max.
Een belangrijk punt na aankoop van elk apparaat Barco MGD 521 M (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Barco MGD 521 M heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Barco MGD 521 M vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Barco MGD 521 M leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Barco MGD 521 M krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Barco MGD 521 M bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Barco MGD 521 M kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Barco MGD 521 M . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.