TW-357 REV E 1 LABS (Laboratory Archival Biological Storage) CRYOSTORAGE SYSTEM OPERATING AND MAINTENANCE INSTRUCTION REVIEW AND UNDERSTAND ALL SAFETY PROCEDURES IN FORM # TW-10 P/ N 7950-8052 BEFORE ATTEMPTING TO INSTALL, OPERAT E OR PERFORM MAINTENANCE ON THIS CRYOSTORAGE SYSTEM.
TW-357 REV E 2.
TW-357 REV E 3 Table of Contents WARNING 4 Safety Precautions for Liquid Nitrogen 4 GENERAL INFORMATION 6 Text Format Notation 6 LABS Cryostorage Specifications 7 INSTALLATION 9 Unpacking and Inspecti.
TW-357 REV E 4 WARNING The following safety precautions are for your protection. Before installing, operating, or maintaining this unit read and follow all safety precautions in this section a nd in reference publications. Failure to observe all safety precautions can result in property damage, personal injury, or possibly death.
TW-357 REV E 5 Dispose of Waste Liquid Nitrogen Safely Dispose of waste liquid nitrogen out-of-doors where its cold temperature cannot damage floors or driveways and where it will evaporate rapidly. An outdoor pit filled with clean sand or gravel will evaporate liquid nitrogen safely and quickly.
TW-357 REV E 6 GENERAL INFORMATION Thank you for purchasing this product. This state of the art CRYOSTORAGE SYSTEM can control either the liquid level and/or the vapor temperature range. The f eatures are designed to provide a safe environment for samples while at the same time tracking all relevant in formation associated with the freezer.
TW-357 REV E 7 LABS Cryostorage Specifications Liquid nitrogen at atmospheric pressure weighs 1. 78 lb./liter (0.8 kg/liter). All Taylor-Wharton Gas Equipment cryostorage systems are designed to support the full weight of liquid nitrogen and a complete stainless steel or aluminum inventory control system with boxes and specimens.
TW-357 REV E 8 KRYOS Specifications Configurations: Designed exclusivel y for all Taylor-Wharton LABS CryoStorage Systems Power Supply: 16.5 VAC, 40 Voltamps – Standard Battery Backup Version 24 VAC, 40 Voltamps (No Battery) (For power outlets other than North American 110 VAC contact Taylor-Wharton Customer Service.
TW-357 REV E 9 INSTALLATION Unpacking and Inspection Inspect shipping containers for external damage. All claims for dam age (apparent or concealed) or partial loss of shipment must be m ade in writing within five (5) days from receipt of goods. If damage or loss is apparent, please notify the shipping agent immediately.
TW-357 REV E 10 Power Supply Connection Connect the 16.5 Volt AC power supply to the control of the LABS cryostorage sy stem; then plug the power supply into a surge protected 110/120 VAC outlet. KRYOS Control Field Installation 1. Unplug power from old unit.
TW-357 REV E 11 OPERATION These instructions are for operators experienced with cryogenic equipment. Before operating the system, become familiar with the safety precautions in this ma nual and in reference publica tions.
TW-357 REV E 12 Control Components Interface Panel The KRYOS Interface panel, which the user will inter act with, contains the vacuum fluorescent display as well as the number keypad, power button, help button and the soft-key control buttons. (See figure 1.
TW-357 REV E 13 Lid Switch Is attached to the hinge and determines whether or not the lid is open on the freezer. This also allows the control to determine whether to activate the Qu ickchill, Auto Defog or Lid Alarm features.
TW-357 REV E 14 Operating Parameters When materials are imm ersed in liquid nitrogen, th ey will assume the temperature of the liquid -196ºC (- 320ºF).
TW-357 REV E 15 Thermocouple Positioning The thermocouple is a separate sensor u sed to monitor and control the temper ature within the refrigerator. The factory-supplied thermocouple in the LABS cryostorage system is positioned to monitor the temperature 30” above the tray.
TW-357 REV E 16 Inventory Control System Layouts (LABS Cryostorage System) Typical 13-2 Rack 100 Vial Polycarbonate 100 Vial Aluminum Boxes 81 Vial Polycarbonate, Cardboard, or Stainless Steel Boxes LABS-20K NOTE: RACKS ARE NOT SQUARE. ORIENTATION IS IMPORTANT.
TW-357 REV E 17 Blood Bag Cassette Blood Bag Storage in Cassette LABS-20K L A B S - 8 0 K L A B S - 4 0 K Fenwal 4R9951 Blood Bags – ask about other Bag types PART NUMBER CAPACITY FRAMES CASSETTES LABS-80K 2100 300 2100 LABS-40K 1204 172 1204 LABS-20K 560 80 560 Figure 6.
TW-357 REV E 18 Blood Bag Cassette Blood Bag Storage in Cassette LABS-20K LABS-80K LABS-40K Fenwal 4R9953 Blood Bags – ask about other Bag types PART NUMBER CAPACITY FRAMES CASSETTES LABS-80K 1505 301 1505 LABS-40K 820 164 820 LABS-20K 400 80 400 Figure 7.
TW-357 REV E 19 Blood Bag Cassette Blood Bag Storage in Cassette LABS-20K LABS-80K LABS-40K Fenwal 4R5462 – 9955 Blood Bags – ask about other Bag types PART NUMBER CAPACITY FRAMES CASSETTES LABS-80K 1050 210 1050 LABS-40K 590 118 590 LABS-20K 290 58 290 Figure 8.
TW-357 REV E 20 Control Operation This section of the operating manual is for Ta ylor-W harton approved equipment that uses the KRYOS controller. Introduction The KRYOS temperature and LN2 level controller is designed for easy operation and uninterrupted, reliable service.
TW-357 REV E 21 Main Display Screen The main display screen consists of 4 lines of information. Line 1: Displays the current status of the control. It indicates if all systems are normal or if any errors have been detected. Error messages disappear when the error is corrected.
TW-357 REV E 22 Table 1.0 Menu System 1. Temperature 1.1 Thermocouple Select 1.2 Calibrate Temperature 1.3 Test Temperature System 2. Level Sensing 2.1 Test Level Sensors 2.2 Set Sensor Offset 2.3 Sensor Positions 2.3.1 Start Set Point 2.3.2 Stop Set Point 2.
TW-357 REV E 23 Help Screens The Help button provides help to the user at any point in the menu system. The help message is displayed and the user is then prompted to press a button to return to the menu system. Temperature Thermocouple Select The chamber temperature is monitored with 1 or 2 Type T therm ocouples.
TW-357 REV E 24 Sensor Type The sensor type can be set through the menu system ( MENU , 2, 4). The sensor type selection should match the sensor type that is being used in the system. This is a 4-sensor, a 7-sensor (F reeze-Guard) or an 8-sensor array.
TW-357 REV E 25 System Alarms The Low LN2 Supply Alarm can be set through the menu system ( MENU , 3, 3, 1). This alarm is activated if the solenoid valve is not closed within a designated tim e period after a fill starts.
TW-357 REV E 26 The control consolidates the events in the internal log (combining “start ev ent” / ”conclude event” information to provide one event with duration), however, when the da ta is sent out the serial port no consolidation of data occurs.
TW-357 REV E 27 OPEN THERMOCOUPLE LOGGING is accessible through the menu system of the control ( MENU , 4, 2, 4). This menu choice turns on/off the logging of the thermocouple open alarm. The choices are ENABLE and DISABLE. Records data in the error log.
TW-357 REV E 28 The MENU PASSWORD can be set through the menu system of the control ( MENU , 5, 2). The menu password requires entry of a 4-digit password before allowing access to the menu system to change any control settings. Follow the directions on the display to set a new password or to disable a password.
TW-357 REV E 29 Menu Password When the menu password is active, it is possible to re view all of the control settings without having the password. When a password is activated and the u ser presses th.
TW-357 REV E 30 Control Range CONTROL RANGE can be accessed through the menu system of the control ( MENU , 6, 2, 3, 2, 2). KRYOS maintains a temperature range around the user-selected te mperature. A tighter range maintains a tem perature very close to the selected temperat ure at the cost of greater LN2 usage.
TW-357 REV E 31 Display Brightness DISPLAY BRIGHTNESS can be set through the menu system of the control ( MENU , 6, 3). This option changes the intensity of the display. The possible choices are 25%, 50%, 75% and 100%. About this Control ABOUT THIS CONTROL can be accessed in the menu system ( MENU , 6, 4).
TW-357 REV E 32 The DC solenoid valve requires that the battery be connected for the sy stem to work properly. If the battery is disconnected or completely discharged the valve opera tion and the control operation may be affected. In particular, if the control is calling for a fill, the sole noid will be called to open.
TW-357 REV E 33 MAINTENANCE Strainer Cleaning Instructions The container will not fill properly if the strainer is clogge d with ice or dirt. To clean the strainer, first close the supply valve to the refrigerator. Vent the fill line of all pressure. Remove and warm the strainer to am bient temperature.
TW-357 REV E 34 Cleaning Your Taylor-Wharton CryoStorage System The cryogenic vessel of all LABS CryoStorage Systems ma y need to be cleaned, and sterilized if the type of stored product is changed or the unit is taken out of service.
TW-357 REV E 35 WARNING: Never use chlorine-based disinfect ants to clean a LABS CryoStorage System. Normal Evaporation Rate (NER) Test Nitrogen consumption is an accumulation of all system components and user introduced evaporation.
TW-357 REV E 36 The Kryos Control installed on the Taylor-Wharton LABS 80K refrigerator operates w ith software version 9.0 (regular operation) or 9.0B (battery backup). This control software has b een specially designed to match the design characteristics of the LABS 80K refrigerator.
TW-357 REV E 37 • Move the sensor up or down so that the appropria te numbers of sensors are in liquid (read “L”). • Get back to the KRYOS main screen and the level indi cated should ma tch the physically measured reading. The Sensor Offset, the Start Fill and the Stop Fill can all be set through the KRYOS menu system ( MENU , 2).
TW-357 REV E 38 Sensor Plug Schematic Figure 9.0 Pin View Schematics.
TW-357 REV E 39 LABS Plumbing Assembly Solenoid Valve Freeze-Guard Sensor Male Branch Tee 90 deg Elbow 90 deg Elbow Male Branch Tee Reducer Relief Valve Tube Strainer Female Connector Relief Valve Figure 10.
TW-357 REV E 40 TROUBLESHOOTING Symptoms The key to troubleshooting your LABS Cryo Storage system is to determine which component in the system is the source of the problem.
TW-357 REV E 41 • Sim ulate an over fill by lowering the next thermi stor into the LN2 . HIGH LEVEL alarm should sound with 10 seconds. • Re-install sensor assem bly and thermocouple into sen sor tube as before.
TW-357 REV E 42 Fill Solenoid Cycles On and Off KRYOS features a Freeze-Guard function whereby the valv e open duration is limited, allowed to shut for a short warm-up and then opened again. To c onfirm or change the valve open duration: • Select MENU , USER OPTIONS, CONTROL OPTIONS, FREEZE-GUARD OPTIONS, VALVE OPEN DURATION.
TW-357 REV E 43 Defog Feature Will Not Operate See LID OPEN ALARM. To activate or adjust times: Select MENU , USER OPTIONS, CONTROL OPTIONS, LID/DEFOG SETTINGS, AUTO DE-FOG. Push Buttons Will Not Respond 1. Check all ribbon cable connectors on back of display panel.
TW-357 REV E 44 Factory Settings To restore factory settings at any time KRYOS can be re-initialized. Please note that a re-initialization clears all logs and passwords and resets settings to those listed in the table below. To reset factory settings in the control, go to the menu choice ABOUT THIS CONTROL.
45 REPLACEMENT PARTS Refrigerator Parts (LABS) 80K/40K/20K Lid Assembly……………………………………………………………LA80-9C85/LA40-9C85/LA20-9C85 Caster, ……………….
46 Fill in top section at installation. Copy this form each time service is required. Fill bottom section with service notes to keep a complete log of each freezer service and maintenance history.
47 4075 Hamilton Blvd. Theodore, Alabama 36582 U.S.A. Telephone (251) 443-8680 Fax (251) 443-2250 In U.S. and Canada: (800) TW TANKS (898-2657) ©2002 Harsco Corporation.
Een belangrijk punt na aankoop van elk apparaat Taylor TW-357 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Taylor TW-357 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Taylor TW-357 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Taylor TW-357 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Taylor TW-357 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Taylor TW-357 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Taylor TW-357 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Taylor TW-357 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.