Gebruiksaanwijzing /service van het product 201A van de fabrikant Teledyne
Ga naar pagina of 31
MANUAL ADDENDUM MODEL 201A AMMONIA ANALYZER TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (T-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.com WEB SITE: www.te le dy ne -ap i.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 SAFETY MESSAGES Your safety and the safety of others is very im portant. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 TABLE OF CONTENTS SAFETY MESSAGES ....................................................................................................... 1 TABLE OF CONTENTS ...................
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 LIST OF FIGURES F IGURE 1 - M201A A MMONIA A NALYZER – B LOCK D IAGRAM ................................................................. 21 F IGURE 2 - M201A E XTERNAL P NEUMATIC C ONNECTION D IAGRAM WITH ECO (E XTERNAL C ALIBRATOR O PTION ) .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 LIST OF TABLES T ABLE 2.1 - Z ERO C ALIBRATION P ROCEDURE - Z ERO G AS T HROUGH THE S AMPLE P ORT ......................... 11 T ABLE 2.2 - E NTER THE E XPECTED S PAN G AS C ONCENTRATIONS P ROCEDURE .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 5.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 1.0 M201A AMMONIA ANALYZER This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n 02246 supplied with this instrument.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 converts some of the ammonia to NO, t hus the TN channel data contains a small ammonia signal. Secondly, the ammonia conv erter does not convert all of the ammonia or ambient NO2 to NO.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Do NOT use copper tubing or fittings designed for household plumbing. Use ONLY Chromatography grade (cleaned, passivated) stainless steel tubing Use ONLY Glass tubing for sample inlet manifold.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 1.2 M201A Analyzer Specifications Ranges 0-50 to 0-2000 ppb in 1 ppb increments Range Modes Single, Independent, AutoRange Noise at Zero 1 0.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.0 CALIBRATION PROCEDURE Calibration of the M201A is done by first calibrating the TN and TNx channels with NO span gas diluted with zero air. In the second phase of calibration, ammonia span gas is introduced to measure the overall system e fficiency for NH3 conversion.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.1 - Zero Calibration Procedure - Zero Gas Through the Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode from sample mode. Input zero gas through the sample port.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode. Input NO span gas through the sample port.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.1.1 Zero/Span Calibration with Z/S Valves or ECO Option The zero gas enters through the zero gas inle t port on the rear panel. Make sure that the umbilical cord that makes the cont rol connections between the analyzer and the converter is connected (see Figure 2 below).
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.5 - Enter the Expected Span Gas Concentrations Procedure Step No. Action Comment 1. Press CAL- CONC- TNX This key sequence causes the M201A to prompt for the expected TNX concentration.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.6 - CALS - Span Cal Procedure - Gas from rear panel span gas inlet Step No. Action Comment 1. Press CALS The M201A enters the calibrate mode. The Z/S valves are activated to route span gas into the analyzer.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.1.3 NH3 Calibration Since the system efficiency is less than 100%, it is necessary to determine a conversion efficiency factor for ammonia and enter it into the system. 1. Set the converter efficiency to 1.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 The instrument is now fully calibrated. 2.1.4 Channel Calibration Quality After the ammonia channel is calibrated via the conv.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 3.0 MAINTENANCE SCHEDULE 3.1 M201A Maintenance Schedule The maintenance requirements of the M201A are essentially like a standard M200A NOx analyzer. Please refer to Table 8-1 in the M200A Operator Manual for the M200A maintenance schedule.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 3.2 – M501NH Converter Rebuild Parts List Part No. Description KIT000191 Converter Rebuild Kit KIT000139 Ceramic Bushings HE0000007 Ceramic Heater 220W 60 VAC Procedure: 1.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 3.3 AutoZero Flow Check Since the M201A is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows: 1. This procedure should be performed with the sample pump running.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 1 - M201A Ammonia Analyzer – Block Diagram 21.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 MODULE AIR ZERO FROM SPAN 02522 UMBILICAL CORD TO VACCUM PUMP POWER UMBILICAL TO CONV CONV FROM UMBILICAL SS TUBING GAS INLET .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 3 - M201A Pneumatic Diagram Basic 23.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 4 - M201A Pneumatic Diagram Z/S Option 24.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 5 - M201A Pneumatic Diagram External Calibrator Option (ECO) 25.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 6 – M501NH Converter Cartridge 26.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 27.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 4.0 M201A SPARE PARTS LIST 000940400 Orifice, 4 mil, 80 cc, Rx Cell 000940600 Orifice, 10 mil, 500 cc, Rx Cell 002270100 Gasket (Rx Cell) Qty.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 M201A Spare Parts List (cont.) 016810200 O3 Generator Assembly for M200A/200AH Low Output 017230000 Dryer Material, Drierite 018720000 Welded Moly Conv.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 M201A Spare Parts List (cont.) KIT000129 Welded Moly, Cartridge only KIT000191 Rebuild, NH3 INNER TUBE & CATALYST OR000000.
Een belangrijk punt na aankoop van elk apparaat Teledyne 201A (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne 201A heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne 201A vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne 201A leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne 201A krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne 201A bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne 201A kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne 201A . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.