Teledyne Analytical Instruments QUICKS T AR T GUIDE M O D E L 3 0 0 0 T A T ra ce Oxygen Analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING S T ARTED This Quickstart Guide is designed to get you s et up and ope rating your Teledyne An a lytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functi ons currently displayed on the VF D screen. • Up & Down Arrows Increment or decrement values of functions currently displayed.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 5 • Analog Output C onnecti on: Terminals for current and volt age a n alog output signals. • Digital Input Terminals: Termi n als for connecti ng d igital input signals. • Alarm Connections: 2 concentration alarms and 1 system alarm.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INS T ALL Mount the Ana lyzer The Model 3000T A is for indoor use in a general purpose are a. It is NOT for hazardous environme nts of any type. It is design ed fo r f lush p anel m ount ing.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibra tion feature: • Connect zero gas to t he Z ERO IN port. • Connect span ga s to the SP AN IN port. 5. Otherwise, c a li b ra t io n ga se s m u st b e T e e' d i nt o t h e s a m p l e i n le t wi t h a p p r o p ri a t e v a l ve s .
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power cord rece ptacle. • Install the fuse: Fuses are not installed at the fac tory. Be sure to install the proper fuse as par t of installati on.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There a re four ded icated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium ra nge Range 3 ID— High range Range 4 ID—Air Cal Ra nge (25%).
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With t h e O-ring in place, align the guide pin with the hole on the cell holder.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPER A TION Once the a naly zer has been installed, it can be configured for your application. Prior to using the instrument you m ust : • Calibrate t he instrument • Define the analysis r anges and choose whether to use autoranging or a fixed range.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 A vailab le Functions The six function butt ons are used to control the analyzer. Some functi ons have submenus that allow additional choices as s hown below. Any function ca n be sel ected at any time by pressing the appropriate button (unless password restrictions appl y).
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting the following system parameters can be done only after t he passw ord is en.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto — Cal PSWD Logou t M ore • Use the ◄► arrow keys to scroll the blin king over to PSWD , • Press Enter to select the password function.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUT O and MAN zero settling. Stop when AUT O appears blinking. Zero : Settling: AUTO <ENT> To B e gin • Press Enter to begin zeroing.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connecte d to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val : 000008.00 <ENT>Span <UP>Mod # • Press ▲ (<UP>) to permit modification ( Mod # ) of span value.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latc hing? Ltch–Y/N (Yes/No). • To change any of th e alarm parameters, use the ◄► arrow keys to move the blinkin g over to the specific par ameter. • Use the ▲▼ arrow keys to change t he va lue of the parameter.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO t o FX/L O , FX/MED , or F X /HI to set the instrument on the desired fixe d range (low, medium , or h igh).
Een belangrijk punt na aankoop van elk apparaat Teledyne 3000TA (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne 3000TA heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne 3000TA vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne 3000TA leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne 3000TA krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne 3000TA bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne 3000TA kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne 3000TA . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.