Gebruiksaanwijzing /service van het product 81-0112-01 Rev C van de fabrikant UVPS
Ga naar pagina of 22
Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV Crosslinker to familiarize yourself with its operation.
UV CROSSLINKERS UVP CONTENTS 1. Introduction 1 2. Important Safety Information 1 3. Description and Specifications 1 4. Operation 3 5. Applications 6 6. Maintenance, Care and Cleaning 7 7. Changing the UV Wavelength 8 8. Calibration Procedure 10 9. Replacement Parts 12 10.
UV CROSSLINKERS UVP 1. INTRODUCTION UVP Crosslinkers offer researchers an instrume nt to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. CL-1000 series features a lay-down type drawer while the CX-2000 has a pull-out dr awer.
UV CROSSLINKERS UVP Features • Microprocessor controlled/UV sensor feedback system • Multiple set functions- preset UV energy exposure preset UV time exposure user set UV energy exposure user set UV time exposure • Maximum UV energy exposure setting of 999,900 m icrojoules per cm2 • Maximum UV tim e exposure setting of 999.
UV CROSSLINKERS UVP 4. OPERATION 1. Place the crosslinker on a level work surface. Be sure there is enough room in front for the door to open easily. 2. Plug the female end of the power cord into the crosslinker. 3. Plug the male end of the power cord into a properly grounded electrical outlet.
UV CROSSLINKERS UVP The new setting will appear on the LED display. Push ENTER on the touchpad. The new setting is now installed. B. Preset ultraviolet time exposure Push PRESET and then TIME on the tactile touchpad.
UV CROSSLINKERS UVP Push START on the touchpad. After a slight delay to energize the UV tubes, The LED will begin to countdown. the unit will stop automatically at the end of the exposure cycle and will beep five times.
UV CROSSLINKERS UVP 5. APPLICATIONS The Crosslinker is a multi-purpose ultraviolet e xposure instrum ent for use in the laboratory. A wide variety of applications for ultraviolet radiation exist in the laboratory.
UV CROSSLINKERS UVP 6. MAINTENANCE, CARE AND CLEANING OF THE CROSSLINKER The UVP UV Crosslinker is built to provide you w ith trouble-free operation. To ensure correct service: NOTE: ALWAYS UNPLUG UNIT FROM ELEC TRICAL SUPPLY BEFORE CLEANING OR DRYING.
UV CROSSLINKERS UVP 7. CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, l ongwave or midrange tubes. However, if user requirements and applications change, the Crosslinkers provide you with the unique ability to change the ultraviolet wavelength and recalibrate the UV sensor and microprocessor to the new UV wavelength.
UV CROSSLINKERS UVP 11. The previous calibrated operation of the crosslinker will continue following any unsuccessful recalibration. Replace the new UV wavelength tubes with the previously removed old wavelength tubes.
UV CROSSLINKERS UVP 8. CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: Chart recorder Two probes with banana plug adapters One radiometer with UVX-25 calibrated sensor Cable with banana plug and jack adaptor to connect radiometer to chart-recorder Power source (100V, 115V or 230V) Procedure: 1.
UV CROSSLINKERS UVP 9. Run the Crosslinker in calibration mode by depressing the STOP/RESET button and turning the power switch on at the same time. At this point, both pens should start recording information on paper.
UV CROSSLINKERS UVP 9. REPLACEMENT PARTS AND SERVICING PROCEDURES PART NO. A. OPERATIONS WITHOUT REMOVING THE TOP COVER 5 x 34-0007-01 Changing tubes B.
UV CROSSLINKERS UVP 10. GENERAL SERVICING PROCEDURES A. Operations Without Removing The Top Cover Changing the Tubes It is recommended to change all FIVE 8-W att 254nm UV tubes at the sam e time. The unit does not require recalibration if you are replacing the old sources with new sources of the same wavelength.
UV CROSSLINKERS UVP 5. Cut all of the cable ties coming from the tube holder following the two wires from the tube holder to where it terminates. Note: ther e will be two diff erent termination points for each tube holder. Once this has been located, cut the wires (marking them to ensure easy replacement).
UV CROSSLINKERS UVP 6. To secure the starter holder, place the holder in your right hand and with your left hand in the chamber, press the now loosened reflector adjacent to the holder inwards. This will push the black spigots inwards thereby securing the starter holder.
UV CROSSLINKERS UVP Replacement of UV sensor UVP recommends returning the Crosslinker to the UVP factory for service. This is located on the right hand side of the internal chassis assembly at the back near the base. 1. Unplug the sensor connector from the PCB board marked J2.
UV CROSSLINKERS UVP C. Operations With The Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro switch located on the bottom left internal chassis assem bly. 2. Release the main power cable to ballasts - black wire with connector (color red).
UV CROSSLINKERS UVP E. General Items Changing fuses The fuses are located on the back panel and are marked with a self-adhesive sticker. 1. Turn fuse cap 15 degrees anti-clockwise. 2. Pull fuse and cap from the holder. 3. Pull fuse from the cap. 4. Replace the fuse.
UV CROSSLINKERS UVP Schematic Wiring Diagram - 19 -.
UV CROSSLINKERS UVP Internet: www.uvp.com Corporate Headquarters : UVP, Inc. 2066 W. 11th Street Upland, CA 91786 Tel: (800)452-6788 or (909)946-3197 Fax: (909)946-3597 / E-Mail: info@uvp.
Een belangrijk punt na aankoop van elk apparaat UVPS 81-0112-01 Rev C (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen UVPS 81-0112-01 Rev C heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens UVPS 81-0112-01 Rev C vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding UVPS 81-0112-01 Rev C leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over UVPS 81-0112-01 Rev C krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van UVPS 81-0112-01 Rev C bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de UVPS 81-0112-01 Rev C kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met UVPS 81-0112-01 Rev C . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.