Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe Revision: 12/06 Copyright © 2001-2006 Campbell Scientific, Inc..
Warranty and Assistance The CS512 OXYGUARD TYPE III DISSOLVED OXYGEN PROBE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materi als and workmanship under norm al use and service for t welve (12) months from date of shipm ent unless specified otherwi se.
i CS512 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. General Information ................................................
CS512 Table of Contents ii Examples 1. Sample Program using P1 .......................................................................... 5 2. Sample Program using P2 .......................................................................... 5 3. Sample Program using VoltSE .
1 Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe An accurate and reliable sensor is a critical element in any measurement system. The Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 2 2. Specifications Principle of Measurem ent: Membrane covered galvanic oxygen probe Output Signal: 2.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 3 3. Optional Probe Accessories PT4-L — Agitator for stagnant condi tions 18026 — DO electrolyte, 1 liter 18027 — Membrane kit , 10 mem branes 4.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 4 FIGURE 4-2. CS512 with PT4 Agitator Optional Repeat Cycle Timer for Agitators : Reciprocating act ion may be controlled by the opti onal solid state Repeat Cycle Tim er. It sends 12 VDC pulses to the agitator coil.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 5 7. Programming In the CR510, CR10X, and CR 23X datalogger, Instruction P1 or P2 can be used. Example 1 uses a single-ended m easurement for the CS512 probe; exampl e 2 uses a differential measurement.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 6 8. Calibration The multiplier is used to calibrate the CS512 probe. To calculate the m ultiplier: 1) Program the datalogger using a multiplier of one. 2) Place the OxyGuard probe in the air, shad ed from the sun.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 7 9. Maintenance The CS512 probe needs little maintenance. Regular cleaning of the m embrane is all that is required. The membrane is ve ry durable and can be cleaned with a cloth or soft paper.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 8 EXAMPLE 8. Sample Program using P86 and P22 Instructions 45: Do (P86) 1: 41 Set Port 1 Hi gh 46: Excitation with Delay (P22 ) 1: 1 Ex Channel 2: 20 Delay W/Ex (uni ts = 0.01 sec) 3: 0 Delay After Ex (uni ts = 0.
A-1 Appendix A. Dissolved Oxygen in Fresh Water Table.
Appendix A. Dissolved Oxygen in Fre sh Water Table A-2.
This is a bl ank page ..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
Een belangrijk punt na aankoop van elk apparaat Campbell Manufacturing CS512 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Campbell Manufacturing CS512 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Campbell Manufacturing CS512 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Campbell Manufacturing CS512 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Campbell Manufacturing CS512 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Campbell Manufacturing CS512 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Campbell Manufacturing CS512 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Campbell Manufacturing CS512 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.