Models 140, 141, and 142 Conductivity Sensors Instruction Sheet PN 51A-140/rev .E December 2010 SENSOR SPECIFICA TIONS For additional information, please visit our website at www .
Keep 1/4 inch (6 mm) clearance between electrodes and piping. The electrodes must be completely sub- merged in the process liquid, i.e., to the level of the threaded connection. See Figure 1 for recommended orientation. If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes.
3 MODEL 140,141, and 142 sensors INST ALLA TION FIGURE 3. Inst alling model 141 sensor DWG . NO. REV . 40014204 B INST ALLA TION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor .
RTD RTD in RTD sense RTD return MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor . Wiring connections in the junction box are shown in Figure 5. Figure 5. Sensor junction box wiring.
5 MODEL 140,141, and 142 sensors WIRING FIGURE 8. Model Xmt-C-10 Wiring (Panel) FIGURE 9. Model Xmt-C-1 1 Wiring (Pipe or Wall) RCV B RCV A RSHLD DRVB RTN SENSE RTD IN SHLD CLEAR WHITE/RED RED DRV A DSHLD WHITE CLEAR GRA Y ORANGE CLEAR TB1 TB2 FIGURE 10.
RETRACTING AND INSERTING THE MODEL 140 SENSOR RETRACTING THE SENSOR 1. Push in on the sensor junction box and slowly loosen the sensor compression fitting nut by reversing the sensor tightening procedure illustrated in Figure 2.
MODEL 140,141, and 142 sensors WIRING 7 3. Slowly open the valve. 4. Insert the sensor up to the desired insertion depth and turn the sensor compression fitting nut until it is finger tight. 5. Position the entire sensor for easy access to the ball valve handle, sensor compression fitting nut and J-box terminal block.
MODEL 140,141, and 142 sensors CALIBRA TION 8 CALIBRA TION Model 140 sensors are not calibrated at the factory . The cell const ant on the label is a nominal value only .
9 MODEL 140,141, and 142 sensors MAINTENANCE MAINTENANCE CLEANING THE SENSOR Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
10 MODEL 140,141, and 142 sensors ACCESSORIES PN Description 23724-00 Ball valve kit for Model 140 sensor 23730-00 Process compression fitting kit (3/4-inch NPT) for Model 140 sensor 23731-00 Process .
11 TROUBLESHOOTING PROBLEM PROBABLE CAUSE SOLUTION Off-scale reading Wiring is wrong. V erify wiring. T emperature element is open or shorted. Check temperature element for open or short circuits. See Figure 14. Sensor is not in the process stream. Be sure the sensor is completely submerged in the process stream.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
Een belangrijk punt na aankoop van elk apparaat Emerson 141 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Emerson 141 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Emerson 141 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Emerson 141 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Emerson 141 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Emerson 141 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Emerson 141 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Emerson 141 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.