Gebruiksaanwijzing /service van het product NGA 2000 van de fabrikant Emerson
Ga naar pagina of 178
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98 NGA 2000 Software Manual CLD - Anal yz er Module Chemiluminescence - Measurement (combined with NGA 2000 - Platf orm / ML T - Anal yz er) CLD Software V er sion 2.3 (Platf orm / ML T Software V er sion 3.
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98 This Operation Manual includes inf or mation about the oper ation of the instr ument. Inf or mation about the additional indications and notes regarding maintenance, troubleshooting and repair are f ound in the accompanying Maintenance & Operation Manual.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 Contents 1 Introduction 1 - 1 2 Structure of Menus 2 - 1 3 Display and Keyboard 3 - 1 3.1 Starting and Initializing ......................................................................
NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 5 Expert Controls and Set up 5 - 1 5.1 Analyzer Module Set up 5 - 3 5.1.1 Calibration/Calibration Gas List ...................................................................... 5 - 4 5.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 Technical Level Configuration 6 - 1 6.1 System Set up 6 - 3 6.1.1 Main Display Configuration ............................................................................ 6 - 4 6.1.2 Front Panel Control .
NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 IV.
1 Introduction 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 1 - 1 This software manual describes step by step how to operate successfully with the ½ 19" and 19" MLT analyzer module and analyzer of the NGA 2000 Series from .
1 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 The following illustrations shall make plain the connection between the hardware configu- ration and the software setup of the modules: Platform CLD FID PMD SIO DIO Other I/O´s (see 5.
2 Structure of Menus 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 2 - 1 Notes: * In the menu structure shown above, you will find the most important ramifications of the MLT software 3.1.X for the MLT platform and analyzer versions.
2 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X] 04/98.
3 Display and Keyboard 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 3 - 1 3.1 Starting and Initializing After switching on the CLD analyzer module (in a platform or part of a NGA network), the initialization procedure will be performed.
3 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 3.3 Operating Keys 3.4 Lines Lines can be selected by the ↓ ↓ -key or the ↑ ↑ -key. The selected line is displayed white on black. You have four different types of lines in the menu: Menu line.
3 Display and Keyboard 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 3 - 3 3.5 Important Functions of the Softkeys DISPLAY ♦ Change from the single component display to the multi component display. MEASURE ♦ Change from the main menu to the single component display.
3 - 4 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 MENU ♦ Change from the single component display to the main menu or from several submenus to higher menus. ESCAPE or <<< ♦ Changing back to the last menu page selected.
3 Display and Keyboard 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 3 - 5 3.7 The STATUS Key From the single component display or from the main menu you can change with the F2 key .
3 - 6 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4 Basic Controls 90003479(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 1 In the chapter "Basic Controls" you will find the most important functions to set up your CLD analyzer module via an NGA front panel: 1) Measurement p. 4 - 3pp (Chapter 4.
4 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4.1.1 Measurement - Multi Component Display: Change of Channel 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 3 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 4 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 INFO CLD F1 F2 F3 F4 F5 SELECT STATUS TAGS OFF MLT25/CH2/R2 ppm SO2 333.0 MLT25/CH3/R2 ppm NO 150.
4.1.2 Measurement - Multi Component Display: Sequence setup 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 5 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 6 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 INFO CLD F1 F2 F3 F4 F5 SELECT STATUS TAGS OFF MLT25/CH3/R2 ppm NO 150.0 MLT25/CH4/R2 %O2 20.00 CLD ppm NOx 95.00 MLT25/CH1/R2 ppm CO 45.00 MLT25/CH2/R2 ppm SO2 333.0 INFO MLT25/CH2/R2 F1 F2 F3 F4 F5 SELECT STATUS TAGS OFF MLT25/CH3/R2 ppm NO 150.
4.1.3 Measurement - Range Setting 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 7 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 8 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ESCAPE INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4.1.3 Measurement - Range Setting 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 9 MEASURE STATUS NEXT LOCK INFO CLD F1 F2 F3 F4 F5 95.0 ppm Main Menu Basic controls... Expert controls and set up... Technical level configuration.
4 - 10 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4.1.4 Measurement - Range and Functional Control 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 11 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 12 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4.1.4 Measurement - Range and Functional Control 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 13 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: AUTO Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4 - 14 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ⇒ ⇒ Range and functional control – Options: • LOCAL: The platform will be controlled via the display. • AUTO: The range will be selected automatically (" auto range").
4.1.5 Measurement - Enable/ Disable Ozonator Power 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 15 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 16 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4.1.5 Measurement - Enable/ Disable Ozonator Power 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 17 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4 - 18 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4.1.6 Measurement - Measurement Mode: NO ↔ ↔ NO x 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 19 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 20 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4.2.1 Calibration - Zeroing 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 21 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 22 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME ESCAPE ZERO INFO CLD F1 F2 F3 F4 F5 95.0 ppm Analyzer zero Are you sure? You must have zero gas flowing through the analyzer. This control does NOT control any auto-calibration module bound to this analyzer! If you are sure, press ZERO again now.
4.2.1 Calibration - Zeroing 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 23 MEASURE STATUS NEXT LOCK INFO CLD F1 F2 F3 F4 F5 0.
4 - 24 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4.2.2 Calibration - Spanning 90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 25 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 26 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME ESCAPE SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Analyzer span Are you sure? You must have span gas flowing through the analyzer. This control does NOT control any auto-calibration module bound to this analyzer! If you are sure, press SPAN again now.
4.2.2 Calibration - Spanning 90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 27 MEASURE STATUS NEXT LOCK INFO CLD F1 F2 F3 F4 F5 225.
4 - 28 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
4.2.3 Calibration - If It Won´t Calibrate 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 29 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 30 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 HOME NO/NOx ZERO SPAN INFO CLD F1 F2 F3 F4 F5 95.0 ppm Basic Controls Measurement range number: 4 Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate.
4.2.3 Calibration - If It Won´t Calibrate 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 4 - 31 DISPLAY STATUS MENU NEXT INFO CLD Sample press.: Detector temperature: Block temperature: Converter temp.: 51.5 C 0.5 C 340 hPa 360 C F1 F2 F3 F4 F5 0 250 95.
4 - 32 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 1 If you press the ENTER key or the → → -key in the line "Expert controls and set up.
5 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Structure of chapter five: At the begin of any chapter you will find the way to a certain submenu of the MLT software starting from the line "Expert controls and set up" in the "Main Menu".
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 3 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ From the menu "Analyzer module set up" you can change to several submenus, where you can set up the measurement and calibration parameters of the CLD analyzer module.
5 - 4 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Calibration gas list ↓ ↓ In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 5 Span gas calibration notes: ♦ The span gas setpoint of NO and NO x have to be the same value, because both components will be calibrated simultaneously! ♦ You can calibrate all ranges together or each range separately with span gas.
5 - 6 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 3) Zeroing: • Change to the submenu "Analyzer zero" with the F3 key (ZERO). • Pay attention to the notes on the screen ! • Flow enough zero gas through the CLD. • Wait for a stable signal (look at the headline of the display).
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 7 ♦ Select the range you want in the line "Measurement range number" and change via the line "Range 1/2/3/4 factors.
5 - 8 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Saving the factors: ♦ You can store the actual zero offset and span factor with the F2 key (STORE) of the menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu page (see next illustration).
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 9 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Calibration parameters ↓ ↓ In the menu "Calibration Parameters" you can set up several parameters of the zero and span calibration.
5 - 10 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Line of variables "Calibration averaging time": The value in this line will determine the time range the analyzer will use to get the calibration average. The precision will be better with longer times.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 11 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ From the menu.
5 - 12 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ Linearization.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 13 Lines of variables "Range 1/2/3/4 linearizer": In these lines you can enable the linearization for the range selected. You have to set up "Disabled" in these lines as long as you will determine the data for linearization.
5 - 14 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ Response time.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 15 Line of variables "LON update rate": In this line you can set up the update rate of the network variable.
5 - 16 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ Range setting ↓ ↓ In the menu "Range Settings" you can set up the begin and the end of each range.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 17 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ Units ↓ ↓ In the menu "Units" you can set up the units of the analyzer for pressure, temperature and gas concentration.
5 - 18 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Gas measurement parameters ↓ ↓ Linearization.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 19 1) Select the range to be linearized: • Press the ENTER key or the → → -key in the line "Range to be linearized" to select the parameter. • Select the range you want with the ↑ ↑ -key or the ↓ ↓ -key and confirm it with ENTER.
5 - 20 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ Input of the desired values: • You will need minimum 6 values to calculate a polynomial curve.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 21 6) Correction of the linearization polynom, if necessary: • You can optimize the calculated linearization with maximum three additional points. • Press the ENTER key or the → → -key in the line "Midpoint correction set up.
5 - 22 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 f) Set up further correction points: • You can repeat the steps c) to e) to determine maximum two further correction points. • Select "Point 2" resp. "Point 3" in the line "Point being measured".
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 23 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Analyzer Parameter List ↓ ↓ In the menu "Analyzer Parameter List" you can set up parameters of the single component display, of the ozonator and the analyzer tag.
5 - 24 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Line of variables " Ozonator power": Options: Enabled/Disabled The setup of this line will change the displayed message of the line " Ozonator status". That line is not available.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 25 ♦ Calibration failure alarm ♦ Calibration failure error allowed ♦ Calibration time out ♦ Zero and span gas calibration o.
5 - 26 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Physical measurement parameters ↓ ↓ In the m.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 27 Menu "Pressure Limits": Menu "Temperature Limits": Set up parameters: ♦ Select any line of variables with the ↓ ↓ -key or the ↑ ↑ -key. ♦ Select the variable with the ENTER key or the → → -key.
5 - 28 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Analyzer module set up ↓ ↓ Displayed parameters ↓ ↓ In the menu "Displayed Parameters" you can determine which parameters shall be displayed in the last four lines of the single component display.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 29 Main Menu — Expert controls and set up ↓ ↓ Auxiliary module controls ↓ ↓ From the menu "Auxiliary module controls.
5 - 30 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Auxiliary module controls ↓ ↓ Set-up SIO module ↓ ↓ From the menu "LOCAL SIO CONFIGURATION SECTIONS" you can change to submenus to control and set up several output configurations of the auxiliary SIO module.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 31 Analog Output Setup: Press the ENTER key or the → → -key in the line "Analog output setup.
5 - 32 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ Go to the line you want with the ↑ ↑ -key or the ↓ ↓ -key. ♦ Select the tag of the channel with the ENTER key or the → → -key: The display will go back to the menu "ANALOG OUTPUT SETUP" automatically.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 33 ♦ Example: • Range from 0 to 1000 ppm • 0% value shall be 400 ppm, 100% value shall be 700 ppm • Analog output is normal.
5 - 34 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 7) Fine adjustment of the analog output: If you press the F4 key in the menu "ANALOG OUTPUT SETUP" you can change to t.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 35 Serial interface setup: Press the ENTER key or the → → -key in the line "Serial interface setup.
5 - 36 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Configuration of Relay Outputs: Press the ENTER key or the → → -key in the line "Configuration of Relay Outputs..." of the menu "LOCAL SIO CONFIGURATION SECTIONS" to change to the corresponding menu: There are three relays on any SIO board.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 37 3) Select any signal: ♦ Change to the line "Choose signal..." with the ↓ ↓ -key. ♦ Press the ENTER key or the → → -key to change to the submenu "Signals": ♦ Go to the line you want with the ↓ ↓ -key or the ↑ ↑ -key.
5 - 38 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 All signals of the control module and of the analyzer modules joint to the platform or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "SIGNALS". The signals shown above will appear by using a CLD module as source module.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 39 Main Menu — Expert controls and set up ↓ ↓ Auxiliary module controls ↓ ↓ Set-up DIO module(s) ↓ ↓ In the menu "DIO-MODULE OUTPUTS" you can set up the output configuration of the DIO modules combined to the control module (platform or MLT analyzer).
5 - 40 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Line of variables "Invert output": Select "Enabled" in the line "Invert output" if you want to invert the measurement signal at the actual digital output, otherwise select "Disabled".
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 41 ♦ Go to the line you want with the ↑ ↑ -key or the ↓ ↓ -key. If there are more than eight source modules available you have to press the F3 key to go to another menu page.
5 - 42 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 • SYS:AM-Span-Gas1, 2, 3, 4 • SYS: Function control • AM: Hold Outputs • CLD-AM: NOx ♦ Select the function with the ENTER key or the → → -key: The display will go back to the menu "DIO-MODULE INPUTS" automatically.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 43 ♦ Go to the line you want with the ↑ ↑ -key or the ↓ ↓ -key. If there are more than eight source modules available you have to press the F3 key to go to another menu page.
5 - 44 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 45 Main Menu — Expert controls and set up ↓ ↓ Auxiliary module set up ↓ ↓ In the menu "Auxiliary module set up".
5 - 46 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 47 Main Menu — Expert controls and set up ↓ ↓ System set up ↓ ↓ From the menu "System Set Up" you can change to several submenus to set up different parameters of the LCD and the single component display of the platform.
5 - 48 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ System set up ↓ ↓ Front panel control ↓ ↓ In the menu "Front panel control" you can set up several display parameters.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 49 Line of variables "Reading refresh rate": In this line you can determine how often the display shall be built up in a certain time interval. Options: 1s to 60s. Standard: 1s.
5 - 50 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ System set up ↓ ↓ Display resolution ↓ ↓ In the menu "Display resolution" you can determine the number of digits for the display of the measured values.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 51 Main Menu — Expert controls and set up ↓ ↓ System set up ↓ ↓ Auxiliary lines ↓ ↓ In the menu "Auxiliary lines" you can determine if the last four lines in the single component display of the current channel shall appear.
5 - 52 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ Go to the line you want with the ↑ ↑ -key or the ↓ ↓ -key. If there are more than eight source modules available you have to press the F3 key to go to another menu page.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 53 Main Menu — Expert controls and set up ↓ ↓ Expert analyzer controls ↓ ↓ In the menu "Expert Controls" you can set up several measurement parameters of the analyzer as range or gas selection.
5 - 54 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ AUTO: The range will be selected automatically ("auto range"). This function will only be available with an I/O bo.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 55 Main Menu — Expert controls and set up ↓ ↓ Expert analyzer controls ↓ ↓ Range settings ↓ ↓ In the menu "Range Settings" you can set up the begin and the end of each range.
5 - 56 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Expert analyzer controls ↓ ↓ Zero/Span calibration ↓ ↓ In the menu "Zero/span calibration" you can set up and start the calibration of the analyzer.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 57 4) Spanning: • Change to the submenu "Analyzer span" with the F4 key (SPAN). • Pay attention to the notes on the screen ! • Flow enough span gas through the CLD. • Wait for a stable signal (look at the headline of the display).
5 - 58 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Set up parameters: ♦ Press the ENTER key or the → → -key in the line "Zero offset" to select the parameter and set up "32700" resp. "524000" with the ↑ ↑ -key or the ↓ ↓ -key.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 5 - 59 Saving the factors: ♦ You can store the actual zero offset and span factor with the F2 key (STORE) of the menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu page (see next illustration).
5 - 60 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Expert controls and set up ↓ ↓ Expert analyzer controls ↓ ↓ Physical Measurements ↓ ↓ In the menu "Physical Measurements" you can find some pressure results and temperature results of the CLD analyzer module.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 1 If you press the ENTER key or the → → -key in the line "Technical level configuration.
6 - 2 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Structure of chapter six: The structure of chapter six is analogous to the structure of chapter five: At the begin of any chapter.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 3 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ From the menu „System set up“ you can change to sev.
6 - 4 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ Main display configuration ↓ ↓ From the menu ".
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 5 Set up the number of digits: ♦ Select one of the two lines of variables with the ↓ ↓ -key or the ↑ ↑ -key. ♦ Select the number with the ENTER key or the → → -key. ♦ Select the digit with the ↑ ↑ -key or the ↓ ↓ -key.
6 - 6 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 In the menu "Auxiliary lines" you can determine if the last four lines in the single component display of the current channel shall appear. If you want to set up the lines of the other channels of the joint analyzer modules, press the F3 key.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 7 In the menu "Units control" you can determine the switch point for the signal value from ppb to ppm, respectively from ppm to vol%. But, the set-ups in this menu are only operative for earlier versions of analyzers or analyzer modules.
6 - 8 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Menu line "Editing controls...": From the menu "Main display configuration" you can change via the menu line "Editing controls.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 9 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ Front panel control ↓ ↓ In the menu "Front Panel Control" you can set up several display parameters.
6 - 10 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Line of variables "Reading refresh rate": In this line you can determine how often the display shall be built up in a certain time interval. Options: 1s to 60s. Standard: 1s.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 11 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ Date and time ↓ ↓ In the menu "Date and Time" you can set up the date and the time of the control module (platform or MLT analyzer).
6 - 12 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ Module binding ↓ ↓ In the menu "Module Bindin.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 13 ♦ Change to the line you want with the ↑ ↑ -key or the ↓ ↓ -key. If there exist more than eight unbound I/O modules (e.g. Autocal I/O) you can change to another menu page with the F3 key (MORE).
6 - 14 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ System Reset ↓ ↓ You can start the following funct.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 15 Main Menu — Technical level configuration ↓ ↓ System set up ↓ ↓ Security codes ↓ ↓ In the menu "Record securi.
6 - 16 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ Put in the code you want with the function keys F1, ..., F5: The sequence of the code numbers will appear in the line "Actual code number". Note: "1" is attached to F1, "2" to F2 and so on.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 17 Main Menu — Technical level configuration ↓ ↓ Service menus ↓ ↓ From the menu "Service menus" you can change .
6 - 18 NGA 2000 90003749X(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Service menus ↓ ↓ Manufacturing data ↓ ↓ Control module data ↓ ↓.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 19 Main Menu — Technical level configuration ↓ ↓ Service menus ↓ ↓ Manufacturing data ↓ ↓ Analyzer module data ↓ ↓.
6 - 20 NGA 2000 90003749X(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Lines of variables "Hardware revision number" / "Software revision number": In these information lines you will find the hardware revision number and the software revision number of the CLD analyzer module.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 21 Main Menu — Technical level configuration ↓ ↓ Service menus ↓ ↓ Service history ↓ ↓ Control module history ↓ ↓ .
6 - 22 NGA 2000 90003749X(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Set up notes: ♦ Select the line you want with the ↑ ↑ -key or ↓ ↓ -key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 23 Main Menu — Technical level configuration ↓ ↓ Service menus ↓ ↓ Service history ↓ ↓ Analyzer module history ↓ ↓ In the menu "Analyzer module service history" you will find the manufacturing date and the service date of the CLD analyzer module.
6 - 24 NGA 2000 90003749X(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Menu line "List notes...": Via the line "List notes..." you can change to the submenu "Analyzer module .
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 25 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ From the menu "Diagnostic menus" you can change to submenus where you will find software error messages of the control module or of the analyzer module.
6 - 26 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Control module diagnostics ↓ ↓ In the menu "Control Unit Diagnostics" you can find software error messages of the control module.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 27 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ From the menu &.
6 - 28 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Power supply vo.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 29 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Primary variabl.
6 - 30 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Lines of variables "Reference duty cycle" and "Duty cycle coefficient": These parameters will be set up during the first startup of the CLD analyzer module by the manufacturer.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 31 Set up parameters: ♦ Press the ENTER key or the → → -key in the line "Zero offset" to select the parameter and set up "32700" resp. "524000" with the ↑ ↑ -key or the ↓ ↓ -key.
6 - 32 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ You can change the stored settings against the manufacturer's settings with the F3 key (RSTR MN: R e st o r e the m an ufacturer's values).
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 33 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Physical Measur.
6 - 34 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Menu "Pressure Limits": Menu "Temperature Limits": Set up parameters: ♦ Select any line of variables with the ↓ ↓ -key or the ↑ ↑ -key. ♦ Select the variable with the ENTER key or the → → -key.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 35 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Temperature con.
6 - 36 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Flow control parameters ↓ ↓ In the menu "NO/ NOx flow balance" you can set up or calculate an NO x to NO flow correction factor.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 37 Line of variables " NOx correction factor": In this line you can set up the NO x to NO flow correction factor. Options: 0.1 to 10 Menu line "Calculate factor using pressure ration.
6 - 38 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Miscellaneous control parameters ↓ ↓ In the menu "Miscellaneous control parameters" you will find some converter parameters and detector parameters.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 39 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Trend display control ↓ ↓ The analyzer stores 24 hours of 15 minute averages.
6 - 40 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 ♦ Converter temperature ♦ Block temperature ♦ Detector temperature ♦ Ozone pressure Line of variables " Timebase": In this line you will find the period of storing values.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 41 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Software diagnostics ↓ ↓ In the menu "Software diagnostics" you can find software error messages of the CLD analyzer module software.
6 - 42 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ Start up analyzer ↓ ↓ In the menu "Analyzer start up" you will find the actual and target temperatures of the converter and of the detector block.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 43 Main Menu — Technical level configuration ↓ ↓ Diagnostic menus ↓ ↓ Analyzer module diagnostics ↓ ↓ NOx converter efficiency ↓ ↓ In this menu you can optimize NOx converter efficiency.
6 - 44 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 Main Menu — Technical level configuration ↓ ↓ Other module diagnostic menus ↓ ↓ Further menus are not available at the moment: In later software revisions it shall be possible to control the internal working status (e.
90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98 NGA 2000 6 - 45 Main Menu — Technical level configuration ↓ ↓ Listing of all modules ↓ ↓ In the menu "Listing of all modules" you will find the tags of all modules combined with the platform or the MLT analyzer.
6 - 46 NGA 2000 90003749(1) [ NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 1 NGA-2000 Software Manual Supplement: System Cal i bration Soft ware Version CLD 2.
System Calibration Supplement - 2 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 Table of Contents 1 Introduction __________________________________________________________________ 3 2 Valves for system calibration _____________________________________________________ 4 2.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 3 1 Introduction In opposite to the possibility to zero and span each analyzer module (channel).
System Calibration Supplement - 4 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 2 Valves for system calibration Before using the system calibration we have to put together the required valve pool. There are three types of I/O-modules supporting this: • DIO - 24 digital outputs / 8 digital inputs (max.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 5 It is also possible to configure the DIO-outputs via LON-Variables (i.
System Calibration Supplement - 6 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 3 Operating System Calibration Because there are many possibilities how to use SYSCAL it must be customized through the Expert Operator.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 7 3 .1 Set-up analyzer modules Before we can run any system calibration features we have to include the different analyzer modules( AM ) into system calibration by setup needed parameters.
System Calibration Supplement - 8 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 Example for an analyzer-system: V1 AM1 AM2 AM3 V2 AM4 V3 AM5 V 4 V 5 V 6 Calibration gas mi.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 9 3 .2 Programming Calibration Seque n ces Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to run the system calibration in an user definable order of up to 40 steps.
System Calibration Supplement - 10 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 For each step we have the choice to select a sp e cific AM or to activate the step for all enabled AMs.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 11 3 .3 Set-up general parameters One general parameter is "Calibration Procedures in Test-Mode". It is located in the menu "Calibration/Test procedures".
System Calibration Supplement - 12 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 A further general parameter is " Timeout for Gastest".
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 13 3 .4 Controlling System Calibration After setting-up there is the possibility to start 3 different modes: system zero-calibration : In this mode a zero-cal of all modules, which are enabled for SYSCAL, will be performed.
System Calibration Supplement - 14 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 3.4 .1 Control via menu system The possibility to control via menu is located in the menu page „Calibration/Test procedures“. There can be started and stopped any kind of system calibration respectively test procedure.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 15 The result of calibration for included AMs is displayed in an own menu page. This can be achieved by softkey „RESULTS“ from different menu pages of SYSCAL.
System Calibration Supplement - 16 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 Here we first choose the specific analyzer module and then we select at the parameter " Gastest of specific module" the required gas.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 17 3.4 .2 Control System Calibration by LON-Variables SYSCAL can also be controlled by LON-Variable CMFUNC. Setting this variable will cause the appropriate function (see following table).
System Calibration Supplement - 18 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 3.4 .3 Control System Calibration by Pr o grammable Input We can use the possibility to control SYSCAL by triggering with a programmable input.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 19 It is also possible to configure the DIO-inputs only per LON-Variables (i.e. SLTA-adapter or AK-protocol). Therefore is the LON-variable "DIOINP_ENTRYSIG" which corresponds to DIOINPSIGC and evaluates enum-values of STINAME.
System Calibration Supplement - 20 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 3.4 .4 Control System Calibration by AK-protocol command We can start and stop SYSCAL over serial interface by AK-protocol commands. Therefore the commands SCAL, STBY and ASTZ has to be used.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 21 3.4 .5 Time controlled System Calibration To activate a time controlled system calibration we have to set-up this in a own menu page.
System Calibration Supplement - 22 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 We can display the calculated ACT-times in menu page „next calibration events“... Menu1Line Menu2Line Menu3Line Expert controls and set up.
System Calibration 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98 NGA 2000 Supplement - 23 3 .5 Calibrate single Analyzers Each specific analyzer has still the possibility to start a calibration over other instances than SYSCAL. For this functionality we now have some additional considerations.
System Calibration Supplement - 24 NGA 2000 90003749(1) [ NGA-e ( CLD-Software 2.3/MLT-Software 3.1.X)] 04/98.
Een belangrijk punt na aankoop van elk apparaat Emerson NGA 2000 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Emerson NGA 2000 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Emerson NGA 2000 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Emerson NGA 2000 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Emerson NGA 2000 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Emerson NGA 2000 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Emerson NGA 2000 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Emerson NGA 2000 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.