Gebruiksaanwijzing /service van het product 1502A van de fabrikant Fluke
Ga naar pagina of 43
1502A Thermometer Readout User’ s Guide Rev. 722101 ENG Hart Scientific.
Limited W arranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Di vision ("Hart") is warranted to be free from de - fects in material and workmanship under normal use and s ervice. The warranty period is three years for the Thermometer Readout.
T able of Contents 1 Bef or e Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.
7 Digital Communications Interface . . . . . . . . . . . . . . . 23 7.1 Serial Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.1.1 Setting the Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.1.2 Automatic T ransmission of Measurements .
iii Figures Figure 1 Connecting a four-wire probe . . . . . . . . . . . . . . . . . . . . . . 12 Figure 2 1502A Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Figure 3 1502A Back Panel . . . . . . . . . . . . . . . . . . . . . .
iv T ables T able1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 T able 2 Matching Certif icate V alues to 1502A ITS-90 Coefficients . . . . . . . 18 T able 3 Setting Coef f icients Rtpw , a8, b8, a4, and b4 . . . . . . .
1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the symbol s that may be used on the i nstrument or in this manual and the meaning of each symbol.
Symbol Description On Canadian Standards Associati on O VERV OL T AGE (Installat ion) CA TEGORY II , P ollution Degree 2 per IEC1010-1 r e - f ers to the le vel of Impulse Wit hstand V oltage protection provided. Equipment of O VERV OL T AGE CA TEGOR Y II is energy-consuming equipment to be suppl ied from the fix ed install ation.
Before initial use, or after transport, or after s torage in humid or semi-hu mid en vironments, or anytime the instrument has not been ener gized for more than 10 days, the instrument needs to be energized for a "dry-out" period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1.
Keep the probe wires clean and aw ay from fluids. 1.3 A uthoriz ed Service Centers Please contact one of the follo wing authorized Service Centers to coordinate service on your Hart product: Fluke Corporation, Hart Scientific Di vision 799 E. Utah V alle y Driv e American Fork, UT 84003-9775 USA Phone: +1.
Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Of fice Service Center 60 Alexandra T errace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 T elefax: +65 6799-5588 E-mail: antng@singa.
2 Intr oduction The 1502A is a low-cost high-accurac y dig ital thermometer readout designed to be used with 25 Ω and 100 Ω R TDs an d SPR Ts. Its unique co mbination of fea - tures makes it suitable for a wide variety of applicatio ns from laborator y mea - surement to industrial processes.
3 Specifications and En vir onmental Conditions 3.1 Specifications Resistance Range 0 Ω to 400 Ω , auto-ranging Resistance Accuracy, one year 1 0 Ω to 20 Ω : 0.0005 Ω 20 Ω to 400 Ω : 0.0025% (25 ppm) of reading Resistance Accuracy, short term 1, 2 0 Ω to 30 Ω : 0.
Weight 2.2 lb. (1. 0 kg) Safety OVERVOLTAGE (Installation) CATEGORY II, Pollut ion Degree 2 per IEC 1010-1 1 Accuracy specificati ons apply within the r ecommended operating temper ature range. Accuracy l imits are increased by a factor of the temperature coefficient out side this range.
4 Quic k Start This section briefly explains the basics of setting up and operating your 1502A thermometer readout. 4.1 Unpac king Unpack the thermometer carefully and i nspect it for any damage that may ha ve occurred during shipment. If there i s shipping damage, notify t he carrier immediately .
to pins 4 and 5. (Pins 1 and 5 source current and pins 2 and 4 sense the poten - tial.) If a shie ld wire is present it should be connected to pi n 3. A two-wire probe can al so be used with the 1502A. It i s connected by attaching one wire to both pi ns 1 and 2 of the plug and the other wire to both pins 4 and 5.
5 P ar ts and Controls 5.1 Fr ont P anel Buttons See Figure 2. The front panel buttons are used to select units of measurement, access operat - ing parameters, and alter operating parameters. The function of each button is as follows: C/Probe —This button selects units of degrees Celsius.
5.2 Rear P anel See Figure 3 . Serial Port - The DB-9 connector i s for interfaci ng the thermometer to a com - puter or terminal with serial RS-232 communications. Probe Connector - At the rear of the thermometer is t he probe connector . The probe must be connected for operation.
6 General Operation This section e xplains basic operation of the 1502A Thermometer . 6.1 Selecting Units T emperature can be displayed in degrees Celsius (indicated with “ C ”), degrees Fahrenheit (indicated with “ F ”), or Kelvin (indicated with “ A ” for absolute).
1502A Thermometer Readout User’s Guide 16 Menu Sample Comm (Cal) Probe Set clock Set filter Set probe type Enter password Set time stamp Set current Set coefficients Set menu lockout Set baud rate S.
when all the digits are correct. If the pa ssw ord is entered correctly the f irst pa - rameter in the menu will appear . 6.4 Selecting the Probe Characterization Before the 1502A can measure temperature accurately it must know ho w to cal - culate temperature from the resistance of the sensor .
digit will flash. Use the U and D b uttons to change a d igit. Once th e sign and digits are correct, press Enter to accept the number . If you decide to cancel any changes you hav e made, you may do so by pressing the Ω / Exit button. This will immediately skip to the ne xt coef fi cient.
Example 1: A PR T was calibrated to ITS-90 and its calibration certificate states v alues for coeff icients Rtpw , a4, b4, a8, and b8. Set the 1502A parameters wi th v alues from the certificate as follows. Example 2: A PR T was calibrated to ITS-90 and its calibration certificate states v alues for coeff icients Rtpw , a5 and b5.
Example 3: A PR T was calibrated to ITS-90 and its calibration certificate states v alues for coeff icients R(273.16K), a6, b6, c6, and d. Set the 1502A parameters wit h val - ues from the certificate as follows: 6.
6.4.5 IPTS-68 Con version The IPTS-68 characterization con verts r esistance to temperature according to the International Practical T emperature Scale of 1968. The applicable coef fi - cients are R 0 (“ r0 ”), α (“ ALPHA ”), δ (“ dELtA ”), a 4 (“ A4 ”), and c 4 (“ C4 ”).
Compare this temperature with th e temperatures listed on the p robe calibration report to verify t hat the coef f icients you entered are correct. 6.5 Filtering While measuring temperature, the readings may appear to v ary . This may be due to actual v ariations in temperature or elect rical noise internal to the 1502A.
7 Digital Comm unications Interface Remote communications allows an e x ternal de vice, such as a computer , to com - municate with the 1502A to obtain measurement data and cont rol its operati on. Communication is accomplished with v arious commands issued to the 1502A through the RS-232 port or optional IEEE-488 port.
7.1.2 A utomatic T ransmission of Measurements The 1502A can be programmed to automati cally send measurements to a re - mote printer or termin al. The transmission interval is set using the “ SA PEr ” sample period parameter . This is set in the Comm menu after the baud rate pa - rameter .
to the time stamp parameter . The display will br iefly show “ ti Sta ” then the time stamp state which is either O N or OFF . Use the U and D buttons to change the state and p ress Enter . ON enables transmission o f the time stamp and OFF disables it.
7.2.1 Setting the Ad dress The IEEE-488 bus requires that each de vi ce has a unique address. The default address of the 1502A is 22 b ut can be changed if necessary . The IEEE-488 ad - dress of the 1502A is set within the Comm menu after t he serial linefeed pa - rameter .
27 7 Digital Communications Interface Remote Commands Command Description Measurement Commands T read measurement (includes label, unit , and time) F[ETCH?] r ead measurement value (SCPI compatible) S.
7.3.1.1 Reading T emperature The most recent temperature measurement can be read using the following command: T<EOS> reads the most recent measurement The syntax of the response is as follo ws: t:_nnnn.nnn_u or t:_nnnn.nnn_u_hh:mm:ss The _’ s represent space characters.
7.3.1.2 A utomatically T ransmitting Measurements By setting the sample period, the 1502A can be programmed to automatically transmit measurements from the RS-232 por t at specif ied intervals.
7.3.2.1 Selecting the Characterization The follo wing commands can be used to se lect the probe characterization and coeff icients: P=90<EOS> selects the ITS-90 characterization P=68<EOS> .
7.3.3.1 Setting the Filter The filter helps to reduce variations in the measuremen ts. The filter can be set remotely using the command: FI=<value><EOS> sets the f ilter time constant FI=0<EOS> disables the filter The value is the f ilter time constant in seconds.
LF=OF<EOS> disables linefeed 7.3.5 Calibration Commands The following commands are used in calibrating the instrument. 7.3.5.1 Entering the P ass wor d In order to set the calibration parameters the passw ord must be issued f irst.
ver. mmmmm,v .vv The m ’ s represent digits of the model number . The v ’ s represent the digit s of the firmw are versi on number . As an example, if the version number was 1.10 the response would be “v er .1502A,1.10". The following IEEE-488.
8 Calibration Pr ocedure The 1502A uses a three-point calibration scheme with a quadrati c polynomial correction function to maintain the accura c y of its resistance measurement. The three calibration poin ts are at 0 Ω , 100 Ω , and 400 Ω . Three calibration pa - rameters determine the correction function: CAL0, CAL100, and CAL400.
power . The *C0=<value><EOS>, *C1=<value> <EOS>, and *C4=<v alue><EOS> commands can be used to set the val ues of the CAL0, CAL100, and CAL400 parameters respecti vely . 8.2 Calibration Procedure Calibrati on requires four-wire 100 Ω and 400 Ω resist ors of 6.
9 Maintenance • The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance.
10 T r oubleshooting In case you run into dif ficulty while operat ing the 1502A, this section provi des some suggestions that may help you solve the probl em. Belo w are se veral sit ua - tions that may arise follo wed by possible causes of the problem and suggested actions you might take .
• Stem conduction error . The problem may be that the actual temperature of the sensor is not what you expect. This is often the result of stem con - duction where heat flowing through the stem of the probe to ambient af - fects the temperature of the probe.
10.1.1.1 Imm unity T esting The instrument was t ested to the requirements for industria l locations. This al - lo ws the instrument t o be used in all types of locations from the l aboratory to the factory floor . Criterion C was used fo r Electrostatic Dischar ge (ESD, IEC 61000-4-2) and Electric F ast T ransit (EFT , Burst, IEC 61000-4-4).
Een belangrijk punt na aankoop van elk apparaat Fluke 1502A (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Fluke 1502A heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Fluke 1502A vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Fluke 1502A leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Fluke 1502A krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Fluke 1502A bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Fluke 1502A kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Fluke 1502A . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.