Manual Supplement © 2008-2009 Fluke Corporation . All rights reserv ed. Printed in the U.S .A. Manual Title: Impulse 6000D/7000DP Users Supplement Issue: 3 Print Date: August 2007 Issue Date: 3/09 Revision/Date: 1, 1/08 Page Count: 10 This supplement contai ns information nece ssary to en sure the accuracy of the above manual.
Impulse 6000D/7000DP Users Manual Supplement 4/08 1 Change #1, 46470, 46697, 47714, 48007, 49680 Replace pages 31 through 35, with the following: Defibrillator Analyzer Specifications Energy Output Measurement Compatible Defibrillator Wa veshapes ....
Manual Supplement Impulse 6000D/700 0DP Users 2 8/08 Accuracy ............................................................. ± 0.1 ms Voltage Range ................................................................. 20 to 5000 V Accuracy ...............
Impulse 6000D/7000DP Users Manual Supplement 4/08 3 Scope Output Autorange ........................................................... 2000:1, 400:1 and 80:1: depe ndant on the range Waveform Playback Output ...........................................
Manual Supplement Impulse 6000D/700 0DP Users 4 3/09 Ventricular Fibrillation ......................................... Coarse and fine Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM Polymorphic Ventricular Tachycardia .....
Impulse 6000D/7000DP Users Manual Supplement 8/08 5 Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 6 4/08 Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 10 0, 150, and 200 Hz Pulse ................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Waveform .
Impulse 6000D/7000DP Users Manual Supplement 4/08 7 Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 8 4/08 Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only) , add the following section: SCOPE OUTPUT Each time a defibrillator is fired into.
Impulse 6000D/7000DP Users Manual Supplement 4/08 9 Pulse playback +4 V, 1 ms pulse Start of defibrillator pulse playback [1] [1] Pulse playback includes 0.4 ms before th e pulse was detected and theref ore displays any ear ly transitional signals. Use the synchronizing marker pulses to display specific information from the Scope Outpu t signal.
Manual Supplement Impulse 6000D/700 0DP Users 10 8/08 Change #3, 48038 On page 38, under Pacemaker Input, change: From: Accuracy …………..±1 % non-inductive (<2 μ H) To: Accuracy………….
Een belangrijk punt na aankoop van elk apparaat Fluke 6000D (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Fluke 6000D heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Fluke 6000D vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Fluke 6000D leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Fluke 6000D krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Fluke 6000D bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Fluke 6000D kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Fluke 6000D . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.