Victoreen ® Digital Preamplifier, Model 943-227-15,for use with Model 943-27 Current Mode Beta Detector Operator Manual February 2008 Manual No. 943-227-15-1, Rev.
Fluke Biomedical 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com.
Table of Contents Section 1: General Information ........................................................................................................... .. 1-1 1.1 General Description ...............................................................
(Blank page).
General Information General Description 1 1- 1 Section 1 General Information 1.1 General Description The Victoreen ® Model 94 3-227-15 Digital Preamplif ier, is designed to operate with the Model 943.
943-227-15 Digital Preamplifier Operator Manual 1- 2 Table 1-1. 943-227-15 Variations a nd Compatible Equipment Preamplifier Variations Application Detector Readout 943-227-15 Digital Preamplifier Tra.
General Information Application 1 1-3 Table 1-2. General Specifications for the 943-227-15 Digital Preamplifier Feature: Specification: Preamplifier Power + 15 v @ 20 mA Preamplifier Input DC current,.
943-227-15 Digital Preamplifier Operator Manual 1-4 Table 1-3. Preamplifier Serial Interface Board 977 -210 -10M Switch Settings Switch Normal Op. Switch Function SW1 Open Microprocessor Reset switch,.
General Information Specifications 1 1- 5 Table 1-6. Field Electrical Con nections, 943 -227-15 and 943-227-15VL 943-227-15 943-27-15VL TB1- Description Connects To 977-210-10M Description Connects To.
943-227-15 Digital Preamplifier Operator Manual 1-6 1.4 Auxiliary Equipment Model Description None 1.5 Recommended Spare Parts Table 1-9. Recommended Spare Parts List for the Model 943-227-15 Digital .
General Information Spare Parts/Receivi ng Inspection 1 1- 7 1.7 Storage Storage of Fluke Biomedi cal instruments must comply with Level B storage requirement s as outlined in ANSI N45.2.2 (1972) Section 6.1.2 (.2 ). The storage area shall comply with ANSI N45.
( Blank Page).
Theory of Operation Functional Description 2 2-1 Section 2 Theory of Operation 2.1 Functional Description The Model 943-227-15 Digital Preamplifi er is an eight- decade, linear reading radiation. The preamplifier is sensitive to moisture and is mounted in a NEMA 4 sealed enclo s ure.
943-227-15 Digital Preamplifier Operator Manual 2-2 INTEGRATING ELECTROMETER PGA ALARM OUTPUT DAC EPROM ANALOG SWITCH ADC ACIA HIGH VOLTAGE POWER SUPPLY RS232 VICO LOOP MICROPROCESSOR CONTROL SIGNAL HIGH VOLTAGE 0 - 10 Vdc 0/-12 Vdc MICROPROCESSOR CONTROL Resistor Resistor DETECTOR ASSEMBLY +VL +TX -TX +RX -RX -VL SIGNAL Figure 2-1.
Theory of Operation Main Circuit Boards 2 2-3 Table 2-2. Electrometer Control Bit Assignments Bit Position 7 6 5 4 3 2 1 0 Description X X X X X 0 0 0 Gain = 1 X X X X X 0 0 1 Gain = 2 X X X X X 0 1 0.
943-227-15 Digital Preamplifier Operator Manual 2-4 The following Test are prov ided on the 977-200-15M board: Table 2-4. 977-200-15M Test Points Test Point Function TP1 Electrometer Output TP2 Analog.
Theory of Operation Communications 2 2-5 Victoreen Communications Loop – 943-227-15VL Communication between t he detector preamplifier and the UDR or 960 controll er may be performed via Victoreen loop, a serial asynchronous communications interface.
943-227-15 Digital Preamplifier Operator Manual 2-6 Table 2-8. Jumper Positions for Pream plifier Serial Interface Board 977-210-10M Jumper Normal Setting Function Alternate Setting Function JMP1 A to B Enables Victoreen Loop (normal) B to C Enables RS232C JMP2 N/A Deleted, Not Used N/A Deleted, Not Used Table 2-9.
Operation Installation 3 3-1 Section 3 Operation 3.1 Installation Installation consists of mounting the equi pment, making the required electrical co nnections, and entering the desired set points. CAUTION Remove all power prior to installing the Preamplifier.
943-227-15 Digital Preamplifier Operator Manual 3-2 NOTE The electrometer circuitry in the preamplifier is highl y sensitive to moisture and physical damage. Use extreme care when the preamplifier case is opened as the high impedance electrometer circuitry is easily damaged.
943-227-15 Digital Preamplifier Operator Manual 3-3.
Operation Configuration 3 3-4 3.3 Operation Operation of the 943-227-15 Di gital Preamplifier is automatic and is controlled by the Mod el 942A-200C Digital Ratemeter or Model 9 60 Digital Process Radi ation monitoring system . Once the power to the controlling device is turned on, the prea mplifier begins operation.
943-227-15 Digital Preamplifier Operator Manual 3-5 For UDR systems, the address of the 942-200-80 Co mmuni cations board in the UDR must be set 404 0 and the baud rate must be set to 4800 baud.
Operation Configuration 3 3-6 3.5 Set Point Error Codes Failure of the 943-227-17 pre amplifier may result in the following error code s being displayed on th e controlling device: Code E0009 is normall y used to identify a Loop co mmunication failure.
(Blank page).
Maintenance, Calibration and Troubleshooting Maintenance 4 4-1 Section 4 Maintenance, Calibration and Troubleshooting 4.1 Maintenance The 943-227-15 Digital P reamplifier is designed to operate for extend ed periods of time with no scheduled maintenance requi red.
943-227-15 Digital Preamplifier Operator Manual 4-2 4.3 Troubleshooting WARNING Extreme care must be used when troubleshooting a system that has power applied. All standard troubleshooting precautions apply. WARNING Once a problem has been located, remove all power before continuing with the repair.
Appendix Connector Designations A A-1 Appendix A Connector Designations A.1 CONNECTOR DESIGNATIONS, 943-227-15 and 943-227-15VL Table A-1. Field Electrical Con nections, 943-2 27-15 and 943-227-15VL 9.
943-227-15 Digital Preamplifier Operator Manual A-2 A.2 SWITCH AND JUMPER POSITIONS, 943-227-15 and 943-227-15VL Table A-2. Preamplifier Serial Interfa ce Boa rd 97 7 -210-10M Switch Settings Switch Normal Op.
Appendix Applicable Drawings B B-1 Appendix B Applicable Drawings B.1 Applicable Drawings 943-227-15 and 943-227-15VL Drawing No . Description 943-227-15 Preamplifier Assembly, RS232 943-227-15VL Prea.
(Blank page).
Appendix Bill of Materials C C-1 Appendix C Bill of Materials C.1 Bill of Materials- 943-227-15 and 943-227-15VL Part Number Description 943-227-15 Preamplifier Assembly, RS232 943-227-15VL Preamplifi.
(Blank page).
Appendix Cable Termination Instructi ons D D-1 Appendix D Cable Termination Instructions D.1 Cable Termination Instructions The procedures and in structions provided in Ap pendix D are provided fo r reference in termin ating the field cables to the Digital Preamplifier.
943-227-15 Digital Preamplifier Operator Manual D-2 Detector Field Cable Preparati on, 943-227-15VL-M1, Conduit Hub: Prior to terminating the cable, the field con duit must be run to the junction box, and the field cable pull ed into the junction box.
Appendix Cable Termination Instructi ons D D-3 HV CONNECTOR NOTES (HV) The mating HV connector, P/N 30-92-1 (Kings 1 705-1 4), may now be terminated per the applicable Kings data sheet attached LOCATI.
943-227-15 Digital Preamplifier Operator Manual D-4.
Appendix Cable Termination Instructi ons D D-5.
943-227-15 Digital Preamplifier Operator Manual D-6.
Appendix Cable Termination Instructi ons D D-7.
943-227-15 Digital Preamplifier Operator Manual D-8.
Appendix Modifications E E-1 Appendix E Modification Sheets, Engineering Instructions E.1 Modification Sheets, Engineering Instructions P / N D e s c r i p t i o n 943-227-15VL-M1 Modification consist.
(Blank page).
(Blank page).
Fluke Biomedical 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com.
Een belangrijk punt na aankoop van elk apparaat Fluke 943-227-15 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Fluke 943-227-15 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Fluke 943-227-15 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Fluke 943-227-15 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Fluke 943-227-15 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Fluke 943-227-15 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Fluke 943-227-15 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Fluke 943-227-15 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.