USER’S MANUAL GENES I S 2002 TM GENES I S 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.
GENESIS 2006 SHOWN 2.
To assure proper record keeping, please retain this information for future use. DATE PURC HASED ____________________________ SERIAL # _____________________________________ DISTRIBUT OR NAME __________.
CONTENTS page 1. UNPACKING ........................................................ 5 2. TO THAW BLOOD PRODUCTS ........................ 5 3. TO REMOVE A THAWED BAG ........................ 9 4. TO MAINTAIN THE DEVICE IN THE READY TO THAW STATUS .....
1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006 unit. C o n t r o l U n i t W a t e r B a t h T h a w i n g B a g D i s p e n s e r Plasma Overwrap Thawing Bags (1000/Box) P.
indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEAN TM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prev ent the formation of turbidity and odor.
c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 ° C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other.
thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time.
3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder.
5. TO CLEAN THE BATH OR TO ADD WATER S A F E T Y : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL) IS TURNED ON. RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET. 5.1 TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF.
when filled with water. Do not drop the bath on a hard surface. GENESIS 2006 : Remove all, or part of the water. The bath weighs approximately thirty (30) pounds when filled with water. Carefully remove the water bath from the device. 5.6 DO NOT USE BLEACH OR HARSH CLEANERS.
6. CONTROL PANEL AND INDICATORS GENESIS 2002 TM CONTROL PANEL 12.
GENESIS 2006 TM CONTROL PANEL 13.
CONTROL KEYS POWER .......... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER SWITCH on the left panel is a master on/off switch to remove electrical power completely from all circuits in the unit. SET TEMP .
INDICATORS HEATING ....... This red indicator is lit when the heating element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off. FAULT ............ If the water temperature exceeds 58 ° C, an alarm will sound and the red fault light will be illuminated.
2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch.
7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrum ent. 8) HEATING light is flashing. • This is normal operation.
8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time setting for each bag location • Forced water circulat.
10. SPECIFICATIONS GENESIS 2002 TM GENESIS 2006 TM External Dim ension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” High Interior Bath Dimension 13.0” L x 7.0” Wide x 7.5” Deep 18.0” L x 8.0” Wide x 7.5” Deep Bath Volume 2.
11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com RETURN INFORMATION : Same as above.
Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006 was purchased S e r i a l # RGA # (Return Goods Authorization #) 6. Mark outside of box clearly, with your return address, and send to GenesisBPS, LLC .
GENESIS 2002 GENESIS 2006 PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006 , Plasma Thawing Water Bath System is warranted against defects in materials and workmanship for a period of one year from the date of purchase.
Een belangrijk punt na aankoop van elk apparaat Genesis Advanced Technologies 2002 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Genesis Advanced Technologies 2002 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Genesis Advanced Technologies 2002 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Genesis Advanced Technologies 2002 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Genesis Advanced Technologies 2002 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Genesis Advanced Technologies 2002 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Genesis Advanced Technologies 2002 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Genesis Advanced Technologies 2002 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.