Gebruiksaanwijzing /service van het product IH6300 van de fabrikant Invacare
Ga naar pagina of 32
Owner’s Operation and Maintenance Manual DEALER: This manual MUST be given to the user of the tub. USER: BEFORE using this tub, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.
Supine Bath Tub 2 Part No 1096926 WARNING DO NOT USE THIS PRODUCT OR AN Y AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READ ING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTION AL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MA NUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OP TIONAL EQUIPMENT.
TABLE OF CONTENTS Part No 1096926 3 Supine Bath Tub TABLE OF CONTENTS SPECIAL NOTES ................................................................................ 5 LABEL LOCATION .....................................................................
TABLE OF CONTENTS Supine Bath Tub 4 Part No 1096926 TABLE OF CONTENTS SECTION 5—MAINTE NANCE ........................................................... 25 Parts Information ................ ...................... ...................................
SPECIAL NOTES Part No 1096926 5 Supine Bath Tub SPECIAL NOTES Signal words are use d in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
LABEL LOCATION Supine Bath Tub 6 Part No 1096926 LABEL LOCATION ! W ARNING ELECTRICAL SHOCK HAZARD. All disassembly and maintenance of the Supine T ub MUST be done by a qualified technician, certified electrician or plumber. Disconnect electrical supply or turn OFF circuit breaker before performing an y maintenance to the Supine T ub.
TYPICAL PRODUCT PARAMETERS Part No 1096926 7 Supine Bath Tub TYPICAL PRODUCT PARAMETERS IH6300 FIXED HEIGHT IH6300ADH ADJUSTABLE HEIGHT Height Tub Rim 32 inches (813 mm) Raised N/A 41 inches (1041 mm .
FEATURES Supine Bath Tub 8 Part No 1096926 FEATURES CONTROL PANEL TUB INTERIOR Model IH6300ADH Adjustable Height Tub Tub Fill ON/OFF Disinfectant Flow Mete r Disinfectant Wand Disinfectant Cabinet Wat.
SECTION 1—GENERAL GUIDELINES Part No 1096926 9 Supine Bath Tub SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - GENERAL GUIDELIN ES contai ns important inform ation for the safe operation and use of this product. Check all parts for shipping damage before using .
SECTION 1—GENERAL GUIDELINES Supine Bath Tub 10 Part No 1096926 When operating wa t e r pressure exceeds 55 p.s.i., pressure regulators (similar to the ¾ ‐ inch W atts U5) MUST be installed.
SECTION 1—GENERAL GUIDELINES Part No 1096926 11 Supine Bath Tub The Supine Tu b is specifically designed to be used with a bath lift. ICCG recommends use of the T raverse Lift Model 1200 ONL Y .
SECTION 2—DISINFECTING THE TUB Supine Bath Tub 12 Part No 1096926 SECTION 2—DISINFECTING THE TUB WARNING Disinfect the tub after EACH use to avoid resi dent infectio n and contamination of the tub. DO NOT disinfect the tub when it is occupied otherwise injury may o ccur.
SECTION 3—OPERATION Part No 1096926 13 Supine Bath Tub SECTION 3—OPERATION Overview The tu b is designed to improve the efficiency and environmental safety of you r facility ʹ s operation.
SECTION 3—OPERATION Supine Bath Tub 14 Part No 1096926 Operating the Tub WARNING DO NOT stand in tub. Tub surfaces are slippery wh en wet. Serious injury may occur.
SECTION 3—OPERATION Part No 1096926 15 Supine Bath Tub Operating Shower/Massage Wand NOTE: STEPS 1 ‐ 3 are to be performed while holding the shower/massage wand over the tu b. Refer to Features on page 8 for additional illustrations.
SECTION 3—OPERATION Supine Bath Tub 16 Part No 1096926 Adjusting Tub Height - Model IH6300ADH ONLY WARNING DO NOT stand on tub frame. Frame is slip pery when wet. Foot injury is possible when adjusting tub height. CAUTION DO NOT adjust height more than three cycles (up and down) within a ten minute period.
SECTION 4—INSTALLATION Part No 1096926 17 Supine Bath Tub SECTION 4—INSTALLATION Preparing the Work Area NOTE: For this proced ure, ref er to FIGURE 4.1. 1. Determine location for the supine bath tub installation.
SECTION 4—INSTALLATION Supine Bath Tub 18 Part No 1096926 Securing the Tub NOTE: For this proced ure, ref er to FIGURE 4.1 on page 17 and FIGURE 4.
SECTION 4—INSTALLATION Part No 1096926 19 Supine Bath Tub Electrical Hook-up WARNING Be certain ALL electrical wo rk is performed by a qualif ied electrician and is in compliance with local electrical codes. Before installing and connec ting the tub, make certai n that the electric power source is SHUT OFF to avoid electrical shock.
SECTION 4—INSTALLATION Supine Bath Tub 20 Part No 1096926 FIGURE 4.3 Electrical Hook-up Electrical Hookup 115 V Electrical IH6300 Fixed Height Use 15 Amp GFCI Breaker Only IH6300ADH Adjustable Heigh.
SECTION 4—INSTALLATION Part No 1096926 21 Supine Bath Tub Electrical Installation DANGER The National Electric Code re quires the use of a ground faul t circuit interrupter on all therapeutic pools and tubs (h ydrotherapeutic tanks). Failure to follow this requirement could result in seri ous or fatal injury.
SECTION 4—INSTALLATION Supine Bath Tub 22 Part No 1096926 Plumbing Hook-Up WARNING Be certain ALL plumbing work is perf ormed by a qualified plumber and is in compliance with local codes. NOTE: For this proced ure, ref er to Refer to FIGURE 4.
SECTION 4—INSTALLATION Part No 1096926 23 Supine Bath Tub Plumbing Installation WARNING Be certain ALL plumbing work is performed by a qualified plumber.
SECTION 4—INSTALLATION Supine Bath Tub 24 Part No 1096926 FIGURE 4.6 Plumbing Installation and Schematic DETAIL “A” - PLUMBING INSTALLATION Cold Water Inlet Hot Water Inlet Shower/Massage Wand H.
SECTION 5—MAINTENANCE Part No 1096926 25 Supine Bath Tub SECTION 5—MAINTENANCE WARNING Before performing any maintenance to the tub, make certain that the electric power source is shut off to avoid electrical shock. Be certain ALL electrical wo rk is performed by a qualif ied electrician and is in compliance with local codes.
SECTION 5—MAINTENANCE Supine Bath Tub 26 Part No 1096926 Every Six Months ❑ Ve r i f y the wa t e r fl ow temperature with a separate thermometer . ❑ V isually inspect plumbing for leaks. ❑ Check that all labels are present and legible.
SECTION 5—MAINTENANCE Part No 1096926 27 Supine Bath Tub Heavy Duty Cleaning Procedure WARNING DO NOT use this product or any available option al equipment without first completely reading and u.
SECTION 5—MAINTENANCE Supine Bath Tub 28 Part No 1096926 Calibrating the Disinfectant Flowmeter WARNING DO NOT use this product or any available optional equipment without first completely readi.
SECTION 5—MAINTENANCE Part No 1096926 29 Supine Bath Tub 15. Adjust the injector by performing one of the foll owing: • Bottom of Floating Ba ll Reads Greater Than 35 ‐ Less disinfectant needs to be added to the mixture.
SECTION 5—MAINTENANCE Supine Bath Tub 30 Part No 1096926 Maintenance Record DATE MAINTENANCE PERSON TYPE OF MAINTENANCE PERFORMED NOTES.
SECTION 6—TROUBLESHOOTING Part No 1096926 31 Supine Bath Tub SECTION 6—TROUBLESHOOTING NOTE: If pr oblems are not rem ed i e d by the suggested means above, contact an Invacar e Dealer .
LIMITED WARRANTY PLEASE NOTE: THE WA RRANTY BELOW HAS BE EN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MA N UFACTURED AFTER JULY 4, 1975.
Een belangrijk punt na aankoop van elk apparaat Invacare IH6300 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Invacare IH6300 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Invacare IH6300 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Invacare IH6300 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Invacare IH6300 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Invacare IH6300 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Invacare IH6300 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Invacare IH6300 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.