Gebruiksaanwijzing /service van het product 16 series van de fabrikant Apex Digital
Ga naar pagina of 41
16 Series Mass and V olumetric Flow Meters Precision Gas Flow Meter Operating Manual.
10/24/06 Rev . 0 DOC-APEXMAN16 Notice: The manufacturer reserves the right to make any changes and improvements to the products described in this manual at any time and without notice.
T able of Contents Page Installation 5 Plumbing 5 Mounting 5 Application 5 Power and Signal Connections 6 RS-232 Digital Output Signal 7 Standard V oltage (0-5 Vdc) Output Signal 7 Optional 0-10 Vdc O.
T able of Contents Page Collecting Data 19 Data Format 19 Sending a Simple Script File to HyperT erminal® 20 Operating Principle 21 Gas Viscosity 21 Other Gases 22 V olume Flow vs.
5 Thank you for purchasing an Apex Gas Flow Meter . Please take the time to nd and read the information contained in this manual. This will help to ensure that you get the best possible service from your instrument.
6 Power and Signal Connections Power can be supplied to your M or V Series meter through either the power jack or the 8 pin Mini-DIN connector as shown in Figure 1. An AC to DC adapter which converts line AC power to DC voltage between 7 and 30 volts is required to use the power jack.
7 RS-232 Digital Output Signal If you will be using the RS-232 output signal, it is necessary to connect the RS-232 Output Signal (Pin 5), the RS-232 Input Signal (Pin 3), and Ground (Pin 8) to your computer serial port as shown in Figure 2.
8 7 4 6 2 1 D B -9 S e r ia l P o rt 5 5 - -- - -- - -- - G r o u n d - - -- - -- -- - -- - -- - -- -- - -- - -- - -- -- - -- - -- - G r o u n d - - -- - -- -- - 8 3 - -- - -- - -- - T ra n s m it -- .
9 An optional industrial connector is also available: 1 6 5 4 3 2 Pin Function Cable Color 1 Power In ( + ) Red 2 RS-232 Output Blue 3 RS-232 Input Signal White 4 Remote T are (Ground to T are) Green 5 Ground (commom for power , communications and signals) Black 6 Signal Out (V oltage or Current as ordered) Brown Figure 4.
10 M Series Mass Flow Meter Operation The M Series Mass Flow Meter provides a multitude of useful ow data in one simple, rugged device. The M Series can have several display “modes” depending on how the device is ordered.
1 1 Gas Absolute Pressure: The M Series ow meters utilize an absolute pressure sensor to measure the line pressure of the gas ow being monitored. This sensor references hard vacuum and accurately reads line pressure both above and below local atmospheric pressure.
12 Select Menu Mode Pushing “Mode” once will bring up the “Select Menu” display . Push the button nearest your selection to go to the corresponding display .
13 Communication Select Mode The Communication Select mode is accessed by pressing the button below “Comm. RS-232” on the Select Menu display . The screen will appear as shown in Figure 9 below . Select Main > Unit ID ( A ) .....A Baud ( 19200 ) .
14 Manufacturer Data “Manufacturer Data” is accessed by pressing the “Mfg. Data” button on the Select Menu display (Figure 7). The “Mfg 1” display shows the name and telephone number of the manufacturer .
15 V Series V olumetric Flow Meter Operation The V Series can have several display “modes” depending on how the device is ordered. All V Series meters have a default Main Mode,a Select Menu Mode, .
16 Flashing Error Message: Our volumetric ow meters and controllers display an error message (VOV = volumetric overrange) when a measured parameter exceeds the range of the sensors in the device. When any item ashes on the display , the ashing parameter is not accurate.
17 to the screen, hit the “Enter” key several times to clear any extraneous information. T ype “*@=@” followed by “Enter” (or using the RS-232 communcation select menu, select @ as identier and exit the screen). If data still does not appear , check all the connections and com port assignments.
18 Gas Select – The selected gas can be changed via RS-232 input. T o change the selected gas, enter the following commands: In Streaming Mode: $$#<Enter> In Polling Mode: Address$$#<Enter> (e.g. B$$#<Enter>) Where # is the number of the gas selected from the table below .
19 Collecting Data: The RS-232 output updates to the screen many times per second. V ery short-term events can be captured simply by disconnecting (there are two telephone symbol icons at the top of the HyperT erminal® screen for disconnecting and connecting) immediately after the event in question.
20 Sending a Simple Script File to HyperT erminal® It is sometimes desirable to capture data for an extended period of time. Standard streaming mode information is useful for short term events, however , when capturing data for an extended period of time, the amount of data and thus the le size can become too large very quickly .
21 Operating Principle All M or V Series Gas Flow Meters are based on the accurate measurement of volumetric ow . The volumetric ow rate is determined by creating a pressure drop across a unique internal restriction, known as a Laminar Flow Element (LFE), and measuring differential pressure across it.
22 Other Gases: M Series Flow Meters can easily be used to measure the ow rate of gases other than those listed as long as “non-corrosive” gas compatibility is observed. For example, a ow meter that has been set for air can be used to measure the ow of argon.
23 Once the corrected mass ow rate at standard conditions has been determined and the density at standard conditions is known (see the density table at the back of this manual), a true mass ow c.
24 This reduces to: P a V a / Z a T a = P s V s / Z s T s , eliminating R and n. Our mass ow meters model gas ows based upon the non-ideal gas characteristics of the calibrated gas. The ow corrections are normally made to 25 C and 14.696 PSIA and the compressibility factor of the gas under those conditions.
25 Gas Number Short Form Long Form Viscosity* 25 deg C 14.696 PSIA Density** 25 deg C 14.696 PSIA Compressibility 25 deg C 14.696 PSIA 0 Air Air 184.918 1.1840 0.9997 1 Ar Argon 225.593 1.6339 0.9994 2 CH4 Methane 1 1 1.852 0.6569 0.9982 3 CO Carbon Monoxide 176.
26 Gas Number Short Form Long Form Viscosity* 0 deg C 14.696 PSIA Density** 0 deg C 14.696 PSIA Compressibility 0 deg C 14.696 PSIA 0 Air Air 172.588 1.2927 0.9994 1 Ar Argon 209.566 1.7840 0.9991 2 CH4 Methane 103.657 0.7175 0.9976 3 CO Carbon Monoxide 165.
27 V olumetric Flow Meters Under Pressure V Series V olumetric Flow Meters are intended for use in low pressure applications. This is because an accurate measurement of the volumetric ow rate by means of dif ferential pressure requires the ow at the differential pressure sensor to be in a laminar state.
28 TROUBLESHOOTING Display does not come on or is weak. Check power and ground connections. Flow reading is approximately xed either near zero or near full scale regardless of actual line ow . Differential pressure sensor may be damaged. A void installations that can subject sensor to pressure drops in excess of 10 PSID.
29 Slower response than specied. M or V Series meters feature an RS-232 programmable Geometric Running Average (GRA). Depending on the full scale range of the meter , it may have the GRA set to enhance the stability/readability of the display , which would result in slower perceived response time.
30 Specication Mass Meter V olumetric Meter Description Accuracy ± 0.8% of Reading ±0.2% of Full Scale At calibration conditions after tare High Accuracy Option ± 0.4% of Reading ±0.2% of Full Scale At calibration conditions after tare Repeatability ± 0.
31 Specication Mass Meter V olumetric Meter Description Accuracy ± 0.8% of Reading ±0.2% of Full Scale At calibration conditions after tare High Accuracy Option ± 0.4% of Reading ±0.2% of Full Scale At calibration conditions after tare Repeatability ± 0.
32 T echnical Data for Moderate Flow Mass & V olumetric Flow Meters >20SLPM Full Scale up to 0-250SLPM Full Scale Specication Mass Meter V olumetric Meter Description Accuracy ± 0.8% of Reading ±0.2% of Full Scale At calibration conditions after tare High Accuracy Option ± 0.
33 Specication Mass Meter V olumetric Meter Description Accuracy ± 0.8% of Reading ±0.2% of Full Scale At calibration conditions after tare High Accuracy Option ± 0.4% of Reading ±0.2% of Full Scale At calibration conditions after tare Repeatability ± 0.
34.
35.
36.
37 Option: T otalizing Mode M or V Series Flow Meters and Controllers can be purchased with the T otalizing Mode option. This option adds an additional mode screen that displays the total ow (normally in the units of the main ow screen) that has passed through the meter or controller since the last time the totalizer was cleared.
38 Option: 9 V olt Battery Pack A Battery Pack that uses a common 9 V olt battery can be mounted to the top of your M or V Series Flow Meter . Power is passed from the battery to the ow meter through the 8 pin Mini-DIN connector .
39 Wrench Access Hole Wrench Access Hole 3/32 Hex Wrench T rapped Mounting Screw T rapped Mounting Screw 8 Pin Mini DIN Plug 9 volt battery Back of Meter Battery Pack Installation/Removal Battery Pack Installation and Removal: The battery must be removed before the battery pack can be installed or removed.
40 CCM CCH LPM LPH CFM CFH CFH 0.0021 0.00003 2.1 189 0.035 60.0 1.0 CFM 0.000035 0.0000005 0.035 0.00059 1.0 0.0166 LPH 0.06 0.001 60.0 1.0 1699.0 28.316 LPM 0.001 0.000017 1.0 0.0166 28.316 0.4719 CCH 60.0 1.0 60000.0 1000.0 169901 1.0 28317.0 CCM 1.
41 Serial Number: ____________________________ Model Number: ____________________________ Calibration Certicate (Store device calibration certicate in the pocket below .
Een belangrijk punt na aankoop van elk apparaat Apex Digital 16 series (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Apex Digital 16 series heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Apex Digital 16 series vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Apex Digital 16 series leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Apex Digital 16 series krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Apex Digital 16 series bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Apex Digital 16 series kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Apex Digital 16 series . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.