Gebruiksaanwijzing /service van het product 9211A van de fabrikant National Instruments
Ga naar pagina of 19
USER GUIDE AND SPECIFICA TIONS USB-9211/9211A 4-Channel, 24-Bit Thermocouple Input Devices This user guide describes how to use the National Instruments USB-9211/9211A devices and lists the specifications.
USB-9211/9211A User Guid e and Specifications 2 ni.com Dimensions Figure 2 shows the USB-9211/9211A devi ce dimensions. Figure 2. USB-9211/9211A Series Devices in Inches (Millimeters) Safety Guidelines Operate the USB-9211/9211A only as described in these operating instructions.
© National Instruments Cor poration 3 USB-9211/9211A User Guide and Specifications Safety Guidelines for Hazardous Vo ltages If hazardous voltages are connected to the module, take the following precautions. A hazardous vo ltage is a voltage greater than 42.
USB-9211/9211A User Guid e and Specifications 4 ni.com Related Documentation If your device uses NI-D A Qmx Base, refer to ni.com to do wnload the software. Installing the USB-9211/9211A Device Installing the Software Before installing the device, you must install the software you plan to use with the device.
© National Instruments Cor poration 5 USB-9211/9211A User Guide and Specifications 3. Align the I/O module with the carrier, as sho wn in Figure 3. Figure 3. Module Installation 4. Squeeze the latches and insert the NI 9211 module into the carrier . 5.
USB-9211/9211A User Guid e and Specifications 6 ni.com Mounting the USB-9211 /9211A to a Panel Threaded inserts are located in the USB-921 1/9211A for mounting it to a panel.
© National Instruments Cor poration 7 USB-9211/9211A User Guide and Specifications Wiring the USB-9211/9211A The USB-9211/9211 A has a 10-terminal, detachabl e high voltage screw terminal enclosure that provides connections for four therm ocouple input channels.
USB-9211/9211A User Guid e and Specifications 8 ni.com Refer to T able 3 for the termin al assignments for each channel. Assembling the High V oltage Screw T erminal Backshell The high voltage screw terminal back shell must be instal led when using hazardous voltages (>42.
© National Instruments Cor poration 9 USB-9211/9211A User Guide and Specifications Figure 7. High V oltage Screw T erminal Backsh ell USB-9211/9211A Circuitr y The USB-9211/9211A channels share a common gro und that is isolated from the chassis and the host comput er.
USB-9211/9211A User Guide and Specifications 10 ni.com Figure 8. Input Cir cuitry for One Channel If thermocouples are connected to the USB-921 1/9211A, the gain and of fset errors resulting from the so ur ce impedance of the thermocouples are negligible for most applications.
© National Instruments Corporation 11 USB- 9211/9211A User Gu ide and Specifications Measurement Accuracy fo r the Different T ypes of Thermocouples Figures 9, 10, 11, 12, and 13 show the typical and maximum errors for the different thermocouple types when used with the USB-9211/9211A over the full temperature range.
USB-9211/9211A User Guide and Specifications 12 ni.com Figure 10. T ype E and T Errors Figure 11. T ype J and N Errors –200 0 200 400 600 800 1000 0 2 3 4 Max ov er T emp Range with A utozero Max at.
© National Instruments Corporation 13 USB- 9211/9211A User Gu ide and Specifications Figur e 12. T ype K Errors Figure 13. T ype R and S Errors Cold-Junction T emperatur e Measurement Accuracy Heat f.
USB-9211/9211A User Guide and Specifications 14 ni.com Minimizing Thermal Gradients Thermocouple wire can be a significan t source of thermal gradients if it conducts heat or cold directly to terminal j unctions. To min imize these errors, follow these guidelines: • Use small-gauge thermocouple wire.
© National Instruments Corporation 15 USB- 9211/9211A User Gu ide and Specifications Cold-junction compensation sensor accuracy 0 to 60 °C ...................... .................. 0.6 °C (1.1 °F) typ, 1.3 °C (2.3 °F) max Conversion time .......
USB-9211/9211A User Guide and Specifications 16 ni.com Offset error from source impedance ...... . 0.05 µ Vt y p ,0 . 0 7 µ Vm a x p e r Ω source impedance due to input current Power Requirements Current consumption from USB .............500 mA, max Suspend mode .
© National Instruments Corporation 17 USB- 9211/9211A User Gu ide and Specifications V oltages Connect only voltages that are within these limits. Channel-to-COM .
USB-9211/9211A User Guide and Specifications 18 ni.com Maximum altitude ................. ..................2,000 m (at 25 °C ambient temperature) Pollution Degree (IEC 60664) .... ............2 Electromagnetic Compatibility Emissions ...............
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for m ore inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are t rademarks or trade names of their respective companie s.
Een belangrijk punt na aankoop van elk apparaat National Instruments 9211A (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen National Instruments 9211A heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens National Instruments 9211A vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding National Instruments 9211A leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over National Instruments 9211A krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van National Instruments 9211A bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de National Instruments 9211A kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met National Instruments 9211A . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.