User Guide
USER
EUR 1
6 1 8 3 6
Global leaders in sleep and respiratory medicine www.resmed.com

ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 65 Milton Park Abingdon Oxfordshire OX14 4RX UK ResMed Offices Australia, Austria, Brazil, Finland, France, Germany, Hong Kong, Japan, Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Mirage Quattro Full Face Mask Protected by patents: AU 777033, EP 1187649, EP 1314445, HK 1057714, US 6412487, US 6439230, US 6513526, US 6581594, US 6701927, US 6796308, US 6823865, US 6823869, US 6860269, US 7021311, US 7069933. Other patents pending.
Protected by design registrations: AU 309495, AU 309789, AU 309800, AU 310242, AU 311114, AU 311132, AU 311133, AU 311134, AU 311135, AU 311136, AU 311137, AU 311186, AU 311187. Others pending.
Mirage and Mirage Quattro are trademarks of ResMed Ltd and Mirage is registered in U.S. Patent and Trademark Office.
Velcro is a registered trademark of Velcro Industries B.V. STERRAD and CIDEX are trademarks of Johnson & Johnson. Alconox is a trademark of Alconox Ltd. Neodisher Medizyme is a trademark of Chemische Fabrik Dr Weigert. Aniosyme is a trademark of Laboratoires Anios.
© 2007 ResMed Ltd.

A Mask parts / Maskenteile / Composants du masque / Parti della maschera / Piezas de la mascarilla / Peças da máscara / Maskeronderdelen
Fitting the mask / Anlegen der Maske / Mise en place du masque / Come indossare la maschera / Colocación de la mascarilla / Colocação da máscara / Masker opzetten en aanpassen
© 2007 ResMed Ltd. © 2007 ResMed Ltd.
1 2 3 4 5 6 7 8
Cushion clip / Maskenkissenclip / Clip de la bulle / Fermaglio del cuscinetto / Traba de la almohadilla /
3 6 9 12 15 18 21 24
Clipe da almofada / Kussenklem
Mask cushion / Maskenkissen / Bulle du masque / Cuscinetto della maschera / Almohadilla de la mascarilla / Almofada da máscara / Maskerkussentje
Mask frame / Maskenrahmen / Entourage rigide / Telaio della maschera / Armazón de la mascarilla / Armação da máscara / Maskerframe
Disassembling for cleaning / Auseinanderbau zu Reinigungszwecken / Démontage pour le nettoyage / Smontaggio per la pulizia / Desmontaje para la limpieza / Desmontagem para limpeza / Demontage voor reiniging
Elbow assembly / Kniestück / Coude / Complesso del gomito / Unidad del codo / Conjunto do cotovelo / Complete bocht
2 3 4 5 6
Valve clip / Ventilclip / Clip de la valve / Fermaglio della valvola / Traba de la válvula / Clipe da válvula / Klepklem
Valve / Ventil / Valve / Valvola / Válvula / Válvula / Klep
Valve hinge / Ventilgelenk / Charnière de la valve / Cardine della valvola / Bisagra de la válvula / Articulação da válvula / Klepscharnier
Elbow / Kniestück / Coude / Gomito / Codo / Cotovelo / Bocht
Valve membrane / Ventilmembran / Membrane de la valve / Membrana della valvola / Membrana de la válvula / Membrana da válvula / Klepmembraan
Swivel / Drehadapter / Pièce pivotante / Giunto rotante / Pieza giratoria / Peça giratória / Draaibare kop
8 9 10 11
© 2007 ResMed Ltd.

Reassembling the mask / Wiederzusammenbau der Maske / Remontage du masque / Riassemblaggio della maschera / Para volver a montar la mascarilla / Montagem da máscara / Compleet masker in elkaar zetten
Elbow Reassembly / Wiederzusammenbau des Kniestücks / Remontage du coude / Riassemblaggio del gomito / Volver a montar el codo / Montagem do cotovelo / Bocht in elkaar zetten
1 2 3 4 5
Full Face Mask / Full Face Maske / Masque facial / Maschera ono-nasale / Mascarilla Facial / Máscara Facial / Volgelaatsmasker
Illustrations Sheet / Illustrationsblatt / Fiche d'illustrations / Foglio delle illustrazioni / Hoja de figuras / Folheto de ilustrações / Illustratieblad
Note: For complete instructions, please read the text section in conjunction with the illustrations on this sheet. Hinweis: Vollständige Anweisungen finden Sie im Textabschnitt und in den Illustrationen auf diesem Blatt. Remarque : pour les instructions complètes, veuillez lire le texte du guide avec cette fiche d’illustrations. Nota: Per istruzioni complete, leggere la sezione con il testo insieme alle illustrazioni su questo foglio. Nota: para tener instrucciones completas, lea el texto en conjunto con las figuras de esta hoja. Nota: para obter instruções completas, leia a secção de texto acompanhando com as ilustrações neste folheto. Let op: Raadpleeg voor de volledige instructies s.v.p. de desbetreffende tekst samen met de illustraties op dit blad.
Mask Reassembly / Wiederzusammenbau der Maske / Remontage du masque / Riassemblaggio della maschera / Volver a montar la mascarilla / Montagem da máscara / Masker in elkaar zetten
6 7 8 9
10 11 12 13
© 2007 ResMed Ltd.
3 6 9 12 15 18 21 24

Thank you for choosing ResMed’s Mirage Quattro Full Face Mask. The Mirage Quattro is a full face mask that covers your nose and mouth. This means that you can receive effective therapy even if you breathe through your mouth. The mask incorporates vent holes and a built-in valve so that you can continue to breathe fresh air if the airflow to your mask is impeded for any reason. Intended Use
The Mirage Quattro channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Quattro is to be used by adult patients (>30 kg) for whom positive airway pressure has been prescribed. The Mirage Quattro is intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment. Before Using the Mask

Select ‘Mir Full’ or ‘Full Face’ as the mask option when using the Mirage Quattro with ResMed flow generators that have mask setting options. Fitting the Mask The diagrams in section A on the illustrations sheet show how the parts of the mask fit together. The diagrams in section B show the sequence for obtaining the best fit and comfort in using your mask. Please follow these instructions carefully. 1 Twist the forehead support dial to Position 24, where the forehead support is fully extended (Fig. B-1). 2 Hold the mask steadily on your face (Fig. B-2) and pull the headgear over your head. 3 Bringing the lower straps below your ears, clip the headgear to the lower part of the mask frame (Fig. B-3).
4 Unfasten the Velcro®, pull the upper headgear straps evenly until they are comfortable and firm, and reattach the Velcro (Fig. B-4). 5 Repeat step 4 with the lower headgear straps (Fig. B-5). 6 Test the functioning of the valve by sealing the swivel with your hand and ensuring that you can still breathe freely. 7 Connect the main air tubing to the swivel (Fig. B-6). Then connect the free end of the air tubing to your flow generator and turn the flow generator on.

8 Lie down in your sleeping position. Twist the forehead support dial (Fig. B-7) until you feel that the mask is evenly positioned on your face. It should be comfortable, with a satisfactory seal. Gently tighten the lower headgear straps if necessary. 9 The fitted mask is shown in Fig. B-8. Optimising Mask Seal and Comfort To receive effective therapy it is important to have the best seal possible between your mask and your face. If you are experiencing leaks, please ensure that you have followed the fitting instructions carefully. Carry out a quick check after fitting your mask:
If you cannot resolve any problems, look at the Troubleshooting section, or talk to your clinician. You may require another mask size. Removing the Mask
To remove the Mirage Quattro, disconnect one of the lower headgear clips and pull the mask and headgear up over your head. Disassembling the Mask
See section C on the illustrations sheet. 1 Detach the air tubing. 2 Unclip the lower headgear clips (Fig. C-1) and unhook the upper headgear straps (Fig. C-2). It is not necessary to remove the headgear clips from the headgear. 3 Remove the elbow assembly from the mask frame (Fig. C-3). It’s easier to pivot the bottom of the elbow away from the mask frame than to pull the elbow straight off. 4 Remove the valve clip from the elbow (Fig. C-4). 5 Peel the valve away from the valve clip (Fig. C-5). Do not pull on the membrane to remove the valve. 6 Detach the swivel from the elbow (Fig. C-6). 7 Note the position on the dial for the next time you use the mask.

8 Fully unscrew and remove the dial then pull the forehead support out of the mask frame (Fig. C-7). 9 Remove the forehead support pads from the forehead support (Fig. C-8). 10 Squeeze and push the upper side tabs on the cushion clip out of the mask frame. Pivot away from the mask, and gently pull out (Fig. C-9). 11 Separate the cushion from the cushion clip (Fig. C-10). 12 Remove the ports cap from the mask frame (Fig. C-11). Cleaning the Mask in the Home

The Mirage Quattro should be disinfected or sterilised when used between patients. All components, except the headgear, are validated for up to 20 disinfection or sterilisation cycles using one of the following processes:
The headgear should be thoroughly washed between patients, as described in “Cleaning the Mask in the Home”. Detailed disinfection and sterilisation instructions are available from the ResMed website, www.resmed.com/masks/sterilization and must be consulted before multipatient use. Reassembling the Mask
See section D on the illustrations sheet.
Before reassembling the mask, check that all components are clean and dry. Check that the valve is not damaged, distorted or torn, and that none of the other components are broken or cracked. Replace any parts that are damaged. Flexible PVC contains elements that can be detrimental to the materials of the mask. If flexible PVC tubing is used continuously, the attachment port should be inspected for cracking. Replace any parts that are damaged. Elbow Reassembly 1 Insert the valve into the valve clip (Fig. D-1). 2 Push the valve clip onto the elbow (Fig. D-2). Make sure both sides click into place (Fig. D-3). 3 Insert the swivel into the end of the elbow (Fig. D-4). 4 Insert the assembled elbow (Fig. D-5) into the mask frame. Mask Reassembly 1 Fit the cushion clip into the grooved edge of the mask cushion. Fit the top section first, then work your way around, pushing the clip into the cushion (Fig. D-6). Ensure that the cushion is not twisted or distorted around the clip. 2 Align the cushion clip with the mask frame. Push the cushion clip into the mask frame, ensuring that all three clips click into place (Fig. D-7). 3 Push the forehead pads firmly into the forehead support (Fig. D-8).

4 Insert the forehead support into the mask frame and screw the dial clockwise onto the pole (Fig. D-9). Turn the dial until it clicks (Fig. D-10). Ensure the dial sits flush against the frame. 5 Insert the ports cap (Fig. D-11). 6 Reattach the top headgear by threading it through the loop (Fig. D-12). 7 Turn the dial until you reach the setting that gives you good comfort and seal (Fig. D-13). Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time. Store the mask in a cool dry place out of direct sunlight. Disposal
The Mirage Quattro does not contain any hazardous substances and may be disposed of with your normal household refuse. Troubleshooting
Problem Possible reason Possible solution
Mask won’t seal properly or is uncomfortable
Mask cushion may be dirty. Clean cushion according to instructions in “Cleaning the Mask in the Home”.
Mask may have been fitted incorrectly or is incorrectly adjusted.
Carefully follow instructions in “Fitting the Mask”. Check headgear and dial are not over- tightened. Extend dial to apply more force to chin. Contract dial to increase force at nose bridge.
Mask cushion is skewed in cushion clip.
Check insertion of cushion and reinsert correctly.
Mask cushion is misaligned or creased on face.
Ensure cushion is properly aligned and that there are no creases in it: undo headgear clip, check cushion and reposition the mask on your face.
Mask is wrong size. Talk to your clinician.
Mask is too noisy Vents are blocked or partially blocked.
Clean vents according to instructions in “Cleaning the Mask in the Home”.
Valve is misassembled. Reassemble valve.
Cushion is not fully clipped in. Check insertion of cushion.
Mask is assembled incorrectly. Reassemble mask.
Valve is torn or distorted and is not sealing.
Replace valve.
Dial inserted into mask frame before forehead support
Misassembly. Push dial base onto a hard surface until the dial comes out of the frame.


Pressure (cm H2O)
Flow (± 6.0 L/min)
Vent Flow Rate (L/min)
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Dead Space Information Dead space: the empty volume of the mask up to the swivel. The dead space of the mask varies according to cushion size. It is 242 mL for the large size.
Standards CE designation in accordance with EC directive 93/42/EEC, class IIa. ISO 14971, ISO 13485, ISO 10993-1 and ISO 17510-2.
Therapy Pressure 4 to 20 cm H2O
Resistance Drop in pressure measured (nominal) at 50 L/min: 0.5 cm H2O at 100 L/min: 1.6 cm H2O
Sound The A-weighted sound power of the mask is 28 dBA.
Environmental Conditions Operating temperature: +5°C to +40°C Operating humidity: 10%–95% relative humidity non-condensing Storage and transport: -20°C to +60°C Storage and transport humidity: 10%–95% relative humidity non-condensing
Pressure–Flow Curve

Mask sizes (Mask fully assembled with elbow assembly - no headgear)
size height (mm) width (mm) depth (mm)
extra small 176 109 108
small 186 109 108
medium 196 109 108
large 206 109 108
REF
User Guide
Indicates a warning or caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device.

Ref. Item
61204 Mirage Quattro Full Face Mask, Extra Small - Europe 1
61205 Mirage Quattro Full Face Mask, Small - Europe 1
61206 Mirage Quattro Full Face Mask, Medium - Europe 1
61207 Mirage Quattro Full Face Mask, Large - Europe 1
61220 Mirage Quattro Full Face Mask, Extra Small - France
61221 Mirage Quattro Full Face Mask, Small - France
61222 Mirage Quattro Full Face Mask, Medium - France
61223 Mirage Quattro Full Face Mask, Large - France
61224 Mirage Quattro Full Face Mask, Extra Small - UK
61225 Mirage Quattro Full Face Mask, Small - UK
61226 Mirage Quattro Full Face Mask, Medium - UK
61227 Mirage Quattro Full Face Mask, Large - UK
61270 Mask frame, Extra Small (1 pk)
61271 Mask frame, Small (1 pk)
61272 Mask frame, Medium (1 pk)
61273 Mask frame, Large (1 pk)
61274 Cushion clip, Extra Small (1 pk)
61275 Cushion clip, Small (1 pk)
61276 Cushion clip, Medium (1 pk)
61277 Cushion clip, Large (1 pk)
61282 Elbow (1 pk)
61286 Valve (AAV) + clip (1 pk)
61287 Valve (AAV) + clip (10 pk)
61288 Forehead support (1 pk)
61289 Dial (1 pk)
61290 Cushion + clip, Extra Small (1 pk)
61291 Cushion + clip, Small (1 pk)
61292 Cushion + clip, Medium (1 pk)
61293 Cushion + clip, Large (1 pk)
61294 Elbow assembly (elbow, valve, clip and swivel) (1 pk)
61295 Ports cap (2 pk)
61296 Ports cap (10 pk)
16565 Swivel (1 pk)
16118 Headgear, Small (1 pk)
16117 Headgear, Standard (1 pk)
Complete System · Replacement Parts

Ref. Item
16119 Headgear, Large (1 pk)
60115 Headgear clip (2 pk)
60116 Headgear clip, (10 pk)
60123 Forehead support pad (1 pk)
60124 Forehead support pad (10 pk)
ResMed acknowledges all consumer rights granted under the EU Directive 1999/44/EG and the respective national laws within the EU for products sold within the European Union.
Pagina's waarnaar wordt verwezen in deze handleiding — fabrikant, ondersteuning en downloads.
Eigenaren en ons team reageren doorgaans binnen een dag — en uw vraag helpt ook de volgende persoon.
Stel als eerste een vraag — andere eigenaren of ons team geven antwoord.