Gebruiksaanwijzing /service van het product BC-41 van de fabrikant Tanita
Ga naar pagina of 23
Please read this Instruction Manual carefully and keep it handy for future reference. BOD Y COMPOSITION ANAL YZER BC-418 INSTRUCTION MANUAL.
3 T able of Contents 2 Table of Contents •••••••••••••••••••••••••••••••••••••••••••••••••••.
W ARNING • Individuals with a Pacemaker or Other Internal Medical Devices This equipment sends a weak electrical current through the body during measurement.
6 1.Product Assembly and Components Printer Paper Paper Width : 58mm Paper Roll Diameter : Maximum 55mm Product Str ucture Please verify that the following components are in the package.
8 Membrane Switch R R L L %Body Fat Range : 1%~75% %Body Fat Increments : 0.1% Max 200kg / 440lb e=d= 0.1kg / 0.2lb Min 2kg / 4lb T= -200kg /-440lb Operation Temperature Range 0 / +35 C Feed Key Advances the print paper. ON/OFF Key Turns the power on or off.
11 Connection and Installation 3.Preparation for Use For accurate measurement, place the Weighing Platform on as flat a surface as possible and adjust the level with the adjustable feet so the bubble in the level gauge is in the center of the frame.
12 13 Setting the date and time 4.Setting Date and Time AGE % FAT ATHLETIC STEP O N FEMALE MALE kg lb ft.in cm CLOTHES WEIGHT AGE % FAT FEMALE MALE ATHLETIC STEP O N TIME SET kg lb ft.in cm CLOTHES WEIGHT AGE % FAT ATHLETIC STEP O N ~ FEMALE MALE kg lb ft.
14 15 The equipment will start up with these settings next time you use it. This process is utilized to select whether or not you use the target body fat ratio function. (See the printout sample on page 16.) AGE % FAT ATHLETIC STEP O N FEMALE MALE kg lb ft.
16 17 ON [1] Goal Setter Mode Print Out STANDARD Input ATHLETIC STANDARD ATHLETIC OFF [0] Short [0] Printing selection Long [1] TARGET BF 00% ON [1] STANDARD ATHLETIC STANDARD ATHLETIC OFF [0] TARGET BF 00% <Printing selection and Goal Setter Mode> For this Printing Selection, please refer to the page 13.
AGE % FAT FEMALE MALE ATHLETIC STEP O N lb kg f t.in cm AGE % FAT FEMALE MALE CLOTHES WEIGHT lb kg f t.in cm CLOTHES WEIGHT ATHLETIC STEP O N 19 Posterior Electrodes Anterior Electrodes AGE % FAT ATHLETIC STEP O N ~ FEMALE MALE kg lb ft.in cm CLOTHES WEIGHT 1.
1. After turning on the unit, press the [ ] key . After a momentary display check, “0.0” will appear on the LCD. If measuring units need to be changed, do so at this time by pressing the [ ]key. An arrow on the LCD will follow the selection of weighing units.
20-39 40-59 60-79 20-39 40-59 60-79 0% 10% 20% 30% 40% 23 22 Explanation of the Print Out BODY TYPE GENDER AGE HEIGHT WEIGHT BMI BMR FAT% FAT MASS FFM TBW DESIRABLE RANGE FAT% FAT MASS TARGET BF% is : Predicted weight : Predicted fat mass : FAT TO GAIN: STANDARD MALE 3 4 179 cm 73.
CAUTION Please pay careful attention to avoid injury from the sharp edge. Please turn off the machine before clearing Paper jams. 1. Remove the Paper Dispenser Cover . If you lift up the Paper Dispenser Cover from the back of the Control Box, it can be easily removed.
26 If problems occur, please refer to the following instructions before you ask for repairs. Problem List 8.T roubleshooting Problem Solution Nothing appears on the display after pushing [ ] key. ON OFF • Please confirm that the Control Box is properly connected to the AC Adapter and the AC Adapter is properly connected to the cable.
• Transmission data will be output after measurements regardless of the status of the personal computer or other device at the receiving end. For that reason, the personal computer or other device to which data will be transmitted must be put in a reception-ready status before measurement.
D? : Confirm current setting D0 : Set clothing weight (5-byte fixed length) D1 : Set gender male = 1 female = 2 D2 : Set body type standard = 0 athlete = 2 D3 : Set height (5-byte fixed length) D4 : S.
Host . / body composition analyzer *Gender, body type, height and age must be entered. *Once all the necessary items of data have been entered, the unit will determine on the basis of age and gender whether to switch to Athlete or Standard mode. Hence even if you choose Athlete mode, the unit may switch to Standard mode.
Reset settings Host . / body composition analyzer If commands accepted : Host , . body composition analyzer If settings incorrect : Host , . body composition analyzer LF CR ! LF CR CHR$(01FH) LF CR CHR$(01FH) Resetting Cancel measurement Host . / body composition analyzer If commands accepted : Host , .
36 Reliability of body composition measurements with an 8-electrode BIA Introduction This device calculates body fat percentage, fat mass, fat fr ee mass, and predicted muscle mass on the basis of data obtained by Dual Energy X-ray Absorptiometry (DXA) using Bioelectrical Impedance Analysis (BIA).
39 Intra-day changes are thought to occur in unique cycles arising from dif ferences in the r hythm of individual daily routines, occupations, and day-to-day activities.
2) Inter-day changes The diagrams below offer examples of actual measur ements made of inter-day changes. A study was done to determine the degree of change in the impedance between the feet during dehydration; the first two days repr esent a normal daily routine, while in the latter two days a state of dehydration was induced using a sauna.
43 Product specifications Output Items Measurement System Measurement Frequency Impedance Measurement Weight Measurement Input Items 8 polar Bioelectrical Impedance Analysis 50kHz Measurement Current 500µA Electrode Material Pressure Contact Stainless Steel Foot Pads, The Hand Grips are Metal-Plated to ABS Resin.
BC4187001(3) This device features radio interference suppression in compliance with valid EC Regulation 89/336/EEC. 14-2,1-chome,Maeno-cho,Itabashi-ku Tokyo,Japan Tel:(03)3968-2123 / (03)3968-7048 Fax.
Een belangrijk punt na aankoop van elk apparaat Tanita BC-41 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Tanita BC-41 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Tanita BC-41 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Tanita BC-41 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Tanita BC-41 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Tanita BC-41 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Tanita BC-41 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Tanita BC-41 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.