INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311-D RANGES: 0-100 PPM; 0-1000 PPM; 1% AND 100% TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CALIF. 91749 TELEPHONE: (626) 934-1500 FAX: (626) 961-2538 TWX: (910) 584-1887 TDYANYL COID EASYLINK: 628-31862 Web: www.
TABLE OF CONTENTS MODEL 311D 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 1 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 2 1.3.5 Compact Packaging 3 1.
1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311-D is a portable trace oxygen analyzer which can be opera ted without an external power source, and reliably calibrated without the use of cumbersome, questionable, “certified” calibration gases.
After the electronics have been properly zeroed (a one-time factory operation), the instrum ent cannot produce an output indication in the absen ce of oxygen; therefore, the need for a “zero” standardization gas is obviated.
Further disassembly may be accom plished by rem oving the backplate assemb ly from its four mounting standoffs, and laying the two separated assem blies out as illustrated on the “Analyzer Wiring Diagram”. The diagram is included among the drawings at the rear of the manual.
necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.
dictated by the customer’s application. If the instrument were being used to certify the oxygen content of a product for delivery, then a calibrati on prior to certification would certainly be in order.
If, on the other hand, the instrument is not to be used imm ediately after calibration, and a low ppm oxygen gas is being employed as a purge, allo w the m anifold to be purged overnight, and then disconnect both male fittings. ALWAYS DISCONNECT THE SOURCE FITTING FIRST, AND IMMEDIATELY THEREAFTER, THE VENT FITTING .
every two (2) days of continuous use. To rech arge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenience outlet.
No tools are required to replace the cell in the in strument. Sim ply unscrew (ccw) the plug at the bottom of the analyzer, and the cell will drop out of the m anifold cavity. Remove the new cell from its package, and carefu lly rem ove the shorting clip.
1. Disconnect cell. 2. Move range switch to “CAL” position. 3. Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally +-.3 mV. 4. Adjust P2 for 0 ± 1 mV at output of A2, pin 6. 5. Verify that the offset is the same on all ranges.
OPERATING TEMPERATURE RANGE: 30 ° TO 125 ° F. RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ± 10% OF AN INTRODUCED GAS HAVING A C.
EFFECT OF CO2 ON B-2C CELL LIFE -11-.
RECOMMENDED SPARE PARTS LIST MODEL 311D QTY. P/N DESCRIPTION 5 F-39 MICRO FUSE, 0.25 AMP., FAST-BLOW (FOR 210-240 VAC) 5 F-51 MICRO FUSE, 0.50 AMP., FAST-BLOW (FOR 100-125 VAC) 1 C-6689-A2C MICRO-FUEL.
DRAWING LIST MODEL 311D B-26473 PICTORAL DIAGRAM C-28339 SCHEMATIC C-26606 W IRING DIAGRAM NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator.
.
Date: 6/20/2006 Time : 01:45:07 PM Title: Indsrsvr01drawingsMASTER26000C26606-14.dwg.
Date: 7/20/04 Time : 10:56:54 AM Title: G:MASTER28000C28339-8.DWG.
Een belangrijk punt na aankoop van elk apparaat Teledyne 311-D (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne 311-D heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne 311-D vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne 311-D leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne 311-D krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne 311-D bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne 311-D kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne 311-D . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.