Gebruiksaanwijzing /service van het product Infrared Gas Analyzer van de fabrikant Teledyne
Ga naar pagina of 80
TELEDYNE ANALYTICAL INSTRUMENTS OPERATING INSTRUCTIONS Series Model 7500 Infrared Gas Analyzer DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUID S OR GASES MAY BE PRE SENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE RE QUIRED WHEN SERVI CING THIS SYSTEM.
2 Model 7500 Instruction Manual T eledyne Analytical Instruments PRELIMINAR Y We are grateful for your purchase of T eledyne Analytical Instrumentʼs Model 7500 Infrared Gas Analyzer .
3 Model 7500 Instruction Manual T eledyne Analytical Instruments SAFETY PRECAUTIONS Please read this section carefully to ensure using the analyzer in the correct way . The cautionary descriptions listed here contain important information about safety , so they should always be observed.
4 Model 7500 Instruction Manual T eledyne Analytical Instruments Caution on piping DANGER In piping, the following precautions should be observed. Wrong piping may cause gas leakage. If the leaking gas contains a toxic component, there is a risk of serious accident being induced.
5 Model 7500 Instruction Manual T eledyne Analytical Instruments Caution on use PROHIBITED • Do not allow metal, nger or others to touch the input/output terminals in the instrument. Otherwise, shock hazard or injury may occur . • Do not smoke nor use a ame near the gas analyzer .
6 Model 7500 Instruction Manual T eledyne Analytical Instruments CONTENTS PREF ACE .............................................................................................................................2 CAUTION ON SAFETY .......................
Teledyne Analytical Instruments 7 Model 7500 Instruction Manual 6.8 Maintenance mode ………………… ………………………………………… ……… 57 6.9 Calibration ……………………… …… ………………………………………… …….
8 Model 7500 Instruction Manual T eledyne Analytical Instruments 1. OVER VIEW The Model 7500 infrared gas analyzer measures the concentrations of CO2, CO, CH4, SO2 and O2 contained in sample gas. CO2, CO, CH4 and SO2 are measured by non-dispersion infrared method, while O2 is measured by paramagnetic method.
9 Model 7500 Instruction Manual T eledyne Analytical Instruments 2. NAME AND DESCRIPTION OF EACH P ART 2.1 Description of each unit Name Description 1. Handle Draws the analyzer unit from the case. 2. Power switch T urns ON/OFF this analyzer . 3. Switch for back light T urn ON/OFF the back light of display .
10 Model 7500 Instruction Manual T eledyne Analytical Instruments 3. INST ALLA TION DANGER This unit is not explosion-proof type. Do not use it in a place with explosive gases to avoid explosion, re or other serious accidents.
1 1 Model 7500 Instruction Manual T eledyne Analytical Instruments 3.2 Installation of analyzer There are two methods of installing the analyzer . For detailed dimensions, see Chapter 9.
12 Model 7500 Instruction Manual T eledyne Analytical Instruments Observe the following when connecting the gas pipes. • The pipes should be connected to the gas inlet and outlet at the rear panel of the analyzer , respectively .
13 Model 7500 Instruction Manual T eledyne Analytical Instruments (2) Sampling gas ow rate A ow rate of sampling gas must be 1 L/min ±0.5 L/min. A ow meter should be provided as shown in Fig. 3-2 to measure ow rate values. (3) Preparation for standard gas Prepare the standard gas for zero/span calibration.
14 Model 7500 Instruction Manual T eledyne Analytical Instruments (6) Example of sampling system conguration The system conguration may vary depending upon the nature of measured gas, coexistent gases or application. A typical conguration diagram is shown in Fig.
15 Model 7500 Instruction Manual T eledyne Analytical Instruments 3.5 Wiring method Caution on wiring CAUTION • Wiring work must be performed with the main power set to OFF to prevent electric shocks. • Enforce construction of class-D grounding wire by all means.
16 Model 7500 Instruction Manual T eledyne Analytical Instruments When a noise source is in the vicinity Do not install the analyzer near power noise generating electric equipment (such as high frequency furnace and electric welder). If the analyzer must be used near such equipment, a separate power line should be used for avoiding noise.
17 Model 7500 Instruction Manual T eledyne Analytical Instruments (4) Contact input (DI): terminal block 2 (13 to 20), terminal block 3 (5 to 10) • It is for a contact input at no voltage. An input is provided when switching to short circuit (on) or open (off).
18 Model 7500 Instruction Manual T eledyne Analytical Instruments (7) List of terminal blocks 1 to 5 *1) Unused terminals are used for internal connection and should not be used as repeating terminals either . *2) O2 sensor input is used when an external O2 analyzer is selected.
19 Model 7500 Instruction Manual T eledyne Analytical Instruments 8) Description on terminal block T erminal block 1 <TN1> T erminal block for analog output (non-isolated output) Output : 4 to 2.
20 Model 7500 Instruction Manual T eledyne Analytical Instruments T erminal block 3 <TN3> Between 1 – 4: For internal connection. Must not be wired. (Must not be used as junction terminal.) Between 5 – 6: Remote hold input. No hold when open.
21 Model 7500 Instruction Manual T eledyne Analytical Instruments T erminal block 4 <TN4> Between 1 – 2: CH4 span calibration contact output Between 3 – 4: CH3 span calibration contact outpu.
22 Model 7500 Instruction Manual T eledyne Analytical Instruments T erminal block 5 <TN5> 1 and 1 1 – 14: For internal connection. Must not be wired. (Must not be used as junction terminal.) Between 2 - 3 and 3 - 4: CH3 alarm output. Conductive at 2 – 3 and open at 3 – 4 when set value is exceeded.
23 Model 7500 Instruction Manual T eledyne Analytical Instruments (9) Timing of calibration contact output 1. Manual calibration (See “6.9 Calibration”) 2.
24 Model 7500 Instruction Manual T eledyne Analytical Instruments 4. OPERA TION 4.1 Preparation for operation (1) Check of gas sampling tube, exhaust tube and wiring Check that the pipes are correctly connected to the gas sampling port and drain port.
25 Model 7500 Instruction Manual T eledyne Analytical Instruments 5. DISPLA Y AND OPERA TION P ANEL This section describes the display and operation panel of the gas analyzer . It also explains the name and description of function on the operation panel.
26 Model 7500 Instruction Manual T eledyne Analytical Instruments 5.2 Overview of display and operation panel Figure 5.2.
27 Model 7500 Instruction Manual T eledyne Analytical Instruments 5.3 Overview of display screen (1) Measurement mode screen On turning on the power switch, the Measurement Mode screen will appear . The measurement screen depends on the number of components.
28 Model 7500 Instruction Manual T eledyne Analytical Instruments • O2 correction concentration average value: CH (component) where “A V / CV” is displayed as “A V / CV CO” in the component .
29 Model 7500 Instruction Manual T eledyne Analytical Instruments (3) Contents of measured channel (CH) The contents in each measured CH corresponding to the type are given below: T ype code Contents .
30 Model 7500 Instruction Manual T eledyne Analytical Instruments 5.4 General operation • Measurement mode The measurement mode can be displayed for up to 5 channels in a single screen. When viewing a channel beyond the 5 channels, press the ( or ) key and the screen can be scrolled one by one channel at a time.
31 Model 7500 Instruction Manual T eledyne Analytical Instruments 6. SETTING AND CALIBRA TION 6.1 Changeover of range This mode is used to select the ranges of measured components. 1. During measurement, press the MODE key to display the User mode. 2.
32 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.2 Calibration setting This mode is used to set calibration concentration and actions. The calibration setting involves calibration concentration, zero calibration, calibration range and auto-calibration component.
33 Model 7500 Instruction Manual T eledyne Analytical Instruments 5. In the “Calibration Concentration Selection” screen that appears, select any concentration item you want to set by pressing the UP , DOWN, and RIGHT key . Note: Analyzers other than the zirconia O2 instrument cannot perform zero setting.
34 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.2.2 Setting of manual zero calibration If zero calibration is to be made manually , select whether to calibrate all components at once or each of them separately upon selection. 1. During measurement, press the MODE key to display the User mode.
35 Model 7500 Instruction Manual T eledyne Analytical Instruments 5. In the “Manual Calibration Selection” screen that appears, select “At once” or “Each” by pressing the UP or DOWN key . When selecting “Both”, the CH (components) to be set can be zero-calibrated at the same time.
36 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.2.3 Setting of calibration range This mode is used to set if the range of each CH (component) at the calibration manual calibration or auto calibration) should be calibrated with a single range or 2 ranges.
37 Model 7500 Instruction Manual T eledyne Analytical Instruments 5. In the “Calibration Selection” screen that appears, select “Both” or “Current” by pressing the UP or DOWN key . • When selecting “both”, Range 1 and Range 2 of the set CH are calibrated together .
38 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.2.4 Setting of auto-calibration component It sets the CH (component) to be calibrated in the auto- calibration. 1. During measurement, press the MODE key to display the User mode. 2. Point the cursor to “Setting about Calibration” by pressing the UP or DOWN key .
39 Model 7500 Instruction Manual T eledyne Analytical Instruments 5. In the “Auto Calibration Selection” screen that appears, select “enable” or “disable” by pressing the UP or DOWN key .
40 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.3 Alarm setting 6.3.1 Setting of alarm values This mode is used to set the upper and lower limit value to provide an alarm output during measurement. Before changing the alarm setting, set the ON/OFF to OFF .
41 Model 7500 Instruction Manual T eledyne Analytical Instruments After setting, the alarm setting is now completed by pressing the ENT key . T o close the “Alarm Setting” T o close the “ Alarm Setting” or to cancel this mode midway , press the ESC key .
42 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.3.2 Hysteresis setting T o prevent chattering of an alarm output near the alarm setting values, set hysteresis. 1. In the “Alarm Setting CH Selection” screen that appears, point the cursor to “Hysteresis” by pressing the UP or DOWN key .
43 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.4 Setting of auto calibration 6.4.1 Auto calibration Auto calibration is automatically carried out when zero calibration and span calibration are set. Before changing the setting of auto calibration, set the ON/OFF to OFF .
44 Model 7500 Instruction Manual T eledyne Analytical Instruments Auto calibration status contact output is closed during auto calibration, and is open in other cases.
45 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.4.2 Forced stop of auto calibration This mode is used to cancel the auto calibration forcedly . 1. In the Menu mode that is displayed, point the cursor to “Setting of Auto Calibration” by pressing the UP or DOWN key .
46 Model 7500 Instruction Manual T eledyne Analytical Instruments “Auto Calibration” screen Example In case where setting the auto calibration components (see Item 6.2.4) to “CH1: enable” and “CH2: enable”. • Zero calibration A message, “Zero calibration” blinks at CH1 and CH2.
47 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.5 Setting of auto zero calibration 6.5.1 Auto zero calibration Auto zero calibration is automatically carried out when zero calibration are set. Components for which a calibration is to be made are determined by setting of auto calibration component in 6.
48 Model 7500 Instruction Manual T eledyne Analytical Instruments Auto calibration status contact output is closed during auto zero calibration, and is open in other cases.
49 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.5.2 Forced stop of auto zero calibration This mode is used to cancel the auto zero calibration forcedly . 1. In the Menu mode that is displayed, point the cursor to “Setting of Auto Zero Calibration” by pressing the UP or DOWN key .
50 Model 7500 Instruction Manual T eledyne Analytical Instruments “Auto Zero Calibration” screen Example In case where setting the auto calibration components (see Item 6.2.4) to “CH1: enable” and “CH2: enable”. • Zero calibration A message, “Zero calibration” blinks at CH1 and CH2.
51 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.6 Peak alarm setting When the number of peaks of which CO concentration exceeds the upper limit value exceeds the setting time, a peak alarm is outputted. This section describes how to perform various settings of peak alarm.
52 Model 7500 Instruction Manual T eledyne Analytical Instruments Setting range • Alarm V alue: 0 to 1000 ppm 5 ppm step (initial value: 500 ppm) • Alarm Count: 1 to 99 times (initial value: 5 times) • Hysteresis: 0 to 20% of full scale (initial value: 0% of full scale) * The hysteresis setting is made in terms of full scale.
53 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.7 Parameter setting It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be set are as follows: Description of setting items • Current T ime: Sets the current day of the week, hour and time.
54 Model 7500 Instruction Manual T eledyne Analytical Instruments 4. In the Parameter Setting screen that appears, enter the numeric values and set the items. Entering the numeric values or setting the items should be carried out by using the or key .
55 Model 7500 Instruction Manual T eledyne Analytical Instruments (2) Auto calibration (3) External hold (4) Screen display during holding The “Holding” message blinks on the measuring screen.
56 Model 7500 Instruction Manual T eledyne Analytical Instruments A verage period It allows you to set an average period of the average value of O2 correction. It enables you to set an average time of 1 to 59 minutes (1- minute step) or 1 to 4 hours (1-hour step).
57 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.8 Maintenance mode This mode is used for check of sensor input values, display of error log les or setting of passwords. First, enter a password and then use it from the next operation.
58 Model 7500 Instruction Manual T eledyne Analytical Instruments • O2 correction reference value setting screen • Password Setting screen Description of Password Setting screen It enables you to set a password to be used when switching the parameter setting mode to the maintenance mode.
59 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.9 Calibration 6.9.1 Zero calibration It is used for zero point adjustment. For zero calibration gas, see 3.3 (3), Preparation for standard gas in Sampling. Use a gas according to application.
60 Model 7500 Instruction Manual T eledyne Analytical Instruments 6.9.2 Span calibration It is used to perform a span point adjustment. Supply calibration gas with concentration set to the span value to perform the span calibration.
61 Model 7500 Instruction Manual T eledyne Analytical Instruments 7. MAINTENANCE 7.1 Daily check (1) Zero calibration and span calibration 1. It is used for zero point adjustment. For calibration, refer to 6.9.1, Zero calibration. 2. After zero calibration, perform span calibration.
62 Model 7500 Instruction Manual T eledyne Analytical Instruments 7.3 Cleaning of measuring cell Entry of dust or water drops in the measuring cell contaminates the interior of the cell, thus resulting in a drift.
63 Model 7500 Instruction Manual T eledyne Analytical Instruments Number Name 1 Screw (for xing light source unit) 2 Screw (for xing detector) 3 Screw (for xing base plate) 4 Base plate 5 Lig.
64 Model 7500 Instruction Manual T eledyne Analytical Instruments (2) How to remove block cell (See Fig. 7-2) 1. For Step 1 to 4, see Item 7.3.1, (1) How to remove pipe cell.
65 Model 7500 Instruction Manual T eledyne Analytical Instruments Number Name 1 Screw (for xing light source unit) (2) Filter 3 Screw (for xing detector) 4 Base plate 5 Light source unit 6 Screw.
66 Model 7500 Instruction Manual T eledyne Analytical Instruments 3) How to remove measuring unit (See Fig. 7-3) 1. For Step 1 to 4, see Item 7.3.1(1), How to remove pipe cell.
67 Model 7500 Instruction Manual T eledyne Analytical Instruments 7.3.2 How to clean cell 1. T o clean the cell inside or infrared ray transmission window , rst clear large dirt of it with a soft brush and then wipe with soft cloth lightly . Donʼt use hard cloth.
68 Model 7500 Instruction Manual T eledyne Analytical Instruments 8. TROUBLESHOOTING 8.1 Error message If errors occur , the following contents are displayed. Error display Error contents Probable causes Error No. 1 Sector motor rotation detector signal is faulty .
69 Model 7500 Instruction Manual T eledyne Analytical Instruments In case of Error No. 5 and No. 7 (2) Error log le If error occurs, the history is saved in an error log le.
70 Model 7500 Instruction Manual T eledyne Analytical Instruments 9. SPECIFICA TIONS 9.1 Specications (1) Standard specications Principle of measurement: CO2, CO, CH4, SO2; Non-dispersion infrar.
71 Model 7500 Instruction Manual T eledyne Analytical Instruments (2) Standard functions Output signal holding: Output signals are held during manual and auto calibrations by activation of holding (turning on its setting). The values to be held are the ones just before start calibration mode.
72 Model 7500 Instruction Manual T eledyne Analytical Instruments CS: measured concentration of sample gas Os : Measured O2 concentration On: Standard O2 concentration (value changeable by setting) *The upper limit value of the fractional part in this calculation is 4.
73 Model 7500 Instruction Manual T eledyne Analytical Instruments 9.2 Code symbols.
74 Model 7500 Instruction Manual T eledyne Analytical Instruments.
75 Model 7500 Instruction Manual T eledyne Analytical Instruments <Code specication for ordering> (1) Code symbols should be specied. (2) Range combination should be the one allowed in separate tables. (note 1) “Y” should be specied when only O2 measurement is necessary .
76 Model 7500 Instruction Manual T eledyne Analytical Instruments Measurable component and range - availability check table -.
77 Model 7500 Instruction Manual T eledyne Analytical Instruments T able 5: T riple component analyzer (CO2, CO, CH4) Covered in combination of T able 2, T able 3 and T able 4.
78 Model 7500 Instruction Manual T eledyne Analytical Instruments 9.3 Outline diagram (1) Main unit.
79 Model 7500 Instruction Manual T eledyne Analytical Instruments (2) Accessory slide rail (unit: mm) * The slide rails are attached to this equipment when designated.
Teledyne Analytical Instruments 80 Mo del 7500 Instruction Manual Appendix Dra wing List D79104 Outline diagram.
Een belangrijk punt na aankoop van elk apparaat Teledyne Infrared Gas Analyzer (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne Infrared Gas Analyzer heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne Infrared Gas Analyzer vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne Infrared Gas Analyzer leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne Infrared Gas Analyzer krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne Infrared Gas Analyzer bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne Infrared Gas Analyzer kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne Infrared Gas Analyzer . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.