Gebruiksaanwijzing /service van het product NALL van de fabrikant Teledyne
Ga naar pagina of 23
INSTRUCTION MANUAL MODEL NALL MASS FLOWMETERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED.
Manual 111-082010_Nall Mass Flowmeters Page 2 of 23 Manual Print History The print history shown below lists the p rinting dates of all revisions and a ddenda created for thi s manual. The revision level letter increases alphabetically as the manual undergoes subseque nt updates.
Manual 111-082010_Nall Mass Flowmeters Page 3 of 23 Table of Contents 1.0 GENERAL INFO RMATI ON ....................................................................................................................................... 5 1.1 F EATURES : .
Manual 111-082010_Nall Mass Flowmeters Page 4 of 23 11.6 S PECIFICATIONS ................................................................................................... E RROR ! B OOKMARK NOT DEFINED . 12.0 WARRA NTY AND REPAI R...................
Manual 111-082010_Nall Mass Flowmeters Page 5 of 23 1.0 General Information 1.1 Features: Hastings Linear Mas s Flowmeters are designed to accurately measure mass flow without corre ctions or compensations for gas pressure and temp erature. Due to a linear electrical out put signal, the flowmeters are ideal for use with totalizers and recorders.
Manual 111-082010_Nall Mass Flowmeters Page 6 of 23 2.0 OPERA TING PRINCIPLE 2.1 Operating Principle Hastings Linear Mass Flowmeters operat e on a unique electrical principle whereby a capillary tube is heated uniformly by a transformer. The temp erature distribution is symmetrical about the midpoint at zero flow (see Figure 2.
Manual 111-082010_Nall Mass Flowmeters Page 7 of 23 3.0 RECEIVING AND INSPECTION 3.1 Initial Inspection: Carefully unpack the Hastings Linear Mass Flowmete r and inspect it for any obvious signs of damage due to shipment. Immediately advise the carrier who de livered the shipment of any suspected damage .
Manual 111-082010_Nall Mass Flowmeters Page 8 of 23 4.0 INST ALLA TION INSTRUCTIONS. 4.1 TRANSDUCER 4.1.1 Orientation of the Transducer The transducer may be mounted in any position, as long as the direction of gas flow thro ugh the transducer is from “IN” to “OUT” as marked on the transducer base.
Manual 111-082010_Nall Mass Flowmeters Page 9 of 23 The use of O-ring seal connectors on the transducers w ith female threads is often more convenient for ma ny low pressure applications, such as with U-Series transducers.
Manual 111-082010_Nall Mass Flowmeters Page 10 of 23 4.4.3 Electrical Zero Check Turn the power supply “ON” and allow the flowmeter 30 minutes to warm up. Stop all flow through the transducer and check the electrical zero. CAUTION: DO NOT ASSUME THAT ALL METERING VA LVES WILL COMPLE TELY SHUT OFF FLOW.
Manual 111-082010_Nall Mass Flowmeters Page 11 of 23 5.0 USING THE FLOWMETER USING THE HASTING S LINEAR MASS FLOWMETER 5.1 Warm-Up Time: When the flowmeter is first turned on, the meter w ill fluctuate somewhat before settling to a stable indication.
Manual 111-082010_Nall Mass Flowmeters Page 12 of 23 Actual flow = meter reading X (KHe) = 50 SCCM X 1.382 = 69.1 SCCM of Helium 5.5.2 Flowmeter Factory Calibrate d for a Special Gas If the Hastings Linear Mass Flowmeter is calibrated for a gas other th an air, it may also be used to measure the flow rate of many other gases.
Manual 111-082010_Nall Mass Flowmeters Page 13 of 23 Example : (A) What is the output sensitivity of HS-1KS? (B) What is the flow rate for a 2.71 volt output? Solution : (A) Sensitivity = rated flow R a t e d o u t p u t = 1000 SCCM 5 . 0 0 V D C = 200 SCCM/VDC (B) Mass Flow = output X sensitivity = 2.
Manual 111-082010_Nall Mass Flowmeters Page 14 of 23 6.0 PRESSURE EFFECTS 6.1 Standard Transducers: The Hastings Linear Mass Flowmeter can measure mass fl ow accurately without co rrections for a variation in line pressure from one psia (.068 ATM) to 250 psig (18 ATMS.
Manual 111-082010_Nall Mass Flowmeters Page 15 of 23 7.0 TEMPERA TURE EFFECTS 7.1 Ambient Temperatures: 7.1.1 Transducer In order to maintain the accuracy of the flowmeter with changes in ambient temperat ure, it is necessary to keep the temperature of the transdu cer between 10° C and 50° C.
Manual 111-082010_Nall Mass Flowmeters Page 16 of 23 8.0 DIFFERENTIAL PRESSURE 8.1 TYPICAL PRESSURE DROPS The differential pressure drop (DP) across the Hastin gs HS-Series Mass Flow Transduce r can vary between 0.50” H2O depending upon the transducer’s range.
Manual 111-082010_Nall Mass Flowmeters Page 17 of 23 8.2.2 Decrease in Line Pressure The relationship stated in Section 8. 2.1 also shows that DP i ncreases prop ortionally as absolute line pressure decreases. This situation is true un til the DP becomes a significant part of the tota l absolute line pressure.
Manual 111-082010_Nall Mass Flowmeters Page 18 of 23 9.0 Calibration 9.1 CALIBRATION PROCEDURES The Hastings Mass Flowmeter has been carefully inspected and calibrated at the f actory before shipment and will give long, reliable service.
Manual 111-082010_Nall Mass Flowmeters Page 19 of 23 TRANSDUCER MILLIVOLT TEST +5 VDC OUTPUT BIAS METER GAIN TRMS AC TEST JUMPER T1 FOR 230V OPERATION FOR 581A ONLY FIG 9.
Manual 111-082010_Nall Mass Flowmeters Page 20 of 23 10.0 Maintenance 10.1 Transducers: 10.1.1 Cleaning With proper care in installing and us e, the transducer will require little or no maintenance. Should the small passages in the transducer become part ially clogged, they can be cleaned with any suitable solvent such as acetone, vythene, etc.
Manual 111-082010_Nall Mass Flowmeters Page 21 of 23 11.0 Accessories OPTIONS AND ACCESSORIES 11.1 Double Point Relay Flowmeter/Alarm ( DNALL-P): The Hastings Flowmeter/Alarm is a complete Linear Mass Flowmeter with all of its charac teristics and ranges.
Manual 111-082010_Nall Mass Flowmeters Page 22 of 23 a) What is the sensitivity of the totalizer? b) What is the total flow indicated by the 13, 500 count? c) What is the average flow rate over the 7 hour period? Solution: a) The full scale flow rate for the TNALL-P/ HS-10S is 10 SCCM.
Manual 111-082010_Nall Mass Flowmeters Page 23 of 23 12.0 W arranty and Rep air 12.1 Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship.
Een belangrijk punt na aankoop van elk apparaat Teledyne NALL (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne NALL heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne NALL vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne NALL leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne NALL krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne NALL bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne NALL kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne NALL . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.