TIC-LF414C Introduction The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the flu.
TIC-LF414C 2 Power consumption: 17W(27VA) or less 19W(29VA) or less (with PROFIBUS) Conformance to European Co mmunity Directives: EMC directive 89/336/EEC The low voltage 93/68/EEC PED 97/23/EC (Note 1) Note : See Table 1 for detail.
TIC-LF414C 3 Digital input DI (opt.) Signal type: 20 to 30Vdc voltage signal Input resistance: 2.7k Ω Number of inputs: one point DI function — One of the following functions can be assigned to the optional DI signal. Range switching — Selects either the higher or lower range in the unidirectional or bidirectional 2-range setting.
TIC-LF414C 4 Conditions when power fails: Parameter setting values are stored in non–volatile memory and the values will be restored when the power r eturns to normal condition.
TIC-LF414C 5 Installation Dimensions Unit : inch( mm) L1 Φ D1 (L2 ) 2. 5 2 ( 64 ) 5.63( 143) 1.42 ( 36) 8.86 ( 225) 5.16( 131) ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lb) 1/2 2.76 9.92 1.93 approx.
TIC-LF414C 6 Unit : inch(mm) (L2 ) L1 8.86( 225) 5.63 ( 143) 1.42( 36) 5.16( 131 ) Φ D1 Note1: Eye bolts are provided at the top fo r flow mete rs sized 8"(200mm). Note2: 1 inch = 25.4mm ANSI class 150 and class 300 dimensions Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lbs) 1-1/2 3.
TIC-LF414C 7 Unit : inch(mm) R ( P T ) 1/2Male S c r e w ( T w o Pl a c e s) 3.46( 88) 1.57 ( 40 ) 4.96 ( 126 ) 3.58 ( 91 ) 1.38( 35) (L2 ) Φ 1.93( Φ 49) 2.52( 64 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 8 Unit : inch(mm) 3.46(88) 1.57 ( 40 ) 4.96(126) 1.38(35) 3.58(91) R(PT)½ Mal e Screw (T wo places) (L2) Φ 2.60( Φ 66) 3.43 ( 87 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part Cable c onnection port at detector.
TIC-LF414C 9 Unit : inch(mm) 3.46( 88) 1 .57 ( 4 0 ) 4.9 6( 126) 1. 38 ( 3 5 ) 3.58(91) R( PT )½M a le Sc r e w ( T w o p l a c e s ) (L 2) L1 Φ D1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 10 矢視 A 信号ケ ー ブ ル グラ ン ド 電源ケ ー ブ ル グ ラ ン ド 封止栓 励磁 ケ ー ブ ル グ ラ ン ド 電流出力ケ ー ブ ル グ ラ ン ド 接地端子 (M 4 ) 取付金具 4 - φ 11 取付金具 82 (3 6 ) 1 5 6 1 3 1 51 178 20 0 1 51 74 ± 0 .
TIC-LF414C 11 ・ Separate type LF410/LF612 and LF414/LF612F flowmeters T erminal board Connected detecto r IV wire 5.5 mm 2 or more Power switch (External double- pole power switch ) Thick walled ste.
TIC-LF414C 12 Wiring Precautions (1) Explosion proof type flow meters are not provided cable glands. Refer to the part Cable connection port at detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100 Ω or less ground resistance).
TIC-LF414C 13 Meter Size To select the meter size: See Table 3 to 4 and find meter sizes within the velocity of 0.3 to 32.8 ft/s (0.1 to 10m/s) for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 3.
TIC-LF414C 14 Table 6. Specification Code (Waf er type detector LF410 Series) Detector ca tegory Model Specific ation Code Normal type Ex. type 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Description Gr.
TIC-LF414C 15 Table 7. Specification Code for converter s Model Specific ation Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Contents L F 6 1 Elec tromagnetic flowmeter c onverter LF610 type LF612 type 0 Comb.
TIC-LF414C TIC-LF414C Specifications are subject to change without notice. Printed in Japan 2008-6 (TDOC) © TOSHIBA Corporation 2008 All Rights Reserved. http://www.toshiba.com/ind / Misuse of this product can result in damages to property or hum an injury.
Een belangrijk punt na aankoop van elk apparaat Toshiba TIC-LF414C (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Toshiba TIC-LF414C heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Toshiba TIC-LF414C vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Toshiba TIC-LF414C leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Toshiba TIC-LF414C krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Toshiba TIC-LF414C bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Toshiba TIC-LF414C kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Toshiba TIC-LF414C . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.