Gebruiksaanwijzing /service van het product MIW-20 van de fabrikant Adams & Co
Ga naar pagina of 12
Adam Equipment MIW-20 BABY SCALE ADAM EQUIPMENT CO. LTD. p.n. 4295, Rev. B, March 2004 © Adam Equipment Company 2004.
© Adam Equipment Company 2004.
CONTENTS 1.0 INTRODUCTION ......................................................................................................... 2 2.0 TECHNICAL SPECIFICATIONS .................................................................................. 3 3.0 LOCATING THE SCALE .
1.0 INTRODUCTION • The MIW-20 scales are precision m edical scales specially designed to weigh babies on a cradle. • The scale has a capacity of 20 kg and readability of 0.01 kg. • The scale displays the weight in kilograms or pounds. • The scale is fast, accurate and simple to use.
2.0 TECHNICAL SPECIFICA TIONS Specifications MIW-20 Maximum Capacity, max 20 Kg (44.4 lbs.) Readability, d 10 g (1/2 oz.) Tare range, -T -20 kg (44.4 lbs.
3.0 LOCA TING THE SCALE • Place the scale on a stable surface. A soft and uneven floor (e.g. carpet, linoleum) could cause an erroneous weight reading. • The ambient temperature of the scale should be kept as constant as possible. Do not place the scale where it may be s ubjected to draughts or heater and direct sunlight.
5.0 KEY DESCRIPTIONS [ON / OFF/TARE] Turns the scale on or off (keep it pressed for 3 seconds). Sets a zero condition when the platform is empty. Sets the display to zero by storing the current weight in the memory to tare the scale. [UNIT] Sets the Units.
• To zero the display, press the [ON/OFF/TARE] key. • In practice when the scale is turned on, the [ON/OFF/TARE] key is pressed to zero the display, if necessary. • After the display is at zero, plac e the baby to be weighed on the cradle of the scale.
8.0 HOLD FUNCTION • Due to the baby’s movement, the we ight shown on the display will change all the time. After the baby is placed on the scale, press [HOLD] to hold the weight value. The display shows ‘ hold ’. The scale will calculate the average of the former weight values taken automatically.
10.0 CALIBRA TION PROCEDURE • Place the scale on a flat platform after inserting a PP3, 9V alkaline battery. • Press the [ON/OFF/TARE] key to turn the power ON.
Manufacturer’s Declaration of Conformity This product has been manufactured in accor dance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatib ility Directive 89/336/EEC Low Voltage Directive 73/23/EEC Adam Equipment Co.
ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sa le of electronic weighing equipments. Products are sold through a world wide di stributor network -supported from our company locations in the UK, USA and SOUTH AFRICA.
Een belangrijk punt na aankoop van elk apparaat Adams & Co MIW-20 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Adams & Co MIW-20 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Adams & Co MIW-20 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Adams & Co MIW-20 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Adams & Co MIW-20 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Adams & Co MIW-20 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Adams & Co MIW-20 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Adams & Co MIW-20 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.