Gebruiksaanwijzing /service van het product 5626/5628 van de fabrikant Fluke
Ga naar pagina of 22
5626/5628 Platinum Resistance Thermometer User’ s Guide Rev. 532801 Hart Scientific.
Rev. 532801 Fluke Hart Scientific • 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 • E-mail: support@ hartscientific.com www.hartscientific.com Subject to change without notice.
T able of Contents 1 Befor e Y ou Start . . . . . . . . . . . . . . . . . . . . . . 1 1 . 1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . 1 1 .2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . 2 1 .2. 1 W arnings . . .
7 . 1 Case Options . . . . . . . . . . . . . . . . . . . . . . . . . 1 5 7 .2 P RT T ermination . . . . . . . . . . . . . . . . . . . . . . . 1 5 8 Tr oubleshooting . . . . . . . . . . . . . . . . . . . . . 1 7 8. 1 Troubleshooting . . . . . . . . . .
iii Figures Figure 1 Hart Model 5626 and 5628 Platinum Resistance Thermometers . . . . . . 5 Figure 2 Standard Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Figure 3 PRT Schematic . . . . . . . . . . . . . . . . . . . . . . . . .
iv T ables T able 1 International Electrical Symb ols . . . . . . . . . . . . . . . . . . . . . . 1 T able 2 Specifications . . . . . . . . . . . . . .
1 Before Y ou Start 1.1 Symbols Used T able 1 lists the Inter national Electrical Symbols . Some or all of these sym - bols may be used on the instrument or in this manual.
Symbol Description Canadian Standards Association OVE RVOL T AG E (Installation) C A TE G ORY I I, Pollution Degree 2 per I E C 1010 - 1 r e f e r s t o t h e l e v e l o f I m p u l s e W i t h s t a n d V o l t a g e p r o t e c t i o n provided .
• DO N OT drop or bang the probe in any way . This w ill cause damage to the probe internally and affect its calibration. • Read Section 5 entitled “PRT Care and Handling Guidelines” before re - moving the P RT from the shipping box or case. Incorrect handling can damage the PRT and void the warranty .
Phone: +86-1 0-6-5 1 2-3436 T elefax: +86-1 0-6-5 1 2-343 7 E-mail: xing ye.han@fluke.com .cn Fluke South East Asia Pte Ltd . Fluke AS EAN Reg ional Office Serv ice Center 60 Alexandra T errace #03-1 6 The Comtech (Lobby D) 1 1 8502 S INGAPORE Phone: +65 6 799-5588 T elefax: +65 6 7 99-5588 E-mail: antng@singa .
2 Introduction 2.1 General The Hart Platinum Resistance Thermometers (P RT) models 5626 and 5628, are designed to b e the best sec ondary standard interpolating instrument converting temperature to resistance. The P RTs are used with a readout de - vic e to detect temperature changes or actual temperature.
calibrated using a primar y standard that has been calibrated in a primary calibration laborator y using know n intrinsic values . 2.3 Calibration In order for any instrument to be used as a standard it must be calibrated.
3 Specifications and W arranty 3.1 Specifications See T able 2 and Figure 2. 7 3 Specifications and Warranty 5626 5628 Temperature Range –200°C to 661°C R tp 100 Ω ±1 Ω at 0.01°C 25.5 Ω ±0.5 Ω at 0.01°C Resistance RatioW(Ga) W(302.9146K) ≥ 1.
3.2 Construction The Model 5626 and 5628 P RT s cover the range from –200°C to 66 1°C . The sensor elements are crafted using high purity platinum w ire wound in a strain free design on a spec ially designed support . 3.3 Electrical Circuit The PRTs are provided w ith a terminal box handle.
4 Installation 4.1 Environmental Issues Ideally, temperature calibration equipment should be used in a calibration laborator y or other facility spec ifically designed for this purpose.
5 PRT Care and Handling Guidelines 5.1 PRT Care CAUT I O N: R E AD T H I S S E CT I O N B E FO R E RE M OVI N G T H E P RT FR O M T H E S H I P P I N G BOX O R CAS E The 5626 and 5628 Platinum Resistance Thermometers (P RTs) are delicate instruments. Care must b e taken in handling the P RT s to maintain calibra - tion accurac y .
• DO allow sufficient time for the thermometer to stabilize before making measurements. This allows for the best accurac y . • DO use the correct drive current with the thermometer to prevent error in temperature or resistance. Hart Scientific recommends 1mA.
6 Operation 6.1 General F or best results, be familiar w ith the operation of the heat source and the r e a d - o u ti n s t r u m e n t .B es u r et of o l l o wt he manufacturer’s instructions for the read-out instrument and the heat source.
6.4 Thermal EMF Two factors contribute to thermal E M F, chemical c onsistency and physical consistenc y . V ariations in chemical structure due to impurities can contrib - ute to thermal E M F . Also discrepancies in crystal structure can contribute to thermal E M F .
7 Accessories 7.1 Case Options The 5626/5628 PRT comes in a rig id case appropriate for the leng th of the probe. • Model Number 26 0 1 P RT Protective Case for 5626-1 2 and 5628-1 2 PR T • Model Number 26 02 PRT Protective Case for 5626-1 5, 5628-1 5, 5626-20, and 5628-20 PRT 7.
8 T roubleshooting 8.1 T roubleshooting In the event that the probe appears to function abnormally, this section may be of use in solv ing the problem . Several possible problem conditions are described along w ith likely causes and solutions . If a problem arises, please read this section carefully and attempt to understand and solve the problem .
Problem Causes and Solutions Temperature readout differ - ent than expected, e.g. the heat source is set at 300°C, the PRT measures 275°C. • Measure the PRT resistance at TPW. • If the resistance of the PRT is less than the rated resis - tance, e.
Een belangrijk punt na aankoop van elk apparaat Fluke 5626/5628 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Fluke 5626/5628 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Fluke 5626/5628 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Fluke 5626/5628 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Fluke 5626/5628 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Fluke 5626/5628 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Fluke 5626/5628 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Fluke 5626/5628 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.