USER GUIDE SCC-TC Series Thermocouple Input Modules The SCC-TC Series thermocouple input modules, SC C-TC01 and SCC-TC02, accept input signals from B-, E-, J-, K-, N-, R-, S-, and T-type thermocouples. The thermocouple inpu ts are filtered and passed into a differential amplif ier with a gain of 100.
SCC-TC Series Thermocouple Input Modules User Guide 2 ni.com When symbol is marked on a product, it denotes a warning advising you to take precautions to av oid electrical shock. When symbol is marked on a product, it denotes a component that may be hot.
© National Instruments Corpor ation 3 SCC-TC Series Th ermocouple Input Mo dules User Guide ❑ Numbers 1 and 2 Phillips screwdri vers ❑ W i re insulation stripper s ❑ NI-D A Q (current version) for Wi ndows 2000/NT/XP Note Software scaling of measurements is not supported on the Macintosh operating system.
SCC-TC Series Thermocouple Input Modules User Guide 4 ni.com A blue stripe on the label identifies the SCC-TC0 X as an analog input module. The label also displays the ico n sho wn in Figure 1. Figure 1. SCC-TC0 X Icon Y ou can plug the SCC-TC0 X into any analog input so cket on the SC-2345.
© National Instruments Corpor ation 5 SCC-TC Series Th ermocouple Input Mo dules User Guide The SCC-TC02 has a f ixed scre w-term inal receptacle and a remov able screw-terminal block, as sho wn in Figure 2.
SCC-TC Series Thermocouple Input Modules User Guide 6 ni.com The amplified thermocouple signal and cold-junction sensor sig nal are measured by E Serie s D A Q device channel X and channel X + 8 respecti vely , where X is 0 to 7 depending on where you plug the SCC-TC0 X .
© National Instruments Corpor ation 7 SCC-TC Series Th ermocouple Input Mo dules User Guide Y ou can use the follo wing formulas to con vert the cold-junction sensor voltage to cold-junction temperature: where T K is the temperature in kelvin. where a = 1.
SCC-TC Series Thermocouple Input Modules User Guide 8 ni.com Detecting Open Thermocouples The SCC-TC0 X contains a 10 M Ω pull-up resistor that connects to + 5 V to detect open thermocouples. To determine if you have an open thermocouple, check whether the corresponding E Series DAQ dev ice channel is saturated.
© National Instruments Corpor ation 9 SCC-TC Series Th ermocouple Input Mo dules User Guide NI-DAQmx Complete the following steps to configure th e SCC-TC module using NI-DAQmx. 1. Open Measurement & A utom ation Explorer . 2. Right-click Devices and Interfaces and select Create New .
SCC-TC Series Thermocouple Input Modules User Guide 10 ni.com 6. Under Accessory: , select SC-23 45 . 7. Click the Configure button. A ne w window appears listing the sock ets (connector reference designators) of the SC-2345. 8. Select the SC-2345 socket where you installed an SCC.
© National Instruments Corpor ation 11 SCC-TC Series Thermocouple Input Modules User Guide 8. Enter the specif ic values for your application in the Setti ngs tab . Context help information for each setting is pro vided on the right side of the screen.
SCC-TC Series Thermocouple Input Modules User Guide 12 ni.com Calibrating the SCC-TC0 X Using a Thermocouple Calibrator To calibrate the system using a thermocouple calibrator, complete the following steps: 1. Make sure that the thermo couple connections at both the SCC and the thermocouple calibrator ar e at the same temperature.
© National Instruments Corpor ation 13 SCC-TC Series Thermocouple Input Modules User Guide where V TC is thermocouple voltage. V ESERIES is E Series D A Q device v oltage. This step provides proper scaling for the th ermocouple amplifier in the SCC-TC0 X .
SCC-TC Series Thermocouple Input Modules User Guide 14 ni.com Amplifier Characteristics Open thermocouple detectio n current.....250 nA max 1 (assuming floating ther mocouple) Common-mode rejection rati o ................110 dB min 1 Output range ....
© National Instruments Corpor ation 15 SCC-TC Series Thermocouple Input Modules User Guide Cold-Junction Sensor Cold-junction sensor accuracy .............. . ± 0.4 ° C max from 15 to 35 ° C, ± 0.75 ° C max from 0 to 15 ° C and 35 to 55 ° C Output.
SCC-TC Series Thermocouple Input Modules User Guide 16 ni.com Physical TC01 dimensions ........................... .........7.8 cm by 2.92 cm by 1.85 cm (3.1 in. by 1.15 in. by 0.73 in.) TC02 dimensions ........................... .........8.89 cm by 2.
© National Instruments Corpor ation 17 SCC-TC Series Thermocouple Input Modules User Guide Note For UL and other safety certifications, refer to the product label, or visit ni.com/hardref.nsf , search by model number or product line, and click the appropriate link in the Certification column.
SCC-TC Series Thermocouple Input Modules User Guide 18 ni.com I/O Connector Pin Assignments Figure 4 shows the I/O connector pins on the bottom o f the module. Figure 4. SCC Module Bottom View T able 1 lists the signal connection corresponding to each pin.
© National Instruments Corpor ation 19 SCC-TC Series Thermocouple Input Modules User Guide 13 + 15 V 14 –15 V 15 — 16 — 17 — 18 — 19 — 20 — Table 1.
National Instruments™, NI™, ni.com ™, and NI-D AQ™ are trademarks of Nationa l Instruments Corporation. Product and com pany names mentioned herein are trademarks or trade names of their respective companies.
Een belangrijk punt na aankoop van elk apparaat National Instruments SCC-TC02 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen National Instruments SCC-TC02 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens National Instruments SCC-TC02 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding National Instruments SCC-TC02 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over National Instruments SCC-TC02 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van National Instruments SCC-TC02 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de National Instruments SCC-TC02 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met National Instruments SCC-TC02 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.