SCXI-1125 Calibration Procedure 2 ni.com Equipment and Other T est Requirements This section describes the equipment, software, documentation, and test conditions required for calibrat ing the SCXI-1125.
© National Instruments Corpor ation 3 SCXI-1125 Calibration Procedur e Documentation The NI-DAQ Function Refe rence Online Help fi le and NI-DAQ User Manual for PC Compa tibles c o n t a i nd e t a i l e di n f o r m a t i o no nu s i n gt h e NI-DAQ driver.
SCXI-1125 Calibration Procedure 4 ni.com Setting Up Y our Module Complete the following steps to set up yo ur module for testing : 1. Install the SCXI-1125 in sl ot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the S CXI module and the E Series D A Q de vice installed in the host computer .
© National Instruments Corpor ation 5 SCXI-1125 Calibration Procedur e 7. Connect the calibrator to the appro priate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front conn ector correspond to t he positiv e and negati ve inputs of the specified channel.
SCXI-1125 Calibration Procedure 6 ni.com 13. Call SCXI_Scale . Set the following parameters: • SCXIChassisID — The v alue assigned b y MAX • moduleSlot — 1 • channel — The same channel num.
© National Instruments Corpor ation 7 SCXI-1125 Calibration Procedur e • calOp — ND_SELF_CALIBRATE • s etOfCalConstants — ND_USER_EEPROM_AREA • calRefVolts — 0.
SCXI-1125 Calibration Procedure 8 ni.com • filterMode — 1 • cutoffFrequency — 10,000 9. Connect the calibrator to the analog input you w ant to adjust. Refer to Figure 1 to determine the pins on the 96-pin f ront connector that correspond to the positiv e and neg ati ve inputs of the specif ied channel.
© National Instruments Corpor ation 9 SCXI-1125 Calibration Procedur e 16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain v alues you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining chan nels you want to adjust.
SCXI-1125 Calibration Procedure 10 ni.com V erifying Adjusted V alues After completing the adjustment s to your SCXI-1125 module, it is important that you verify the analog input operation by repeating t he steps listed in the Veri fying Gain and Offset Errors section of this docu ment.
© National Instruments Corpor ation 11 SCXI-1125 Calibration Procedur e Front and Rear Signal Connector Diagrams Figure 1. Front Signal Connector Pin Assignments CHSGND +5 V RSVD CH7+ CH6+ CH5+ CH4+ .
SCXI-1125 Calibration Procedure 12 ni.com Figure 2. Rear Signal Connector Pin Assignments SERCLK DIGGND SLOT0SEL* DAQD*/A SERDATIN OUTREF MCH7+ MCH6+ MCH5+ MCH4+ MCH3+ MCH2+ MCH1+ MCH0+ AOGND SCANCLK .
© National Instruments Corpor ation 13 SCXI-1125 Calibration Procedur e Specifications The following table contains the sp ecifications you should use w hen verifying and calibrating y our SCXI-1125 module. If you r device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit valu es.
SCXI-1125 Calibration Procedure 14 ni.com 200 – 0.025 – 0.024750 – 0.02468642 – 0 .02481358 250 0.02 0.019800 0.01985126 0.019 74874 250 0 0.000000 0.000022 – 0.000022 250 – 0.02 – 0.019800 – 0.01974874 – 0.0 1985126 500 0.01 0.009900 0.
Een belangrijk punt na aankoop van elk apparaat National Instruments SCXI -1125 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen National Instruments SCXI -1125 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens National Instruments SCXI -1125 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding National Instruments SCXI -1125 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over National Instruments SCXI -1125 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van National Instruments SCXI -1125 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de National Instruments SCXI -1125 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met National Instruments SCXI -1125 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.