DP26-RTD Differential Temperature
Meter with RTD

® On-Line Service www.omega.com
Internet e-mail info@omega.com
Servicing North America:
USA:PP One Omega Drive, P.O. Box 4047 ISO 9001 CertifiedP Stamford CT 06907-0047 PP TEL: (203) 359-1660PP FAX: (203) 359-7700 PP e-mail: info@omega.com
Canada:@ P 976 Bergar PP Laval (Quebec) H7L 5A1 PP TEL: (514) 856-6928PP FAX: (514) 856-6886 PP e-mail: [email protected]
For immediate technical or application assistance:
USA and Canada:P Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA® PP Customer Service: 1-800-622-2378 / 1-800-622-BEST® PP Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN®
Mexico and P TEL: (001)800-TC-OMEGA®P FAX: (001) 203-359-7807 Latin American:P En Español: (001) 203-359-7803P PP e-mail: espanol@omega.com
Servicing Europe:
Czech Republic:P Frystatska 184, 733 01 Karviná PP TEL: +420 59 6311899PP FAX: +420 59 6311114 PP e-mail: [email protected]
Germany/Austria:P Daimlerstrasse 26, D-75392 Deckenpfronn, Germany PP TEL: +49 7056 9398-0PP FAX: +49 7056 9398-29 PP Toll Free in Germany: 0800 639 7678 PP e-mail: [email protected]
United Kingdom:P One Omega Drive ISO 9002 CertifiedP River Bend Technology Centre PP Northbank, Irlam Manchester M44 5BD United Kingdom PP TEL: +44 161 777 6611P FAX: +44 161 777 6622 PP Toll Free in England: 0800 488 488 PP e-mail: sales@omega.co.uk
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications. !
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
® · ® · OMEGAnet

Manual Objectives This manual shows you how to set up and use the Differential RTD Meter.
* Checking voltage jumpers, or changing voltage power
* Mounting the panel
* Selecting the input type
* Selecting a decimal point position
* Selecting reading configuration (Fahrenheit or Celsius)
* Setting setpoint configuration
* Setting setpoint deadbands
* Entering temperature offset
* Accessing the deviation mode
* Selecting meter display
* Enabling/disabling analog output
* Selecting analog output as current or voltage
* Selecting analog output or proportional control
* Routing analog output
* Selecting proportional band
* Using manual reset (offsetting setpoint errors)
* Scaling analog output
For first-time users: Refer to the QuickStart Manual for basic operation and set-up instructions.



Information that is especially important to note is identified by these labels:



Page 4.8 SELECTING OUTPUT CONFIGURATION (OT.CF) 26 4.8.1 Enabling or Disabling The Analog Output 26 4.8.2 Selecting Analog Output as Current or Voltage 27 4.8.3 Selecting Analog Output or Proportional Control 27 4.9 ROUTING ANALOG OUTPUT (OUT) 29 4.10 SELECTING PROPORTIONAL BAND (P.BND) 29 4.11 USING MANUAL RESET (M.RST) 31 4.12 SCALING THE ANALOG OUTPUT (OT.S.O) 32 4.13 RTD TEMPERATURE OFFSET ERROR CORRECTION 36
Section

Figure 2-1. Front-Panel 5 Figure 2-3. Connector Label (AC-Powered and DC-Powered Detail) 9 Figure 3-1. Main Board Power Jumpers (W1, W2, W3) 12 Figure 3-2. Main Board Jumper Positions 13 Figure 3-3. Upper Option Board Installation 13 Figure 3-4. Meter - Exploded View 15 Figure 3-5. Panel Cut-Out 15 Figure 3-6. 2-Wire RTD Input Connection 16 Figure 3-7. 3-Wire RTD Input Connection 16 Figure 3-8. 4-Wire RTD Input connection 16 Figure 3-9. Main Power Connections - AC Powered Unit 17 Figure 3-10. Main Power Connections - DC Powered Unit 18 Figure 3-11. Relay Output Connections 18 Figure 3-12. Analog Output Connections 19 Figure 3-13. Isolated Analog Output Connections 19 Figure 4-1 Proportional Band 29 Figure 12-1. Meter Dimensions 53
Table A-1. Sections of the Manual ii Table 2-1. Accessories and Add-Ons 4 Table 2-2. Connector Description 10 Table 3-1. S3 Jumper Functions 14 Table 3-2. Main Power Connections - AC Powered Unit 18 Table 6-1. Truth Table for Display Values 39 Table 8-1. Display Messages 42 Table 9-1. Configuration Menu 43 Table 10-1. Front Panel Displays 45 Table 10-2 Run Mode Displays 48 Table 11-1. Setpoint Configuration Displays 49 Table 13-1. Factory Preset Values 54

1.1 UNPACKING Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, use the phone numbers listed on the back cover to contact the Customer Service Department nearest you.
Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary. Verify that you receive the following items in the shipping box:
1 Meter with all applicable connectors attached.
1 Set Mounting brackets
If you ordered any of the available options (except the "BL" Blank Lens option), they will be shipped in a separate container to avoid any damage to your indicator/controller.
1 Owner's Manual · 1 QuickStart Manual

This device is marked with the international caution symbol. It is important to
read this manual before installing or commissioning this device as it contains
This instrument is a panel mount device protected in accordance with EN 61010-1:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the main supply cord.
Furthermore, to provide protection against excessive energy being drawn from the main supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
Failure to follow all instructions and warnings may result in injury!

2.1 DESCRIPTION The Differential RTD meter is a value packed indicator/controller. Four full digits accurately display your temperature. Select from DIN (Alpha = .00385) or NIST (Alpha = .00392); 2,3, or 4 wire RTD. Your meter may be a basic indicator or it may include analog output or dual relay output. Analog or dual relay output must be ordered at time of purchase. Analog output is fully scalable and may be configured as a proportional controller, or to follow your display. Dual 5 amp, form C relays control critical processes. Front panel peak detection and memory is also standard. A mechanical lockout has been included to guard against unauthorized changes.
2.2 STANDARD FEATURES The following is a list of features:
* 4-digit, 14 segment LED Display * NEMA 4/Type 4 Front Bezel * ±0.5° C accuracy * Automatic Decimal Adjustment * Non-volatile memory * 115 or 230 V ac 50/60 Hz power supply or 10-32 Vdc or 26-56 Vdc * T1, T2, T1-T2 display
The following is a list of optional features
* Dual 5 amps, form C relay outputs * Scalable analog output * Analog out Proportional control * Easy setup for proportional control * Front-panel deviation correction * Easy RTD error offset calibration

Table 2.1 Accessories and Add-ons
Add-On Options
FS Special Calibration/Configuration SPC4 NEMA-4 Splash Proof Cover SPC18 NEMA-4 Splash Proof Cover, NEW
TP1A Trimplate panel adaptor. Adapts DIN1A/DIN2A cases to larger panel cutouts RP18 19-In. Rack Panel for one (1) 1/8 DIN instrument RP28 19-In. Rack Panel for two (2) 1/8 DIN instrument RP38 19-In. Rack Panel for three (3) 1/8 DIN instrument

Figure 5-1 shows the location of each part of the front of the meter.
- T2 T1
RESET MENU DEV T1/T2/ SETPTS
/ /
Digital LED Display: -1.9.9.9. or 9.9.9.9. 4-digit, 14 segment, 13.8 mm (0.54") high LED display with programmable decimal point.
These meter display windows light when appropriate:
5 Pushbuttons for programming the meter.

METER BUTTONS
In the run mode, press this button to sequentially recall the previous setpoint settings. After using the /T1/T2 and /DEV buttons to alter these settings as desired, press the SETPTS button to store these new values.
Unless you press the SETPTS, /T1/T2 or /DEV button within 20 seconds to store your input, the meter will scroll to setpoint 2 the return to the run mode.
If the dual relay option is not installed or if the “L.3=1” on the “LK.CF” menu, pressing the SETPTS button will display the meter's firmware version.
/T1/T2 Button
In the run mode, pressing the /T1/T2 button will scroll to show T1, T2, and T1-T2.
During the configuration mode or Setpoint modes, use the /T1/T2 button to change the values of the flashing digit shown on the display and/or toggle between menu choices, such as "R.1=F" or "R.1=C". When configuring your setpoint values, press the /T1/T2 button to increment the flashing digit from 0 to 9 by 1's.
In the run mode press the /DEV button to display the deviation of reading from Setpoint 1.
In the configuration or setpoint modes, press the /DEV button to scroll to the next digit.
SETPTS Button · /DEV Button

In the run mode, press the MENU button to terminate the current measuring process and enter you into the configuration mode.
In the configuration mode, press the MENU button to store changes in the non-volatile memory and then advance you to the next menu item.
To lock the MENU button, install S3E jumper. Refer to Table 3-1 for more information about the S3 jumpers.
In the run mode, press the RESET button to:
* reset the latched setpoints. The display will show "SP.RS" (Setpoint Reset).
* perform tuning for Proportional Control if the meter is setup to serve as a proportional controller (0.3=P refer to Section 4.8.3 and Section 7) the meter will show “TUNE”.
Also in the run mode, if you hard reset or power off/on the meter, it shows "RST" followed by "D.RTD".
In the configuration mode, press the RESET button once to review the previous menu. Pressing the RESET button twice results in a hard reset and returns you to the run mode.
In the setpoint mode, press RESET button to go back to the run mode. THe meter will show “RUN”.
MENU Button · RESET Button

2.5.1 PUSH BUTTON LOCKOUT To lock the RESET, MENU and SETPTS buttons, verify that the S3-A jumper is installed and the S3-E jumper is not installed, then follow these steps:
In the run mode-
To unlock the RESET, MENU and SETPTS buttons, follow these steps: 1. Press and hold down the RESET button. The meter shows “LOCK”. Do not release the RESET button.
To lock all front panel buttons, remove the S3-A jumper (refer to Figure 3-2).
To lock the MENU button only, verify that the S3-B jumper is removed, then install the S3-A and S3-E jumpers, (refer to Table 3-1).

2.6 BACK OF THE METER Figure 2-2 shows the connector label mounted at the top of the meter housing. Table 2-2 gives a brief description of each connector at the back of the meter.


Table 2-2 Connector Description
Connector Description TB1-1 Setpoint 1: Normally open (N.O.1) connection TB1-2 Setpoint 1: Normally closed (N.C.1) connection TB1-3 Setpoint 1: Common (COM1) connection TB1-4 Setpoint 2: Normally open (N.O.2) connection TB1-5 Setpoint 2: Normally closed (N.C.2) connection TB1-6 Setpoint 2: Common (COM2) connection TB1-7 ac line connection (no connections on dc-powered units) TB1-8 ac neutral connection (+ Input on dc-powered units) TB1-9 ac Earth ground (-dc-power return on dc-powered units) TB1-10 Analog 1 voltage output TB1-11 Analog 2 current output TB1-12 Analog 3 return TB2-1 +E2: Positive excitation for input #2 TB2-2 +S2: Positive signal input for input #2 TB2-3 -S2: Negative signal input for input #2 TB2-4 -E2: Negative excitation for input #2 TB2-5 +E1: Positive excitation for input #1 TB2-6 +S1: Positive signal input for input #1 TB2-7 -S1: Negative signal input for input #1 TB2-8 -E1: Negative excitation for input #1 TB5-1 Isolated Analog Voltage Output TB5-2 Isolated Analog Current Output TB5-3 Isolated Analog Output Return

2.7 DISASSEMBLY You may need to open up the meter for one of the following reasons:
Disconnect the power supply before proceeding.
To remove and access the main board, follow these steps:

If you power off/on the meter, or perform a hard reset (press the RESET button twice), the meter shows "RST", followed by "D.RTD".
3.1 RATING/PRODUCT LABEL This label is located on top of the meter housing (Refer to Figure 3-3).
To check voltage jumpers, or to change from 115 V to 230 V ac:
Figure 3-1 shows the location of solder jumpers W1 through W3 on the main board.

3.2 MAIN BOARD POWER JUMPERS (Continued) Figure 3-2 shows the main board jumpers.



S2 jumpers are for sensor break indications:
* S2A jumper is not used * S2B jumper is for positive sensor break on input 1 (i.e. heating) * S2C jumper is for positive sensor break on input 2 (i.e. heating) * S2D is not used
S3 jumpers are used for the following (refer to Table 3-1):
* To enable or disable the front panel push-buttons * To allow for an extremely low resistance load for analog output * To disable the MENU button * To perform calibration procedure
Test pins TP1 - TP11 are for testing purposes. Do not use as reading errors may result.
Table 3-1 S3 Jumper Functions Jumper Description S3-A Install to enable front panel push-buttons. Remove to disable all front panel push-buttons S3-B Removed. Install for factory calibration only. S3-C Normally removed. Install for analog voltage output when load is less than 1 Kohm impedance. Care should be taken when installing this jumper. S3-D Removed. Not used. S3-E If installed without S3-B, the MENU button locks out. While in run mode, if you press the MENU button, the display shows “LOCK".

PANEL CUT-OUT
REAR COVER (REMOVED)
"NEW" STYLE
MOUNTING
CONNECTOR
LABEL
BRACKET
PRODUCT
LABEL
CASE
"OLDER" STYLE
MOUNTING BRACKET 2 PCS GASKET
FRONT BEZEL
Figure 3-3 Meter - Exploded View
6,4 (.25) MAX 0,8 (.03) MIN
1,5 R(.06) 4 PLCS
45,00 + 0,61/-0,00 (1.772 + .024/–.000)
92,00 + 0,81/–0,00 (3.622 + .032/–.000)

3.4 CONNECTING SENSOR INPUT Figure 3-5 through 3-7 illustrate how to connect you power sensors.
+S2 -E2
RTD
+E1 +S1 -S1 -E1
RTD
Figure 3-5 2-Wire RTD Input Connection
+E2 +S2
-S2 -E2 RTD
+E1
+S1
-E1 -S1
RTD
Figure 3-6 3-Wire RTD Input Connection
+E2 +S2 -S2 -E2 RTD
4 3 1 2
+S1 +E1
-S1 -E1
Figure 3-7 4-Wire RTD Input Connection

3.5 CONNECTING MAIN POWER Connect the ac main power connections as shown in Figure 3-8.

Table 3-2 shows the wire color and respective terminal connections for both USA and Europe.


Connect the dc main power connections as shown in Figure 3-9..

3.6 ANALOG AND RELAY OUTPUT CONNECTIONS
If you have purchased a meter with analog or dual relay or isolated analog output, refer to the following figures for output connections.



























































/DEV MENU DESCRIPTION /T1/T2/
P.BND Proportional Band select from 0000 through 9999
Scroll right one digit Changes the value of the flashing digit
M.RST Manual Reset select from 0000 through 9999
Scroll right one digit Changes the value of the flashing digit
T1.OF RTD Temperature Offset Channel 1 Shows prior value entered. Scroll right one digit
Changes the value of the flashing digit
T2.OF RTD Temperature Offset Channel 2 Changes the value of the flashing digit Shows prior value entered. Scroll right one digit
/DEV DISPLAY DESCRIPTION /T1/T2/ RESET
LATCHED RESET
Press RESET button to reset your setpoints.
Run Mode

/DEV DESCRIPTION MENU /T1/T2/
SP 1 Setpoint 1
Press to scroll to the next digitto the right
Press to change the value of the flashing digit
Select from -1999 through 9999
Scroll right one digit SP 2 Setpoint 2
Press to change the value of the flashing digit
Select from -1999 through 9999

NIST (0.00392) type 2,3, or 4 wire
Lead Resistance for Specified Accuracy:
2 Wire Up to 55 milliohm/lead 3 Wire Up to 10 ohms/lead balanced 4 Wire Up to 20 ohms/total unbalanced
Isolation: Dielectric strength to 2500V transient per 3mm spacing based on EN 61010 for 260Vrms or dc working voltage
200 mA for current input ranges

~ AC units 115/230 V~(AC) ±10%, 50/60 Hz 7 W max, power consumption (Non-Isolated Analog Output) 8 W max, power consumption (Isolated Analog Output)
DC units 10-32 Vdc 6 W max, power consumption (Non-Isolated Analog Output) 7 W max, power consumption (Isolated Analog Output)
Power Fuse 115 V 125 mA @ 250 (T) 230 V 63 mA @ 250 (T) UL 248-14 (Listed Fuse)
Power Fuse 115 V 175 mA @ 250 V Slow-Blow 230 V 80 mA @ 250 V Slow-Blow

Signal level: Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.
Line Regulation: 0.02% /V ac Step Response Time: 2 - 3 seconds to 99% of the final value
2 Form "C" on/off relays. Configurable for latched and unlatched by software.
Max current: 5 AMPS, resistive load.
ISOLATED ANALOG OUTPUT (TB5, if applicable) Same as non-isolated analog output except isolated to 1000 Vdc.
Signal Level: Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.
Only one analog output is available on each unit and it must be factory installed.

20,3 (.80)
RETAINER
151,4 (5.96)
CASE
REAR COVER
SIDE VIEW TOP VIEW
PANEL THICKNESS
6,4 (.25) MAX 0,8 (.03) MIN
1,5 R(.06) 4 PLCS
45,00 + 0,61/-0,00 (1.772 + .024/–.000)
92,00 + 0,81/–0,00 (3.622 + .032/–.000)

MENU ITEM FACTORY PRESET VALUES
INPT Input Type: 385.3 (3 wire DIN RTD) DEC.P Decimal Point Position: FFF.F RD.CF Reading Configuration: R.1=F (Fahrenheit) S1.CF Setpoint 1 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S.3=0 (Setpoint 1 is assigned to T1-T2) S2.CF Setpoint 2 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S.3=0 (Setpoint 2 is assigned to T1-T2) S1.DB Setpoint 1 Deadband: .0003 S2.DB Setpoint 2 Deadband: .0003 OT.CF Output Configuration: O.1=E (Analog output is enabled) O.2=C (Analog output is current) O.3=A (Analog output follows the display value) OUT Analog Output Routing: T1 - T2 OT.S.O Output Scale and Offset: 0-1000 = 4-20 mA dc T1.OF RTD for T1: 000.0 T2.OF RTD for T2: 000.0 SP1 Setpoint 1 Value: 000.0 SP2 Setpoint 2 Value: 000.0

This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory. Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min
Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc. This unit should not be used in Measurement Categories II, III, IV. Transients Overvoltage Surge (1.2 / 50uS pulse)
EMC EN61326:1997 + and A1:1998 + A2:2001 Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA: 1. Purchase Order number to cover the COST of the repair, 2. Model and serial number of product, and 3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. USED UNDER LICENSE. Other US and International Patents pending or applied for.

Process Measurement and Control?
OMEGA…Of Course! Shop on line at www.omega.com
TEMPERATURE Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers
PRESSURE, STRAIN AND FORCE Transducers & Strain Gauges Load Cells & Pressure Gauges Displacement Transducers Instrumentation & Accessories
FLOW/LEVEL Rotameters, Gas Mass Flowmeters & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers
pH/CONDUCTIVITY pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment
DATA ACQUISITION Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters
HEATERS Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments
M1865/0206 11869ML-96 Rev.B