Inhoudsopgave 80 secties omega.com e-mail: [email protected] For latest product manuals · omegamanual.info · DP26-RTD Differential Temperature · Meter with RTD · … TonenVerbergen
omega.com e-mail: [email protected] For latest product manualsp.1omegamanual.infop.1DP26-RTD Differential Temperaturep.1Meter with RTDp.1PREFACEp.3Standard Procedures:p.3Optional Procedures:p.3NOTES, WARNINGS and CAUTIONSp.5List of Figures Figurep.8List of Tables Tablep.8SECTION 1. INTRODUCTIONp.9QTY DESCRIPTIONp.9SECTION 1.2 SAFETY CONSIDERATIONSp.10SECTION 2. ABOUT THE METERp.112.3 AVAILABLE ACCESSORIESp.12Table 2.1 Accessories and Add-onsp.12Add-On Optionsp.12Accessoriesp.122.4 FRONT OF THE METERp.13••••p.13C Fp.13- T2 T1p.13RESET MENU DEV T1/T2/ SETPTSp.13Figure 2-1 Front Panelp.132.4 FRONT OF THE METER (Continued)p.14METER BUTTONSp.142.5 FRONT PANEL LOCKOUTp.162.5.2 JUMPER LOCKOUTp.16Figure 2-2 Connector label (ac Powered and dc Powered detail)p.172.6 BACK OF THE METER (Continued)p.18Table 2-2 Connector Descriptionp.18SECTION 3. GETTING STARTEDp.203.2 MAIN BOARD POWER JUMPERSp.20Figure 3.1 Main Board Power Jumpersp.20Figure 3.2 Main Board Jumper Positionsp.21Figure 3.3 Upper Option Board Installationp.213.2 MAIN BOARD POWER JUMPERS (Continued)p.223.3 MOUNTING THE METERp.23PANEL CUT-OUTp.23REAR COVER (REMOVED)p.23"NEW" STYLEp.23MOUNTINGp.23CONNECTORp.23LABELp.23BRACKETp.23PRODUCTp.23CASEp.23"OLDER" STYLEp.23MOUNTING BRACKET 2 PCS GASKETp.23FRONT BEZELp.23Figure 3-3 Meter - Exploded Viewp.23PANEL THICKNESSp.236,4 (.25) MAX 0,8 (.03) MINp.231,5 R(.06) 4 PLCSp.23Figure 3-4 Panel Cut-Outp.23+E2 -S2p.24+S2 -E2p.24RTDp.24+E1 +S1 -S1 -E1p.24Figure 3-5 2-Wire RTD Input Connectionp.24+E2 +S2p.24-S2 -E2 RTDp.24+E1p.24+S1p.24-E1 -S1p.24Figure 3-6 3-Wire RTD Input Connectionp.24+E2 +S2 -S2 -E2 RTDp.246 7 RTDp.24+S1 +E1p.24-S1 -E1p.24Figure 3-7 4-Wire RTD Input Connectionp.24Figure 3-8 Main Power Connections (ac powered unit)p.25Table 3-2 Main Power Connections (ac powered unit)p.253.5 CONNECTING MAIN POWER (Continued)p.26Figure 3-9 Main Power Connections (dc powered unit)p.263.6 ANALOG AND RELAY OUTPUT CONNECTIONSp.26Figure 3-10 Relay Output Connectionsp.263.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)p.27Figure 3-11 Analog Output Connectionsp.27Figure 3-12 Isolated Analog Output Connectionsp.27
Page 1 of the Omega DP26-RTD user manual — omega.com e-mail: info@omega.com For latest product manuals
Pagina 1 · klik om te vergroten
Pagina 1
Omega DP26-RTD — omega.com e-mail: info@omega.com For latest product manuals diagram (page 1)
Afb. 1 · omega.com e-mail: [email protected] For latest product…
Omega DP26-RTD — omega.com e-mail: info@omega.com For latest product manuals diagram (page 1)
Afb. 2 · omega.com e-mail: [email protected] For latest product…

omega.com e-mail: [email protected] For latest product manuals

omegamanual.info

DP26-RTD Differential Temperature

Meter with RTD

Page 2 of the Omega DP26-RTD user manual — omega.com e-mail: info@omega.com For latest product manuals
Pagina 2 · klik om te vergroten
Pagina 2

® On-Line Service www.omega.com

Internet e-mail info@omega.com

Servicing North America:

USA:PP One Omega Drive, P.O. Box 4047 ISO 9001 CertifiedP Stamford CT 06907-0047 PP TEL: (203) 359-1660PP FAX: (203) 359-7700 PP e-mail: info@omega.com

Canada:@ P 976 Bergar PP Laval (Quebec) H7L 5A1 PP TEL: (514) 856-6928PP FAX: (514) 856-6886 PP e-mail: [email protected]

For immediate technical or application assistance:

USA and Canada:P Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA® PP Customer Service: 1-800-622-2378 / 1-800-622-BEST® PP Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN®

Mexico and P TEL: (001)800-TC-OMEGA®P FAX: (001) 203-359-7807 Latin American:P En Español: (001) 203-359-7803P PP e-mail: espanol@omega.com

Servicing Europe:

Czech Republic:P Frystatska 184, 733 01 Karviná PP TEL: +420 59 6311899PP FAX: +420 59 6311114 PP e-mail: [email protected]

Germany/Austria:P Daimlerstrasse 26, D-75392 Deckenpfronn, Germany PP TEL: +49 7056 9398-0PP FAX: +49 7056 9398-29 PP Toll Free in Germany: 0800 639 7678 PP e-mail: [email protected]

United Kingdom:P One Omega Drive ISO 9002 CertifiedP River Bend Technology Centre PP Northbank, Irlam Manchester M44 5BD United Kingdom PP TEL: +44 161 777 6611P FAX: +44 161 777 6622 PP Toll Free in England: 0800 488 488 PP e-mail: sales@omega.co.uk

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.

The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications. !

This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.

® · ® · OMEGAnet

Page 3 of the Omega DP26-RTD user manual — Optional Procedures:
Pagina 3 · klik om te vergroten
Pagina 3

PREFACE

Manual Objectives This manual shows you how to set up and use the Differential RTD Meter.

Standard Procedures:

* Checking voltage jumpers, or changing voltage power

* Mounting the panel

* Selecting the input type

* Selecting a decimal point position

* Selecting reading configuration (Fahrenheit or Celsius)

* Setting setpoint configuration

* Setting setpoint deadbands

* Entering temperature offset

* Accessing the deviation mode

* Selecting meter display

Optional Procedures:

* Enabling/disabling analog output

* Selecting analog output as current or voltage

* Selecting analog output or proportional control

* Routing analog output

* Selecting proportional band

* Using manual reset (offsetting setpoint errors)

* Scaling analog output

For first-time users: Refer to the QuickStart Manual for basic operation and set-up instructions.

Page 4 of the Omega DP26-RTD user manual — PREFACE
Pagina 4 · klik om te vergroten
Pagina 4
Omega DP26-RTD — PREFACE diagram (page 4)
Afb. 3
Page 5 of the Omega DP26-RTD user manual — NOTES, WARNINGS and CAUTIONS
Pagina 5 · klik om te vergroten
Pagina 5

NOTES, WARNINGS and CAUTIONS

Information that is especially important to note is identified by these labels:

  • NOTE
  • WARNING
  • CAUTION
  • IMPORTANT
Page 6 of the Omega DP26-RTD user manual — NOTES, WARNINGS and CAUTIONS
Pagina 6 · klik om te vergroten
Pagina 6
Page 6 of the Omega DP26-RTD user manual — NOTES, WARNINGS and CAUTIONS
Page 7 of the Omega DP26-RTD user manual — NOTES, WARNINGS and CAUTIONS
Pagina 7 · klik om te vergroten
Pagina 7

Page 4.8 SELECTING OUTPUT CONFIGURATION (OT.CF) 26 4.8.1 Enabling or Disabling The Analog Output 26 4.8.2 Selecting Analog Output as Current or Voltage 27 4.8.3 Selecting Analog Output or Proportional Control 27 4.9 ROUTING ANALOG OUTPUT (OUT) 29 4.10 SELECTING PROPORTIONAL BAND (P.BND) 29 4.11 USING MANUAL RESET (M.RST) 31 4.12 SCALING THE ANALOG OUTPUT (OT.S.O) 32 4.13 RTD TEMPERATURE OFFSET ERROR CORRECTION 36

5. ACCESSING THE DEVIATION MODE 38
6. SELECTING METER DISPLAY 39
7. TUNING THE PROPORTIONAL CONTROLLER 41
8. DISPLAY MESSAGES 42
9. MENU CONFIGURATION 43
10. FRONT PANEL DISPLAYS 45
11. SETPOINT CONFIGURATION DISPLAYS 48
12. SPECIFICATIONS 49
13. FACTORY PRESET VALUES 53
14. CE APPROVALS 54

Section

Page 8 of the Omega DP26-RTD user manual — List of Tables Table
Pagina 8 · klik om te vergroten
Pagina 8

List of Figures Figure

Figure 2-1. Front-Panel 5 Figure 2-3. Connector Label (AC-Powered and DC-Powered Detail) 9 Figure 3-1. Main Board Power Jumpers (W1, W2, W3) 12 Figure 3-2. Main Board Jumper Positions 13 Figure 3-3. Upper Option Board Installation 13 Figure 3-4. Meter - Exploded View 15 Figure 3-5. Panel Cut-Out 15 Figure 3-6. 2-Wire RTD Input Connection 16 Figure 3-7. 3-Wire RTD Input Connection 16 Figure 3-8. 4-Wire RTD Input connection 16 Figure 3-9. Main Power Connections - AC Powered Unit 17 Figure 3-10. Main Power Connections - DC Powered Unit 18 Figure 3-11. Relay Output Connections 18 Figure 3-12. Analog Output Connections 19 Figure 3-13. Isolated Analog Output Connections 19 Figure 4-1 Proportional Band 29 Figure 12-1. Meter Dimensions 53

List of Tables Table

Table A-1. Sections of the Manual ii Table 2-1. Accessories and Add-Ons 4 Table 2-2. Connector Description 10 Table 3-1. S3 Jumper Functions 14 Table 3-2. Main Power Connections - AC Powered Unit 18 Table 6-1. Truth Table for Display Values 39 Table 8-1. Display Messages 42 Table 9-1. Configuration Menu 43 Table 10-1. Front Panel Displays 45 Table 10-2 Run Mode Displays 48 Table 11-1. Setpoint Configuration Displays 49 Table 13-1. Factory Preset Values 54

Page 9 of the Omega DP26-RTD user manual — QTY DESCRIPTION
Pagina 9 · klik om te vergroten
Pagina 9

SECTION 1. INTRODUCTION

1.1 UNPACKING Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, use the phone numbers listed on the back cover to contact the Customer Service Department nearest you.

Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.

The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary. Verify that you receive the following items in the shipping box:

QTY DESCRIPTION

1 Meter with all applicable connectors attached.

1 Set Mounting brackets

If you ordered any of the available options (except the "BL" Blank Lens option), they will be shipped in a separate container to avoid any damage to your indicator/controller.

1 Owner's Manual · 1 QuickStart Manual

Page 10 of the Omega DP26-RTD user manual — SECTION 1.2 SAFETY CONSIDERATIONS
Pagina 10 · klik om te vergroten
Pagina 10

SECTION 1.2 SAFETY CONSIDERATIONS

This device is marked with the international caution symbol. It is important to

read this manual before installing or commissioning this device as it contains

This instrument is a panel mount device protected in accordance with EN 61010-1:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the main supply cord.

Furthermore, to provide protection against excessive energy being drawn from the main supply in case of a fault in the equipment, an overcurrent protection device shall be installed.

  • Do not exceed voltage rating on the label located on the top of the instrument housing.
  • Always disconnect power before changing signal and power connections.
  • Do not use this instrument on a work bench without its case for safety reasons.
  • Do not operate this instrument in flammable or explosive atmospheres.
  • Do not expose this instrument to rain or moisture.
  • Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
  • Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards. EMC Considerations
  • Whenever EMC is an issue, always use shielded cables.
  • Never run signal and power wires in the same conduit.
  • Use signal wire connections with twisted-pair cables.
  • Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.

Failure to follow all instructions and warnings may result in injury!

Page 11 of the Omega DP26-RTD user manual — SECTION 2. ABOUT THE METER
Pagina 11 · klik om te vergroten
Pagina 11

SECTION 2. ABOUT THE METER

2.1 DESCRIPTION The Differential RTD meter is a value packed indicator/controller. Four full digits accurately display your temperature. Select from DIN (Alpha = .00385) or NIST (Alpha = .00392); 2,3, or 4 wire RTD. Your meter may be a basic indicator or it may include analog output or dual relay output. Analog or dual relay output must be ordered at time of purchase. Analog output is fully scalable and may be configured as a proportional controller, or to follow your display. Dual 5 amp, form C relays control critical processes. Front panel peak detection and memory is also standard. A mechanical lockout has been included to guard against unauthorized changes.

2.2 STANDARD FEATURES The following is a list of features:

* 4-digit, 14 segment LED Display * NEMA 4/Type 4 Front Bezel * ±0.5° C accuracy * Automatic Decimal Adjustment * Non-volatile memory * 115 or 230 V ac 50/60 Hz power supply or 10-32 Vdc or 26-56 Vdc * T1, T2, T1-T2 display

The following is a list of optional features

* Dual 5 amps, form C relay outputs * Scalable analog output * Analog out Proportional control * Easy setup for proportional control * Front-panel deviation correction * Easy RTD error offset calibration

Page 12 of the Omega DP26-RTD user manual — Accessories
Pagina 12 · klik om te vergroten
Pagina 12

2.3 AVAILABLE ACCESSORIES

Table 2.1 Accessories and Add-ons

Add-On Options

FS Special Calibration/Configuration SPC4 NEMA-4 Splash Proof Cover SPC18 NEMA-4 Splash Proof Cover, NEW

Accessories

TP1A Trimplate panel adaptor. Adapts DIN1A/DIN2A cases to larger panel cutouts RP18 19-In. Rack Panel for one (1) 1/8 DIN instrument RP28 19-In. Rack Panel for two (2) 1/8 DIN instrument RP38 19-In. Rack Panel for three (3) 1/8 DIN instrument

Page 13 of the Omega DP26-RTD user manual — Figure 2-1 Front Panel
Pagina 13 · klik om te vergroten
Pagina 13

2.4 FRONT OF THE METER

Figure 5-1 shows the location of each part of the front of the meter.

C F

- T2 T1

RESET MENU DEV T1/T2/ SETPTS

/ /

Figure 2-1 Front Panel

Digital LED Display: -1.9.9.9. or 9.9.9.9. 4-digit, 14 segment, 13.8 mm (0.54") high LED display with programmable decimal point.

These meter display windows light when appropriate:

1. Setpoint 1 status 2 - Setpoint 2 status C - °C unit indicator F - °F unit indicator T1 - T1 on display T2 - T2 on display T1-T2 - T1-T2 on display

5 Pushbuttons for programming the meter.

Page 14 of the Omega DP26-RTD user manual — 2.4 FRONT OF THE METER (Continued)
Pagina 14 · klik om te vergroten
Pagina 14

2.4 FRONT OF THE METER (Continued)

METER BUTTONS

In the run mode, press this button to sequentially recall the previous setpoint settings. After using the /T1/T2 and /DEV buttons to alter these settings as desired, press the SETPTS button to store these new values.

Unless you press the SETPTS, /T1/T2 or /DEV button within 20 seconds to store your input, the meter will scroll to setpoint 2 the return to the run mode.

If the dual relay option is not installed or if the “L.3=1” on the “LK.CF” menu, pressing the SETPTS button will display the meter's firmware version.

/T1/T2 Button

In the run mode, pressing the /T1/T2 button will scroll to show T1, T2, and T1-T2.

During the configuration mode or Setpoint modes, use the /T1/T2 button to change the values of the flashing digit shown on the display and/or toggle between menu choices, such as "R.1=F" or "R.1=C". When configuring your setpoint values, press the /T1/T2 button to increment the flashing digit from 0 to 9 by 1's.

In the run mode press the /DEV button to display the deviation of reading from Setpoint 1.

In the configuration or setpoint modes, press the /DEV button to scroll to the next digit.

SETPTS Button · /DEV Button

Page 15 of the Omega DP26-RTD user manual — 2.4 FRONT OF THE METER (Continued)
Pagina 15 · klik om te vergroten
Pagina 15

2.4 FRONT OF THE METER (Continued)

In the run mode, press the MENU button to terminate the current measuring process and enter you into the configuration mode.

In the configuration mode, press the MENU button to store changes in the non-volatile memory and then advance you to the next menu item.

To lock the MENU button, install S3E jumper. Refer to Table 3-1 for more information about the S3 jumpers.

In the run mode, press the RESET button to:

* reset the latched setpoints. The display will show "SP.RS" (Setpoint Reset).

* perform tuning for Proportional Control if the meter is setup to serve as a proportional controller (0.3=P refer to Section 4.8.3 and Section 7) the meter will show “TUNE”.

Also in the run mode, if you hard reset or power off/on the meter, it shows "RST" followed by "D.RTD".

In the configuration mode, press the RESET button once to review the previous menu. Pressing the RESET button twice results in a hard reset and returns you to the run mode.

In the setpoint mode, press RESET button to go back to the run mode. THe meter will show “RUN”.

MENU Button · RESET Button

Page 16 of the Omega DP26-RTD user manual — 2.5.2 JUMPER LOCKOUT
Pagina 16 · klik om te vergroten
Pagina 16

2.5 FRONT PANEL LOCKOUT

2.5.1 PUSH BUTTON LOCKOUT To lock the RESET, MENU and SETPTS buttons, verify that the S3-A jumper is installed and the S3-E jumper is not installed, then follow these steps:

In the run mode-

1. Press and hold down the RESET button. The meter shows “SP.RS” . Do not release the RESET button.
2. Press the MENU button and hold RESET button down (3-5 seconds) until the meter shows “LOCK” .
3. Release both buttons.

To unlock the RESET, MENU and SETPTS buttons, follow these steps: 1. Press and hold down the RESET button. The meter shows “LOCK”. Do not release the RESET button.

2. Press the MENU button and hold RESET button down (3-5 seconds) until the meter shows “UN.LK” .
3. Release both buttons.

2.5.2 JUMPER LOCKOUT

To lock all front panel buttons, remove the S3-A jumper (refer to Figure 3-2).

To lock the MENU button only, verify that the S3-B jumper is removed, then install the S3-A and S3-E jumpers, (refer to Table 3-1).

Page 17 of the Omega DP26-RTD user manual — Figure 2-2 Connector label (ac Powered and dc Powered detail)
Pagina 17 · klik om te vergroten
Pagina 17

2.6 BACK OF THE METER Figure 2-2 shows the connector label mounted at the top of the meter housing. Table 2-2 gives a brief description of each connector at the back of the meter.

Omega DP26-RTD — Figure 2-2 Connector label (ac Powered and dc Powered detail) diagram (page 17)
Afb. 4 · Figure 2-2 Connector label (ac Powered and dc Powere…

Figure 2-2 Connector label (ac Powered and dc Powered detail)

Page 18 of the Omega DP26-RTD user manual — 2.6 BACK OF THE METER (Continued)
Pagina 18 · klik om te vergroten
Pagina 18

2.6 BACK OF THE METER (Continued)

Table 2-2 Connector Description

Connector Description TB1-1 Setpoint 1: Normally open (N.O.1) connection TB1-2 Setpoint 1: Normally closed (N.C.1) connection TB1-3 Setpoint 1: Common (COM1) connection TB1-4 Setpoint 2: Normally open (N.O.2) connection TB1-5 Setpoint 2: Normally closed (N.C.2) connection TB1-6 Setpoint 2: Common (COM2) connection TB1-7 ac line connection (no connections on dc-powered units) TB1-8 ac neutral connection (+ Input on dc-powered units) TB1-9 ac Earth ground (-dc-power return on dc-powered units) TB1-10 Analog 1 voltage output TB1-11 Analog 2 current output TB1-12 Analog 3 return TB2-1 +E2: Positive excitation for input #2 TB2-2 +S2: Positive signal input for input #2 TB2-3 -S2: Negative signal input for input #2 TB2-4 -E2: Negative excitation for input #2 TB2-5 +E1: Positive excitation for input #1 TB2-6 +S1: Positive signal input for input #1 TB2-7 -S1: Negative signal input for input #1 TB2-8 -E1: Negative excitation for input #1 TB5-1 Isolated Analog Voltage Output TB5-2 Isolated Analog Current Output TB5-3 Isolated Analog Output Return

Page 19 of the Omega DP26-RTD user manual — 2.6 BACK OF THE METER (Continued)
Pagina 19 · klik om te vergroten
Pagina 19

2.7 DISASSEMBLY You may need to open up the meter for one of the following reasons:

  • To check or change the 115 or 230 Vac power jumpers.
  • To install or remove jumpers on the main board.

Disconnect the power supply before proceeding.

To remove and access the main board, follow these steps:

  • Disconnect the main power from the meter.
  • Remove the back case cover.
  • Lift the back of the main board upwards and let it slide out of the case.
Page 20 of the Omega DP26-RTD user manual — Figure 3.1 Main Board Power Jumpers
Pagina 20 · klik om te vergroten
Pagina 20

SECTION 3. GETTING STARTED

If you power off/on the meter, or perform a hard reset (press the RESET button twice), the meter shows "RST", followed by "D.RTD".

3.1 RATING/PRODUCT LABEL This label is located on top of the meter housing (Refer to Figure 3-3).

3.2 MAIN BOARD POWER JUMPERS

To check voltage jumpers, or to change from 115 V to 230 V ac:

1. Remove the main board from the case. Refer to Section 2.7.
2. Locate the solder jumpers W1, W2, and W3 (located near the edge of the main board alongside the transformer - refer to Figure 3-1).
3. If your power requirement is 115 V ac, solder jumpers W1 and W3 should be wired, but jumper W2 should not. If your power requirement is 230 V ac, solder jumper W2 should be wired, but jumpers W1 and W3 should not.

Figure 3-1 shows the location of solder jumpers W1 through W3 on the main board.

Figure 3.1 Main Board Power Jumpers

Page 21 of the Omega DP26-RTD user manual — Figure 3.3 Upper Option Board Installation
Pagina 21 · klik om te vergroten
Pagina 21

3.2 MAIN BOARD POWER JUMPERS (Continued) Figure 3-2 shows the main board jumpers.

Omega DP26-RTD — Figure 3.2 Main Board Jumper Positions diagram (page 21)
Afb. 5 · Figure 3.2 Main Board Jumper Positions

Figure 3.2 Main Board Jumper Positions

Omega DP26-RTD — Figure 3.2 Main Board Jumper Positions diagram (page 21)
Afb. 6 · Figure 3.2 Main Board Jumper Positions

Figure 3.3 Upper Option Board Installation

Page 22 of the Omega DP26-RTD user manual — 3.2 MAIN BOARD POWER JUMPERS (Continued)
Pagina 22 · klik om te vergroten
Pagina 22

3.2 MAIN BOARD POWER JUMPERS (Continued)

S2 jumpers are for sensor break indications:

* S2A jumper is not used * S2B jumper is for positive sensor break on input 1 (i.e. heating) * S2C jumper is for positive sensor break on input 2 (i.e. heating) * S2D is not used

S3 jumpers are used for the following (refer to Table 3-1):

* To enable or disable the front panel push-buttons * To allow for an extremely low resistance load for analog output * To disable the MENU button * To perform calibration procedure

Test pins TP1 - TP11 are for testing purposes. Do not use as reading errors may result.

Table 3-1 S3 Jumper Functions Jumper Description S3-A Install to enable front panel push-buttons. Remove to disable all front panel push-buttons S3-B Removed. Install for factory calibration only. S3-C Normally removed. Install for analog voltage output when load is less than 1 Kohm impedance. Care should be taken when installing this jumper. S3-D Removed. Not used. S3-E If installed without S3-B, the MENU button locks out. While in run mode, if you press the MENU button, the display shows “LOCK".

Page 23 of the Omega DP26-RTD user manual — Figure 3-4 Panel Cut-Out
Pagina 23 · klik om te vergroten
Pagina 23

3.3 MOUNTING THE METER

PANEL CUT-OUT

REAR COVER (REMOVED)

"NEW" STYLE

MOUNTING

CONNECTOR

LABEL

BRACKET

PRODUCT

LABEL

CASE

"OLDER" STYLE

MOUNTING BRACKET 2 PCS GASKET

FRONT BEZEL

Figure 3-3 Meter - Exploded View

1. Cut a hole in your panel, as shown in Figure 3-3. For specific dimensions refer to Figure 3-4. 2. Insert the meter into the hole. Be sure the front bezel gasket is flush to the panel. 3. Slide on mounting bracket to secure. 4. Proceed to Section 3.4 to connect your sensor input and main power.

PANEL THICKNESS

6,4 (.25) MAX 0,8 (.03) MIN

1,5 R(.06) 4 PLCS

45,00 + 0,61/-0,00 (1.772 + .024/–.000)

92,00 + 0,81/–0,00 (3.622 + .032/–.000)

Figure 3-4 Panel Cut-Out

Page 24 of the Omega DP26-RTD user manual — 6 7 RTD
Pagina 24 · klik om te vergroten
Pagina 24

3.4 CONNECTING SENSOR INPUT Figure 3-5 through 3-7 illustrate how to connect you power sensors.

+E2 -S2

+S2 -E2

RTD

+E1 +S1 -S1 -E1

RTD

Figure 3-5 2-Wire RTD Input Connection

+E2 +S2

-S2 -E2 RTD

+E1

+S1

-E1 -S1

RTD

Figure 3-6 3-Wire RTD Input Connection

+E2 +S2 -S2 -E2 RTD

4 3 1 2

6 7 RTD

+S1 +E1

-S1 -E1

Figure 3-7 4-Wire RTD Input Connection

Show 6 more labels / short text on this page
Channel 2 · Channel 1 · Channel 2 · Channel 1 · Channel 2 · Channel 1
Page 25 of the Omega DP26-RTD user manual — Table 3-2 Main Power Connections (ac powered unit)
Pagina 25 · klik om te vergroten
Pagina 25

3.5 CONNECTING MAIN POWER Connect the ac main power connections as shown in Figure 3-8.

Omega DP26-RTD — Figure 3-8 Main Power Connections (ac powered unit) diagram (page 25)
Afb. 7 · Figure 3-8 Main Power Connections (ac powered unit)

Figure 3-8 Main Power Connections (ac powered unit)

Table 3-2 shows the wire color and respective terminal connections for both USA and Europe.

Table 3-2 Main Power Connections (ac powered unit)

Omega DP26-RTD — Table 3-2 Main Power Connections (ac powered unit) diagram (page 25)
Afb. 8 · Table 3-2 Main Power Connections (ac powered unit)
Page 26 of the Omega DP26-RTD user manual — Figure 3-10 Relay Output Connections
Pagina 26 · klik om te vergroten
Pagina 26

3.5 CONNECTING MAIN POWER (Continued)

Connect the dc main power connections as shown in Figure 3-9..

Omega DP26-RTD — 3.5 CONNECTING MAIN POWER (Continued) diagram (page 26)
Afb. 9 · 3.5 CONNECTING MAIN POWER (Continued)

Figure 3-9 Main Power Connections (dc powered unit)

3.6 ANALOG AND RELAY OUTPUT CONNECTIONS

If you have purchased a meter with analog or dual relay or isolated analog output, refer to the following figures for output connections.

Omega DP26-RTD — Figure 3-9 Main Power Connections (dc powered unit) diagram (page 26)
Afb. 10 · Figure 3-9 Main Power Connections (dc powered unit)

Figure 3-10 Relay Output Connections

Page 27 of the Omega DP26-RTD user manual — Figure 3-12 Isolated Analog Output Connections
Pagina 27 · klik om te vergroten
Pagina 27

3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)

Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 27)
Afb. 11 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)

Figure 3-11 Analog Output Connections

Omega DP26-RTD — Figure 3-11 Analog Output Connections diagram (page 27)
Afb. 12 · Figure 3-11 Analog Output Connections

Figure 3-12 Isolated Analog Output Connections

Page 28 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 28 · klik om te vergroten
Pagina 28
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 28)
Afb. 13 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 29 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 29 · klik om te vergroten
Pagina 29
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 29)
Afb. 14 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 30 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 30 · klik om te vergroten
Pagina 30
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 30)
Afb. 15 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 31 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 31 · klik om te vergroten
Pagina 31
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 31)
Afb. 16 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 32 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 32 · klik om te vergroten
Pagina 32
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 32)
Afb. 17 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 33 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 33 · klik om te vergroten
Pagina 33
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 33)
Afb. 18 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 34 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 34 · klik om te vergroten
Pagina 34
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 34)
Afb. 19 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 35 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 35 · klik om te vergroten
Pagina 35
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 35)
Afb. 20 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 36 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 36 · klik om te vergroten
Pagina 36
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 36)
Afb. 21 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 37 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 37 · klik om te vergroten
Pagina 37
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 37)
Afb. 22 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 38 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 38 · klik om te vergroten
Pagina 38
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 38)
Afb. 23 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 39 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 39 · klik om te vergroten
Pagina 39
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 39)
Afb. 24 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 40 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 40 · klik om te vergroten
Pagina 40
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 40)
Afb. 25 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 41 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 41 · klik om te vergroten
Pagina 41
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 41)
Afb. 26 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 42 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 42 · klik om te vergroten
Pagina 42
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 42)
Afb. 27 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 43 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 43 · klik om te vergroten
Pagina 43
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 43)
Afb. 28 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 44 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 44 · klik om te vergroten
Pagina 44
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 44)
Afb. 29 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 45 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 45 · klik om te vergroten
Pagina 45
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 45)
Afb. 30 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 46 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 46 · klik om te vergroten
Pagina 46
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 46)
Afb. 31 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 47 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 47 · klik om te vergroten
Pagina 47
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 47)
Afb. 32 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 48 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 48 · klik om te vergroten
Pagina 48
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 48)
Afb. 33 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 49 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 49 · klik om te vergroten
Pagina 49
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 49)
Afb. 34 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 50 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 50 · klik om te vergroten
Pagina 50
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 50)
Afb. 35 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 51 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 51 · klik om te vergroten
Pagina 51
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 51)
Afb. 36 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 52 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 52 · klik om te vergroten
Pagina 52
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 52)
Afb. 37 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 53 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 53 · klik om te vergroten
Pagina 53
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 53)
Afb. 38 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 54 of the Omega DP26-RTD user manual — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Pagina 54 · klik om te vergroten
Pagina 54
Omega DP26-RTD — 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued) diagram (page 54)
Afb. 39 · 3.6 ANALOG AND RELAY OUTPUT CONNECTIONS (continued)
Page 55 of the Omega DP26-RTD user manual — Table 10-2 Run Mode Display
Pagina 55 · klik om te vergroten
Pagina 55

Table 10-1 Front-Panel Displays (continued)

/DEV MENU DESCRIPTION /T1/T2/

P.BND Proportional Band select from 0000 through 9999

Scroll right one digit Changes the value of the flashing digit

M.RST Manual Reset select from 0000 through 9999

Scroll right one digit Changes the value of the flashing digit

T1.OF RTD Temperature Offset Channel 1 Shows prior value entered. Scroll right one digit

Changes the value of the flashing digit

T2.OF RTD Temperature Offset Channel 2 Changes the value of the flashing digit Shows prior value entered. Scroll right one digit

Table 10-2 Run Mode Display

/DEV DISPLAY DESCRIPTION /T1/T2/ RESET

LATCHED RESET

Press RESET button to reset your setpoints.

Run Mode

Page 56 of the Omega DP26-RTD user manual — SETPOINT 2
Pagina 56 · klik om te vergroten
Pagina 56

SECTION 11. SETPOINT CONFIGURATION DISPLAYS

Table 11-1 Setpoints Configuration Displays

/DEV DESCRIPTION MENU /T1/T2/

SP 1 Setpoint 1

Press to scroll to the next digitto the right

Press to change the value of the flashing digit

SETPOINT 1

Select from -1999 through 9999

Scroll right one digit SP 2 Setpoint 2

Press to change the value of the flashing digit

SETPOINT 2

Select from -1999 through 9999

Page 57 of the Omega DP26-RTD user manual — ACCURACY AT 25°C
Pagina 57 · klik om te vergroten
Pagina 57

SECTION 12. SPECIFICATIONS

SIGNAL INPUT

RTD typesDIN (0.00385) type 2,3, or 4 wire

NIST (0.00392) type 2,3, or 4 wire

Lead Resistance for Specified Accuracy:

2 Wire Up to 55 milliohm/lead 3 Wire Up to 10 ohms/lead balanced 4 Wire Up to 20 ohms/total unbalanced

Isolation: Dielectric strength to 2500V transient per 3mm spacing based on EN 61010 for 260Vrms or dc working voltage

Noise RejectionNormal Mode Rejection (NMR) = 60dB Common Mode Rejection (CMR) = 120dB
Protection240 V rms max for voltage input ranges

200 mA for current input ranges

Standard Display4-digit, 14-segment LED, 13.8 mm (0.54”)
Symbol(-1.9.9.9. ~ 9.9.9.9.)

ANALOG TO DIGITAL

TechniqueDual slope
Internal resolution15 bits
Read rate3/seconds for each channel
PolarityAutomatic

ACCURACY AT 25°C

Accuracy±0.50°C after 30 minute warm-up
Temperature Stability0.04°C/°C
Step response1 second to 99% of the final value
Page 58 of the Omega DP26-RTD user manual — MECHANICAL
Pagina 58 · klik om te vergroten
Pagina 58

INPUT POWER INFORMATION:

~ AC units 115/230 V~(AC) ±10%, 50/60 Hz 7 W max, power consumption (Non-Isolated Analog Output) 8 W max, power consumption (Isolated Analog Output)

DC units 10-32 Vdc 6 W max, power consumption (Non-Isolated Analog Output) 7 W max, power consumption (Isolated Analog Output)

External Fuse RequiredIEC 127-2/III

Power Fuse 115 V 125 mA @ 250 (T) 230 V 63 mA @ 250 (T) UL 248-14 (Listed Fuse)

Power Fuse 115 V 175 mA @ 250 V Slow-Blow 230 V 80 mA @ 250 V Slow-Blow

ENVIRONMENT

Operating temperature0 to 50°C (32° to 122°F)
Storage temperature-40° to 85°C (-40° to 185°F)
Relative humidity90% at 40°C (non-condensing)

MECHANICAL

Panel cutout1/8 DIN 3.62 x 1.78" (92 x 45 mm)
Weight1.27 lb (574 g)
Case materialPolycarbonate, 94 V-O UL rated
Page 59 of the Omega DP26-RTD user manual — ALARM OUTPUTS (if applicable)
Pagina 59 · klik om te vergroten
Pagina 59

ANALOG OUTPUT (if applicable)

Signal typeCurrent or voltage

Signal level: Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output

Function: May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

Linearity0.2%
Load Regulation1.1%

Line Regulation: 0.02% /V ac Step Response Time: 2 - 3 seconds to 99% of the final value

ALARM OUTPUTS (if applicable)

2 Form "C" on/off relays. Configurable for latched and unlatched by software.

Max current: 5 AMPS, resistive load.

Max voltage250 V ac or 30 V dc

ISOLATED ANALOG OUTPUT (TB5, if applicable) Same as non-isolated analog output except isolated to 1000 Vdc.

Signal TypeCurrent or voltage

Signal Level: Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output

Function: May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

Linearity0.2%
Step Response Time2 - 3 seconds to 99% of the final value

Only one analog output is available on each unit and it must be factory installed.

Page 60 of the Omega DP26-RTD user manual — Figure 12-1 Meter Dimensions
Pagina 60 · klik om te vergroten
Pagina 60

96,0 (3.78) 48,0 (1.89)

FRONT BEZEL

20,3 (.80)

RETAINER

151,4 (5.96)

CASE

REAR COVER

SIDE VIEW TOP VIEW

PANEL THICKNESS

6,4 (.25) MAX 0,8 (.03) MIN

1,5 R(.06) 4 PLCS

45,00 + 0,61/-0,00 (1.772 + .024/–.000)

92,00 + 0,81/–0,00 (3.622 + .032/–.000)

Figure 12-1 Meter Dimensions

Page 61 of the Omega DP26-RTD user manual — Table 13-1 Factory Preset Values
Pagina 61 · klik om te vergroten
Pagina 61

SECTION 13. FACTORY PRESET VALUES

Table 13-1 Factory Preset Values

MENU ITEM FACTORY PRESET VALUES

INPT Input Type: 385.3 (3 wire DIN RTD) DEC.P Decimal Point Position: FFF.F RD.CF Reading Configuration: R.1=F (Fahrenheit) S1.CF Setpoint 1 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S.3=0 (Setpoint 1 is assigned to T1-T2) S2.CF Setpoint 2 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S.3=0 (Setpoint 2 is assigned to T1-T2) S1.DB Setpoint 1 Deadband: .0003 S2.DB Setpoint 2 Deadband: .0003 OT.CF Output Configuration: O.1=E (Analog output is enabled) O.2=C (Analog output is current) O.3=A (Analog output follows the display value) OUT Analog Output Routing: T1 - T2 OT.S.O Output Scale and Offset: 0-1000 = 4-20 mA dc T1.OF RTD for T1: 000.0 T2.OF RTD for T2: 000.0 SP1 Setpoint 1 Value: 000.0 SP2 Setpoint 2 Value: 000.0

Page 62 of the Omega DP26-RTD user manual — CE APPROVALS INFORMATION
Pagina 62 · klik om te vergroten
Pagina 62

CE APPROVALS INFORMATION

This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.

Electrical Safety EN61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory. Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min

  • Power to Input/Output: 2300 Vac (3250 Vdc)
  • Power to Input/Output: 500 Vac (720 Vdc) (Low Voltage dc Power Option*)
  • Power to Relays Output: 2300 Vac (3250 Vdc)
  • Relay 1 to Relay 2: 2300 Vac (3250 Vdc)
  • Isolated Analog to Inputs: 1000 Vac (1420 Vdc)
  • Analog to Inputs: No Isolation Measurement Category I

Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc. This unit should not be used in Measurement Categories II, III, IV. Transients Overvoltage Surge (1.2 / 50uS pulse)

  • Input Power: 2500 V
  • Input Power: 500 V (Low Voltage dc Power Option*)
  • Isolated Analog: 500 V
  • Input/Output Signals: 500 V Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation)

EMC EN61326:1997 + and A1:1998 + A2:2001 Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.

  • EMC Emissions Table 4, Class B of EN61326
  • EMC Immunity** Table 1 of EN61326 Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters

Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.

Page 63 of the Omega DP26-RTD user manual — WARRANTY/DISCLAIMER
Pagina 63 · klik om te vergroten
Pagina 63

WARRANTY/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.

If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS/INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.

FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product.

FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA: 1. Purchase Order number to cover the COST of the repair, 2. Model and serial number of product, and 3. Repair instructions and/or specific problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.

PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. USED UNDER LICENSE. Other US and International Patents pending or applied for.

Page 64 of the Omega DP26-RTD user manual — Where Do I Find Everything I Need for
Pagina 64 · klik om te vergroten
Pagina 64

Where Do I Find Everything I Need for

Process Measurement and Control?

OMEGA…Of Course! Shop on line at www.omega.com

TEMPERATURE Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers

PRESSURE, STRAIN AND FORCE Transducers & Strain Gauges Load Cells & Pressure Gauges Displacement Transducers Instrumentation & Accessories

FLOW/LEVEL Rotameters, Gas Mass Flowmeters & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers

pH/CONDUCTIVITY pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment

DATA ACQUISITION Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters

HEATERS Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters

ENVIRONMENTAL MONITORING AND CONTROL Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments

M1865/0206 11869ML-96 Rev.B

Bronnen

Referenties en links

Pagina's waarnaar wordt verwezen in deze handleiding — fabrikant, ondersteuning en downloads.

Vraag stellen over de Omega DP26-RTD

Eigenaren en ons team reageren doorgaans binnen een dag — en uw vraag helpt ook de volgende persoon.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Nog geen vragen over de Omega DP26-RTD

Stel als eerste een vraag — andere eigenaren of ons team geven antwoord.

Origineel document

Omega DP26-RTD handleiding-pdf

64 pagina's·1.2 MB·Thermometer·PDF
Klik op een bovenstaande pagina om deze hier te openen
Pagina 1 / …
100%
Pagina's laden…