Gebruiksaanwijzing /service van het product HH503 van de fabrikant Omega Vehicle Security
Ga naar pagina of 14
OMEGA HH503 Digital Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.
1 INTRODUCTION This instrument is a 4½ digit, compact-sized portable digital thermometer designed to use external K-type and J-type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K-type and J-type thermo- couples.
2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable Measurement Range: J-TYPE -200°C to 1050°C, (-328°F to 1922°F) K-TYPE -200°C to 1370°C, (-328°F to 2498°F) Resolution: 0.
ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.
4 2 1 3 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
5 Cold Junction Temperature Display mode (second display) Press C.J. TEMP key to display the thermocouple input connector cold junction temperature on second display. Press C.J. TEMP key again to exit this mode. K/J T1 Input Thermocouple Type Select The K/J key switch the thermocouple input selects the K-type or J-type thermocouple as input.
6 7 The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated value is retained as the average reading, but the actual minimum and maximum reading will continue to be captured.
7 8 9 Sec. Minu. Selecting the Time scale Reading the third displayed (Time) in either second or minute. When the thermometer turned on, it is set to second. To change the Time scale, press Sec. Minu. key. Maximum Time reading is 100 hours. If 100 hours is exceeded, reset Time to zero and restart.
8 10 11 12 4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits set value that was in use when thermometer was last turned off set values. Relative value Recall display mode. Press REL RECALL key to display the Relative set value on second display.
9 OPERA T OR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed.
1 HH 502/HH 503 CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter.
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warr ants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warr anty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermoc.
Een belangrijk punt na aankoop van elk apparaat Omega Vehicle Security HH503 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Omega Vehicle Security HH503 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Omega Vehicle Security HH503 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Omega Vehicle Security HH503 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Omega Vehicle Security HH503 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Omega Vehicle Security HH503 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Omega Vehicle Security HH503 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Omega Vehicle Security HH503 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.