omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info EP510, IP510, EP511, IP511 SERIES Electr oneumatic Conver ters Shop online at U ser ’ s Guide MADE IN.
Servicing Nor th America: U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047 ISO 90 01 Cer tif ied Stamford, CT 06907-0047 USA Toll Free: 1-800-826-6342 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.
A IR CONNECTIONS S UPPLY Connect air supply to 1/4 NPT port marked "IN." If the above specifications are not met, possibility of internal clogging exists.
M EP510 AND IP510 OUNTING GENERAL NEMA 1 and NEMA 4X transducers can be mounted in any position. DIRECT MOUNTING PIPE • Any NEMA 1 transducers may be supported by its own plumbing for air supply and output. NEMA 4X transducers may also be supported using 1/2" explosion proof conduit in the electrical port.
E LECTRICAL CONNECTIONS P RECAUTIONS TO BE OBSERVED DURING INSTALLATION These Transducers were tested at the EMC Test Centre, Dunfermline, Fife, KY11 5LB to the Electromagnetic Compatibility Directive effective January 1, 1996. The relevant EMC specifications tested were the following: EN 50081-1 (1992) and EN 50082-2 (1992).
WARNING: In explosive atmospheres, electrical power must be removed from the transducer before attempting to remove the cover and until the cover is fully reinstalled. Failure to do so may result in electrical spark or explosion. CAUTION: Only attempt electrical adjustments as described below.
R EPLACING ORIFICE: T his can be accomplished without removing the unit from its mounting or plumbing. 1. Turn off supply pressure and unscrew the brass orifice assembly located on the side of the housing with the gauge port. NOTE: Small sealing o-ring may remain inside of the housing.
EXPLOSION & DUST-IGNITION PROOF CERTIFICATIONS F ACTORY MUTUAL and CANADIAN STANDARDS ASSOCIATION WARNING: The following ratings are valid ONLY WHEN the cover is installed for FM and CSA. BOTH the cover and the locking device, using tamper resistant screws must be installed.
M- 410 9/1 10 4 I P5 10, EP 51 0, IP 511 , EP5 11 SE RIE S A 9 TABLE 5: KITS KIT NAME PART NUMBER 1. Panel Mounting Kit 010135-000 2. Valve Mounting Kit 010134-000 3. 2" Pipe Mounting Kit (Valve Mounting Kit is required) 010143-000 4. DIN Rail Adapter 010115-000 5.
M- 410 9/1 10 4 I P5 10, EP 51 0, IP5 11 , E P5 11 SE RIE S A 10 510 and 511 Series I/P AND E/P TRANSDUCER SPECIFICATIONS Accuracy: ± 0.10% of output span, typical ± 0.25% of output span, maximum Hysteresis: 0.01% of output span, typical 0.10% of output span, maximum Dead Band: Not measurable Repeatability: 0.
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months fro m da te of pu rcha se. OMEGA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M4109/0909 Where Do I Find Ever ything I Need for Pr ocess Measurement and Control? OM EGA…Of Course! Shop online at omega.com SM TEMPERA TURE 䡺 ⻬ Thermocouple, RTD & Thermistor Probes, Conn.
Een belangrijk punt na aankoop van elk apparaat Iomega EP510 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Iomega EP510 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Iomega EP510 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Iomega EP510 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Iomega EP510 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Iomega EP510 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Iomega EP510 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Iomega EP510 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.