MIL W AUKEE ELECTRIC T OOL CORPORA TION – TEST & MEASUREMEN T PRODUCT REP AIR SER VICE INSTRUCTIONS — CALIBRA TION 2 216- 20 Light Commercial Digital Multimeter Environmental Condition Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.
i T ABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 S pecifications 3 General S pecification 3 V oltage S pecif ication 4 Current S pecifications 5 Resist .
1 Introduction W arning T o avoid shock or injury , do not perform the verification test s or calibration procedu res described in the manual unless y ou are qualified to do so. The information provided in this document is for the use of qu alified personnel only .
2 SAFETY Review the following safety precaution s to avoid injury and prevent damag e to this product or any product s connected to it. T o avoid potential hazards, use the product only as spe cified. CAUTION: These st atement s identify conditions or practices that could result in damage to the equipment or other pro perty .
3 SPECIFICA TIONS All specifications are warranted unless noted typical and apply to the 2216 -20 DMM. S tated accuracies are at 23°C±5°C at tha n 80% relative humidity and without the battery indicator displayed.
4 Measurement Characteris tics Accuracy is ±(% reading + number of digit s) at 23°C ± 5°C, less than 80% R.H. T emperature coeffici ent: 0.1 * (S pecifi ed accuracy)/°C , < 18°C , > 28°C (1) DC V olt age Range Resolution Accuracy 600.0mV 0.
5 (3) DC Current Range Resolution Accuracy 6.000A 0.001A 10.00A 0.01A ± (1% reading + 3 digit s) Response Ti me: < 1.5 sec. Maximum measurement time: 5 minutes at 10A with at least 20 minutes rest time. Overload Protection: AC/D C 20A for 10 sec ond.
6 (6) Continuity Range Resolution Accuracy 600.0 Ω 0.1 Ω ± (1% reading + 5 digit s) Open Circuit V olt age: Approx. 0.6V Continuity: Built-in buzzer sounds whe n measured resistance i s less than 30 Ω and sounds off when measured resist ance is more than 200 Ω , Between 30 Ω to 200 Ω the buzzer maybe sound or of f either .
7 (1 1) Auto Po wer Off The meter au tomatically turns the power off after powering up and no operation for 20 ± 5 minutes. T o power up, turn the knob back to OFF f irst. (12) Measuring Rate: 3 t i mes / sec. (13) Fuse: 10A/600V Fast Action Fuse (1 4) Batt er y : Alkaline 1.
8 Physical and Environment al Characteri stics n o i t p i r c s e D s c i t s i r e t c a r a h C Dime nsio ns (H ×W ×D) 170mm × 80mm × 57m m g K 3 .
9 Required Equipment Required equipment is listed in T able B. If the recommended models are not available, equip ment with equivalent specifications may be used.
10 Basic Maintenance W arning T o avoid shock, remove the test leads an d any input signals before opening the case o r replacing the battery or fuses. Opening the Meter Ca se Caution T o avoid unintentional shock circuit, always place the uncovered Meter assembl y on a protective surface.
11 T esting Fuses T o test the internal fuses of the meter . 1. T urn the rotary selector s witch to the Ω position. 2. T o test FS1, plug a test lead into V Ω Hz input terminal, and touch the probe to the A input terminal. The displ ay should indicate between 0.
12 Performance T est s The following performance test s verify the co mplete operability of the Meter an d check the accuracy of each Meter function against the Meter’s specifications.
13 T esting the V olt age Function To verify accuracy in the AC and DC voltage ranges, do the following: 1. T urn t he ro t ar y switch to “ V ” position . 2. Connect the Calibrator to the V Ω and CO M inputs on the Meter. 3. Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1.
14 T esting the Resist ance Function T o verify the accuracy of the resist ance function, do the following: 1. Connect the Calibrator to V Ω Hz and CO M on t he Me ter . 2. T u rn t he ro tary swi tch to Ω . 3. Apply the inputs for st ep 1-6 in T able 3.
15 T esting the Ampere Function T o verify the accuracy of AC current measurement function s, do the following: 1. Connect the Calibrator to the A and CO M inpu t s on t he Me ter . 2. T u rn the ro ta ry swit ch to A~. 3. Apply the inputs for step s 1-4 in T able 5.
16 T able 7 Frequency T est: Step Source Level Waveform Readin g 1 8.0Vp-p 2.00Hz Square Wa ve 1.98Hz to 2.02Hz 2 8.0Vp-p 80.00Hz Square Wa ve 79.90Hz to 80.10Hz 3 8.0Vp-p 800.0Hz Square Wa ve 799.0Hz to 801.0Hz 4 8.0Vp-p 8.000kHz Square Wa ve 7.990kHz to 8.
17 Calibration Procedure Recalibrate the meter: It is recommended that the meter may be calibrated once year . Use the following procedure to calibrate the meter . Envir on ment a l C ondi t ion 1. Perform calibration at an ambient temperature and relative humidity (23°C±2°C a nd R.
18 (E) DCA Calibration (V R7) 1. Rotate the rotary switch to DCA Function. 2. Connect the output of a DCA calibrator to V and COM input terminals of the meter. 3. Output DCA 5.000A from the calibrator. 4. Adjust VR7 until the display shows 5.000A ± 0.
Een belangrijk punt na aankoop van elk apparaat Milwaukee 2216-20 (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Milwaukee 2216-20 heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Milwaukee 2216-20 vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Milwaukee 2216-20 leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Milwaukee 2216-20 krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Milwaukee 2216-20 bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Milwaukee 2216-20 kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Milwaukee 2216-20 . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.