T200U-NOy ANALYZER ADDENDUM (for use with T200 Operators Manual, PN 06858 and with the T200U Addendum, PN 06861) © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.
.
07303C DCN6646 i SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a saf ety alert symbol, and placed throughout this manual and inside the instrument.
ii 07303C DCN6646 CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes.
07303C DCN6646 iii WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAP I), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipm ent is thoroughly inspected and tested. Should equipment failure occur, TAPI assures its customers that prom pt service and support will be available.
Model 200EU-NOy Opti on Manual Addendum iv 07303C DCN6646 This page intentionally left blank..
Model 200EU-NOy Opti on Manual Addendum 07303C DCN6646 v TABLE OF CONTENTS 1.0 INTR ODUCTION ...................................................................................................... 7 2.0 SPECIF ICATIONS .................................
Model 200EU-NOy Opti on Manual Addendum vi 07303C DCN6646 List of Figures Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom) ...................... 13 Figure 2. Rear Panel Pneumatic Connecti ons .....................
07303C DCN6646 7 1.0 INTRODUCTION The T200U-NOy analyzer consists of an external Converter designed to support a NOx analyzer by converting multiple , unstable compounds grouped under the name NOy .
Model T200EU-NOy Opti on Manual Addendum 1.0 Int roduction 8 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 9 2.0 SPECIFICATIONS 2.1 Specifications Table 1 Specifications for the External Converter and the 501Y Bypass Pump Converter Remotely mounted molybdenum external Converter with temperature controller in a NEMA – 4 enclosure. Converter Temperature 315 o C 7 o C with read-out on front panel of bypass pump chassis.
Model T200EU-NOy Option M anual Addendum 2.0 Specifi cations 10 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 11 3.0 GETTING STARTED The NOy converter has been designed to operate with the T200U low level NOx analyzer, with some additional modification.
Model T200U-NOy Opti on Manual Addendum 12 07303C DCN6646 3.2 Electrical and Pneumatic Connections NOTE To maintain compliance with EMC standards, it is required that the cable length (at the rear of .
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 13 Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom).
Model T200U-NOy Opti on Manual Addendum 14 07303C DCN6646 Figure 2. Rear Panel Pneumatic Connections Note: Pneumatic tubes and rear panel labels incl ude numbers to facilitate m atching connections between instruments.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 15 3.3 Initial Operation 1. After confirming proper supply volta ge, turn on the power to the analyzer and the 501Y bypass pump, and plug in the external Pum p.
Model T200U-NOy Opti on Manual Addendum 16 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 17 4.0 THE T200U-NOY EXTERNAL CONVERTER 4.1 Principle of Operation The T-API Model T200U-NOy is designed to measure the concentration of NO, NO 2 , and other compounds that are too unsta ble to be measured when taken in through the normal ambient air sam ple inlet system.
Model T200U-NOy Opti on Manual Addendum 18 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 19 5.0 CALIBRATION AND CAL CHECKS Unlike most NOx analyzers, the T200U-NOy does not have a sample inlet port on its rear panel. The sample inlet is located on the external Converter. Follow the steps in the T200U manual to: Enter the expected NO and NOy span gas concentrations in the analyzer.
Model T200U-NOy Opti on Manual Addendum 20 07303C DCN6646 5.1 Calibration or Cal Check Procedure Calibration of the analyzer can be check ed or adjusted using gas introduced through the CAL IN (3) port on the 501Y rear panel.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 21 Table 5 Span Calibration Procedure Step Action Comment 1. Press CAL The analyzer enters the calibrate mode. 2. Wait For Stability Type STB < 2 PPB Wait for reading to stabilize at span value.
Model T200U-NOy Opti on Manual Addendum 22 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 23 6.0 MAINTENANCE 6.1 Maintenance Schedule The schedule for preventative maintenance is presented in Table 6 below. Qualified Maintenance Personnel Required Table 6 Preventative Mainte.
Model T200U-NOy Opti on Manual Addendum 24 07303C DCN6646 6.2 Replacing the Sample Particulate Filters The particulate filter should be inspected often for signs of plugging or contamination. It is also comm on for dirt particles to cause instrument drift, and affect accuracy.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 25 Figure 3. Replacing the Particulate Filter.
Model T200U-NOy Opti on Manual Addendum 26 07303C DCN6646 6.3 Checking Analyzer Flow Rate The external Pump is capable of maintain ing the reaction cell pressure at less than 5.0 In-Hg-A. If a higher pressure is noted, the Pump may need servicing. Check the Pump and pneumatic system for leaks or rebuild the Pum p.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 27 8. Turn the 501Y power back on. The in sulation can emit a burnt odor for the first 24 hours; this is normal. Allow t he Converter to burn-in for 24 hours, then re-calibrate the instrument. Figure 4.
Model T200U-NOy Opti on Manual Addendum 28 07303C DCN6646 Figure 5. Pneumatics.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 29 6.6 Checking for Leaks If a leak checker is not available, it is po ssible to leak check the instrument using the sample pump plus a shut-off valve. 1. Turn off power to the analyzer and the 501Y, and unplug the external Pump.
Model T200U-NOy Opti on Manual Addendum 30 07303C DCN6646 If you have a leak checker: 1. Turn off power to the analyzer and 501Y, and unplug the external Pump.
07303C DCN6646 31 7.0 TROUBLESHOOTING, ADJUSTMENTS CAUTION! The operations outlined in this section must be performed by qualified maintenance personnel only! This section of the manual contains in formation on diagnosing and repairing instrument performance problem s.
Model T200U-NOy Opti on Manual Addendum 32 07303C DCN6646 7.1.3 501Y Bypass Pump Diagnostic Procedures The 501Y bypass pump is capable of maintaining a total bypass flow rate of ~2.2 lpm. If flow checks indicate a fall off in bypass flow, the pump may need servicing.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 33 Indications that would require re-programming are: Failure / replacement of controller. Replacement of thermocouple elem ent. Controller’s loss of PID parameters leading to instability of the tem p setpoint.
Model T200U-NOy Opti on Manual Addendum 34 07303C DCN6646 SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL hold until p-n1 SET TO 0 – REVERSE ACTION SEL hold until .
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 35 Figure 7 Electrical Diagram.
Model T200U-NOy Opti on Manual Addendum 36 07303C DCN6646 This page intentionally left blank..
07303C DCN6646 37 8.0 T200U-NOY SPARE PARTS LIST The following spare parts list supplements the spare parts list for the T200U analyzer. Note Use of replacement parts other than those s upplied by API m ay result in non-compliance with European Standard EN 61010-1.
Model T200U-NOy Opti on Manual Addendum 38 07303C DCN6646 Reference: 06848 11/8/2010 1:13:10 PM PARTNUMBER DESCRIPTION 041800600 PCA, PMT PREAMP, VR, 041920000 ASSY, THERMISTOR 042680100 ASSY, VALVE (.
Model T200U-NOy Opti on Manual Addendum 07303C DCN6646 39 Reference: 06848 11/8/2010 1:13:10 PM PARTNUMBER DESCRIPTION CN0000073 POWER ENTRY, 120/60 (KB) FL0000001 FILTER, SS (KB) FL0000003 FILTER, DF.
Model T200U-NOy Opti on Manual Addendum 40 07303C DCN6646 This page intentionally left blank..
Een belangrijk punt na aankoop van elk apparaat Teledyne T200U-NOy (of zelfs voordat je het koopt) is om de handleiding te lezen. Dit moeten wij doen vanwege een paar simpele redenen:
Als u nog geen Teledyne T200U-NOy heb gekocht dan nu is een goed moment om kennis te maken met de basisgegevens van het product. Eerst kijk dan naar de eerste pagina\'s van de handleiding, die je hierboven vindt. Je moet daar de belangrijkste technische gegevens Teledyne T200U-NOy vinden. Op dit manier kan je controleren of het apparaat aan jouw behoeften voldoet. Op de volgende pagina's van de handleiding Teledyne T200U-NOy leer je over alle kenmerken van het product en krijg je informatie over de werking. De informatie die je over Teledyne T200U-NOy krijgt, zal je zeker helpen om een besluit over de aankoop te nemen.
In een situatie waarin je al een beziter van Teledyne T200U-NOy bent, maar toch heb je de instructies niet gelezen, moet je het doen voor de hierboven beschreven redenen. Je zult dan weten of je goed de alle beschikbare functies heb gebruikt, en of je fouten heb gemaakt die het leven van de Teledyne T200U-NOy kunnen verkorten.
Maar de belangrijkste taak van de handleiding is om de gebruiker bij het oplossen van problemen te helpen met Teledyne T200U-NOy . Bijna altijd, zal je daar het vinden Troubleshooting met de meest voorkomende storingen en defecten #MANUAl# samen met de instructies over hun opplosinge. Zelfs als je zelf niet kan om het probleem op te lossen, zal de instructie je de weg wijzen naar verdere andere procedure, bijv. door contact met de klantenservice of het dichtstbijzijnde servicecentrum.