Page 2 of the Geratherm GP-6621 - desktop user manual — Precautions
Pagina 2 · klik om te vergroten
Pagina 2

Precautions

  • This manual and the product are not substitutes for

visiting the doctor. Neither the information contained herein nor this prod-

uct may be used to diagnose or treat health problems, or to prescribe drugs. If you have or suspect that you have a medical problem, please seek immediate ad- vice from your doctor. • Do not conduct any measurements if the temperature

is low (below +5 °C) or high (over +40 °C), or if the relative humidity is beyond the range of 15 % to 90 %, as this can lead to inaccurate readings. • Wait 30 to 45 minutes, before taking a measurement if

you have just had a caffeinated drink or a cigarette. • Relax for at least 5 to 10 minutes before taking a

measurement. • Please wait 3 to 5 minutes between measurements,

so that your blood vessels can return to the state they were in prior to measurement. You may have to adapt the waiting time to your personal physiology. • It is recommended that you use the same arm for each

measurement (preferably the left) and take the meas- urement at about the same time every day. • Sit down comfortably with your elbows placed on the

table and both feet on the ground. Please do not inter- lock your legs during the measurement. • Place the unit at the level of the heart. Relax your

hand. Your palm should be facing up. • Take the measurement at room temperature in a quiet

and stress-free environment. • The unit should not be moved or shaken during the

measurement. Please do not speak during the meas- urement. • Please keep in mind that blood pressure naturally

varies depending on the time of day and is affected by many different factors. Blood pressure is usually highest at work and reaches its lowest level during the sleep phase. • Blood pressure measurements should be assessed

by a doctor or trained healthcare professional who is familiar with your medical history. If you use the unit and regularly record the results, please keep your doctor informed with regard to the ongoing changes in your blood pressure. • If you suffer from a cardiovascular disease (such as

atherosclerosis), diabetes, a liver or kidney disease,

Page 3 of the Geratherm GP-6621 - desktop user manual — Precautions
Pagina 3 · klik om te vergroten
Pagina 3

Precautions

severe hypertension or peripheral circulatory disor- ders, etc., please consult your doctor before using this unit. • The performance of this device can be infl uenced as

severe arrhythmias such as atrial or ventricular pre- mature beats or atrial fi brillation are presented during measurement. • The blood pressure measurements conducted with

this unit are equivalent to measurements obtained by a trained observer in accordance with the values achieved using the cuff/stethoscope auscultation method and are within the specifi ed DIN EN ISO 81060-2 standard limits. • If the cuff causes any discomfort during the measure-

ment, press the “POWER” button to turn off the unit immediately. • If the pressure is over 300 mmHg and the cuff does

not defl ate automatically, pull off the Velcro strap to detach the cuff. • Do not use this appliance on infants, children or per-

sons who cannot express their own intentions. • To avoid accidental strangulation, keep the product

away from children and do not place the hose around the neck. • Measuring too frequently may result in circulatory dis-

orders, which can cause unpleasant sensations such as localised bleeding under the skin or temporary numbness in your arm. These symptoms do not usu- ally last long. However, if you have not recovered after some time, please consult your doctor. • Please take into consideration the electromagnetic

compatibility of the unit (e.g. disruptions to the power supply, radio frequency interference, etc.) see annex. Please only use the unit indoors. To avoid inaccurate results due to electromagnetic interference between electrical and electronic equipment, please do not use the unit near mobile phones or microwave ovens. In the case of devices whose maximum power exceeds 2 W, the minimum distance from your blood pressure monitor should be 3.3 metres. • The unit is not waterproof. Never immerse this instru-

ment in any liquids. • Do not use the instrument if you think it is damaged or

if you notice anything unusual.

Page 4 of the Geratherm GP-6621 - desktop user manual — Information you should know before operating the unit
Pagina 4 · klik om te vergroten
Pagina 4

Warranty

The warranty for this blood pressure monitor is valid for any error on the part of the manufacturer under normal use for 3 years from the date of purchase. If your unit does not function properly due to defective parts or assembly, we will repair it free of charge. With the exception of the battery and cuff, all parts of the unit are subject to this warranty. Damage caused by im- proper handling of your unit is not guaranteed. We recommend that the accuracy of the unit be checked after 2 years from manufacturing date by an authorized laboratory. This checking procedure is not a service provided under

the warranty.

Information you should know before operating the unit

What is blood pressure? A force is created by the heart as the ventricle forcibly ejects blood into the blood vessels and through the trans- port system. Another force is created by the arteries as they resist the blood fl ow. Blood pressure is the result of these two forces.

What are systolic and diastolic pressures? Systolic pressure represents the highest pressure at the height of the heart‘s contraction. Diastolic pressure is the lowest pressure when the heart is resting.

Is my blood pressure normal? See the following blood pressure classifi cation chart re- leased by the WHO (World Health Organization) for evalu- ation of your blood pressure level.

Severe hypertension (Grade 3)

Moderate Hypertension (Grade 2)

Mild Hypertension (Grade 1)

Show 5 more labels / short text on this page
Diastolic mm Hg · Systolic mm Hg · High - normal · Normal · Optimal
Page 5 of the Geratherm GP-6621 - desktop user manual — About the unit
Pagina 5 · klik om te vergroten
Pagina 5

Information you should know before operating the unit

What about low blood pressure? In general, a lower blood pressure reading is better unless it causes some uncomfortable symptoms such as fainting and/or lightheadedness.

Fluctuation and variation in blood pressure The following chart shows possible blood pressure fl uctua-

tions during a 24hr period.

Geratherm GP-6621 - desktop — Information you should know before operating the unit diagram (page 5)
Afb. 1 · Information you should know before operating the uni…

The following factors will infl uence blood pressure meas- urement results and cause variations.

  • Bathing
  • Breathing
  • Conversation
  • Drinking alcohol
  • Exercise
  • Moving
  • Mental tension
  • Meals
  • Temperature change
  • Thoughts
  • Smoking etc.

About the unit

Geratherm GP-6621 - desktop — About the unit diagram (page 5)
Afb. 2 · About the unit
Show 5 more labels / short text on this page
mmHg · Explanation of Display · Memory · Detection · Pulse
Page 6 of the Geratherm GP-6621 - desktop user manual — About the unit
Pagina 6 · klik om te vergroten
Pagina 6
Geratherm GP-6621 - desktop — About the unit diagram (page 6)
Afb. 3 · About the unit

The cuff is designed to fi t arm sizes between 22 and 32cm (8.7 to 12.6 inches.)

Geratherm GP-6621 - desktop — About the unit diagram (page 6)
Afb. 4 · About the unit
1. Insert the batteries into the battery compartment matching correct polarities „+“ and „-“. 2. Replace all batteries if the low battery indicator appears. 3. Remove the batteries if the unit will not be usedfor a long period of time. 4. REMOVE all of the BATTERIES when utilizing the optional AC adapter.

Keep batteries away from small children. Do not throw batteries into fi res: they could explode.

Page 7 of the Geratherm GP-6621 - desktop user manual — Preparation for Use
Pagina 7 · klik om te vergroten
Pagina 7

Preparation for Use

Geratherm GP-6621 - desktop — Preparation for Use diagram (page 7)
Afb. 5 · Preparation for Use
1. Wrap the cuff around the left arm. The arm should be bare skin.
2. Fasten the cuff. Don‘t pull it strongly and make the cuff too tight. The cuff‘s edge should be approximately 1 inch from crook of the arm.
3. Attach the cuff on the right hand as shown in the fi gure if it is not possible to measure on the left arm.

Measuring Posture 1. Sit upright and ensure that the measuring site is at heart level. Relax and measure in a natural position.

2. Measure and record blood pressure at the same time every day to establish your blood pressure pattern.

AC Adapter (accessory) Use the device only with a medical approved stabi- lized AC adapter (Input: 100~240 V, AC, 60/50Hz; Output: 6V, DC, 800mA).

2. If AC adapter power is interrupted during measurement, the device must be reset by disconnecting the AC adapter from the device.
3. Only use medical approved adapters that comply with the specifi cations in this manual. Using other adapters could cause damage your blood pressure monitor.
Page 8 of the Geratherm GP-6621 - desktop user manual — How to Operate the Unit
Pagina 8 · klik om te vergroten
Pagina 8

How to Operate the Unit

1.Setting the Clock

To enter the setting mode by press- ing and holding the memory key for over 3 seconds. The display will show a blinking year. Press the power key to adjust the year when it is blinking. Then press and release the memory key again, the “month” will appear and blink to be adjusted, and then “date”, “hour” and “minute” will fol- low up as you press and release the memory key again.

Geratherm GP-6621 - desktop — How to Operate the Unit diagram (page 8)
Afb. 6 · How to Operate the Unit

To adjust the clock Press the memory key to adjust the values of the ”month”, “date”, “hour”, and “minute” when they are blinking.

Geratherm GP-6621 - desktop — How to Operate the Unit diagram (page 8)
Afb. 7 · How to Operate the Unit
2. Measuring Blood Pressure

Press and release the POWER key to start measuring automatically

ã Measuring and Buzzer sounds.

When the measurement is fi nished, the display will show the measured blood pressure and pulse value. Press the power key to turn off the unit. Otherwise the unit will turn off automatically 150 seconds later.

ã Initial display · ã Pumping

Page 9 of the Geratherm GP-6621 - desktop user manual — How to Operate the Unit
Pagina 9 · klik om te vergroten
Pagina 9

How to Operate the Unit

Geratherm GP-6621 - desktop — How to Operate the Unit diagram (page 9)
Afb. 8 · How to Operate the Unit
3. Recalling Memories

Press and release the MEMORY key to recall the last read- ing (Example: the no.7 memory). The display will show the memorized pressure and pulse value.

Geratherm GP-6621 - desktop — How to Operate the Unit diagram (page 9)
Afb. 9 · How to Operate the Unit

Press and release the MEMORY key again to recall the no.6 memory. Press and release repeatedly to read the no 5, 4, 3, 2, 1 memorized pressures.

Page 10 of the Geratherm GP-6621 - desktop user manual — How to Operate the Unit
Pagina 10 · klik om te vergroten
Pagina 10

How to Operate the Unit

4. Erasing Memories

You may erase one or all records in memory.

Delete one record 1. Press and release MEMORY key to choose the record that you wish to delete

2. Press and hold the POWER key until the display shows “dEL”.
3. Press and release the power key again. After 3 beeps are heard, the unit erases the record.

Delete all records 1. Press and release MEMORY key to choose the record that you wish to delete.

2. Press and hold the POWER key until the display shows “dEL”.
3. Press and release the MEMORY key then the display shows “dEL ALL”.
4. Press and release the POWER key. After 3 beeps are heard, the unit erases all records.
Page 11 of the Geratherm GP-6621 - desktop user manual — Care and Maintenance
Pagina 11 · klik om te vergroten
Pagina 11

Care and Maintenance

Geratherm GP-6621 - desktop — Care and Maintenance diagram (page 11)
Afb. 10 · Care and Maintenance

Do not drop the unit. It is not shock-proof.

Do not modify or disassemble the unit or the arm cuff.

Use a cloth mois- tened with water or neutral detergent to clean the body of the unit and then wipe it dry.

Page 12 of the Geratherm GP-6621 - desktop user manual — Care and Maintenance
Pagina 12 · klik om te vergroten
Pagina 12

Care and Maintenance

Geratherm GP-6621 - desktop — Care and Maintenance diagram (page 12)
Afb. 11 · Care and Maintenance

Keep the unit in a suitable place. Avoid high tempera- ture, direct sunlight, high moisture, and dust.

Remove the batteries if the unit will not be used for a long time.

Do not press the POWER button if the cuff has not been properly wrapped around the arm.

Page 13 of the Geratherm GP-6621 - desktop user manual — Disposal
Pagina 13 · klik om te vergroten
Pagina 13

Error Messages

Geratherm GP-6621 - desktop — Error Messages diagram (page 13)
Afb. 12 · Error Messages

The pressure measured was lower than 20 mm Hg. Please measure again.

Excessive body movement is detected during measurement. Please measure again.

The pressure can not be measured due to signal noise. Please measure again.

The pumping pressure is higher than 300 mm Hg. Please measure again.

Low battery. Check and replace the 4 batteries if necessary.

Disposal

Observe the applicable regulations when disposing of the device and batteries. This product must not be disposed of together with domes-

tic waste. All users are obliged to hand in all electrical or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or commercial collection point so that they can be disposed of in an environmentally ac- ceptable manner. Please remove the batteries before disposing of the de- vice/unit. Do not dispose of old batteries with your house- hold waste, but at a battery collection station at a recycling site or in a shop.

Geratherm GP-6621 - desktop — Disposal diagram (page 13)
Afb. 13 · Disposal
Page 14 of the Geratherm GP-6621 - desktop user manual — Specifi cations
Pagina 14 · klik om te vergroten
Pagina 14

Specifi cations

Model no. GP-6621

Display System Liquid Crystal Display

Power Source 4 alkaline “AA“ type batteries (1.5 V) or AC adapter (input: 100~240 V, AC, 60/50 Hz; output: 6 V, DC, 800 mA)

Measuring Range Blood pressure20 to 300 mmHg Heart pulse rate: 40 to 200 pulse beats/min
Accuracy Blood pressureHeart pulse rate: ±3 mm Hg
pulse rate±5 % Infl ation

Pressure Exhaust Electric solenoid valve

Low Battery Indicator Yes

Auto Power-Off 150 secs. without any operation

Battery Life Around 250 measurements

Protection Level IP20 (protected against

foreign bodies < 12,5 mm)

Cuff Size 22 to 32 cm / 8.7 - 12.6 inches

Operating Environment +5 °C to +40 °C; RH 15 % to 90 % (+41 °F to +104 °F)

Storage Environment -25 °C to +55 °C; RH 15 % to 90 % (-4 °F to +122 °F)

Dimensions 171 mm x 139 mm x 54 mm

Weight around 550 g (including batteries)

Specifi cations are subject to change for

improvement without notice.

Show 5 more labels / short text on this page
Measuring Method Oscillometric · Micro rolling pump · Memory 85 sets · Display LCD (year/date/time, · pressure and pulse)
Page 15 of the Geratherm GP-6621 - desktop user manual — -25°C
Pagina 15 · klik om te vergroten
Pagina 15

Quality Standard

Geratherm® is certifi ed in accordance with Council Direc- tive 93/42/EEC and EN ISO 13485 and is entitled to affi x the CE-mark 0197 (Notifi ed Body: TÜV Rheinland LGA Products GmbH).

The blood pressure monitor conforms to

  • EN 60601-1 +A1 Medical electrical equipment. Part 1:

General requirements for basic safety and essential performance. • IEC/EN 60601-1-11 Medical electrical equipment. Part

1-11: General requirements for basic safety and essen- tial performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. • DIN EN ISO 81060-1 Non-invasive sphygmomanometers – Part 1: Require-

ments and test methods for non-automated measure- ment type (ISO 81060-1) • DIN EN ISO 81060-2 Non-invasive sphygmomanometers – Part 2: Clinical

investigation of automated measurement type (ISO 81060-2) • EN 1060-3 Non-invasive sphygmomanometers – Part 3: Supple-

mentary requirements for electro-mechanical blood pressure measuring systems

Symbol index

Follow the instruc- tions for use Type BF Equipment

Keep dry Batch code (mm/yyyy; month/year)

Store between 15 % and 90 % R.H. Serial number

SN

+55°C

Store between -25 °C and +55 °C Manufacturer

-25°C

The device must not be disposed of with household waste.

90% · 15%

Page 16 of the Geratherm GP-6621 - desktop user manual — Annex
Pagina 16 · klik om te vergroten
Pagina 16

Annex

Information on electromagnetic compatibility (EMC)

Electronic devices such as PCs and mobile phones can lead to the exposure of medical devices in operation to electromagnetic interference from other devices. This can lead to malfunction of the medical device and create a po- tentially unsafe situation. Medical devices should also not interfere with any other devices.

The EN 60601-1-2 standard regulates the requirements for

EMC (electromagnetic compatibility) and defi nes the levels of immunity to electromagnetic interference and the maxi- mum electromagnetic emission levels for medical devices.

This blood pressure monitor, which is manufactured by Geratherm Medical AG, complies with the EN 60601-1-2 standard in relation to both immunity and emissions. However, special precautions should be observed: please only use the device indoors and not in the vicinity of mobile phones or microwave ovens. In the case of de- vices whose maximum power exceeds 2 W, the minimum distance from your blood pressure monitor should be 3.3 metres.

Page 17 of the Geratherm GP-6621 - desktop user manual — Annex
Pagina 17 · klik om te vergroten
Pagina 17

Annex

Guidance and manufacturer´s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environments listed below, and should only beused in such environments:

Emissions test Compliance Electromagnetic environment - guidance

RF energy is used only to maintain device’s operation. Therefore, its RF emissions are so low that it’s not likely to cause any interfer- ence in nearby electronic equipment.

RF emissions CISPR 11 Group 1

RF emissions CISPR 11 Class B The device is suitable for use in all establish- ments, including domesticestablishments, and those directly connected to the public low-voltage power supply networkthat sup- plies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2 Class A

Voltage fl uctuations / fl icker emissions IEC 61000-3-3 complies

Guidance and manufacturer´s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environments listed below, and should only beused in such environments:

Immunity test IEC 60601 – test level

Electromagnetic environment - guidance

Electrostatic discharge (ESD) IEC 61000-4-2

± 6 kV contact ± 8 kV air

± 6 kV contact ± 8 kV air

Floors should be wood, concrete or ceramic tile. If fl oors are covered with synthetic material, the relative humidity should be at least 30 %.

Power frequency (50/60 Hz) magnetic fi eld IEC 61000-4-8

Power frequency magnetic fi elds should be at levels characteristic of a typical location in a typicalcommer- cial or hospital environment.

3 A/m 3 A/m

Recommended separation distances between portable and mobile RF communication

The device is intended for use in an electromagnetic environment where radiated RF disturbancesare under control. User can help prevent electromagnetic interference by keeping the device at a minimum distance from portable and mobile RF communications equipment (transmitters). Below table details the maximum output power of transmitter:

Rated maximum output power of transmitters in Watt

Seperation distance according to frequency of transmitter / m

150 kHz to 80 MHz d = 1.2 √P

80 MHz to 800 MHz d = 1.2 √P

800 MHz to 2.5 GHz d = 2.3 √P

0.01 0.1 1 10 100

0.12 0.38 1.2 3.8 12

0.12 0.38 1.2 3.8 12

0.23 0.73 2.3 7.3 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmit- ter in watts (W) according to the transmitter manufacturer.

Compliance level

Page 18 of the Geratherm GP-6621 - desktop user manual — Annex
Pagina 18 · klik om te vergroten
Pagina 18

Annex

Guidance and manufacturer´s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments:

IEC 60601 – test level

Compliance level Electromagnetic environment - guidance

Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: Conducted RF IEC 61000-4-6

3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.2 √P

Radiated RF IEC 61000-4-3

3 V/m 80 MHz to 2.5 GHz 3 V/m d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recom- mended separation distance in metres (m).

Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity ofequipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and refl ection from structures, objects and people.

Guidance and manufacturer´s declaration – electromagnetic immunity a) Field strengths from fi xed transmitters, such as base stations for radio (cellular/cord- less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be considered. If the measured fi eld strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.

b) Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.

Geratherm Medical AG Fahrenheitstrasse 1 98716 Geschwenda Germany www.geratherm.com

Immunity test

Page 19 of the Geratherm GP-6621 - desktop user manual — Annex
Pagina 19 · klik om te vergroten
Pagina 19
Page 19 of the Geratherm GP-6621 - desktop user manual — Annex
Page 20 of the Geratherm GP-6621 - desktop user manual — FR
Pagina 20 · klik om te vergroten
Pagina 20

Tensiomètre numérique à écran

de contrôle

Geratherm GP-6621 - desktop — Tensiomètre numérique à écran diagram (page 20)
Afb. 14 · Tensiomètre numérique à écran

FR

MODE D’EMPLOI

Bronnen

Referenties en links

Pagina's waarnaar wordt verwezen in deze handleiding — fabrikant, ondersteuning en downloads.

Vraag stellen over de Geratherm GP-6621 - desktop

Eigenaren en ons team reageren doorgaans binnen een dag — en uw vraag helpt ook de volgende persoon.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Nog geen vragen over de Geratherm GP-6621 - desktop

Stel als eerste een vraag — andere eigenaren of ons team geven antwoord.

Origineel document

Geratherm GP-6621 - desktop handleiding-pdf

164 pagina's·2.8 MB·Bloeddrukmeter·PDF
Klik op een bovenstaande pagina om deze hier te openen
Pagina 1 / …
100%
Pagina's laden…