This unit uses an oscillometric measurement method in order to measure systolic and diastolic blood pressure, as well as the heart rate. The measurement is conducted on the upper arm. All val- ues can be read on an LCD screen. This blood pressure monitor complies with the standards for medical devices. The blood pressure monitor can be used by adults with an arm circumference of 22 cm to 42 cm.
This product uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a “zero pressure” equivalent to the air pressure. Then it starts inflating the cuff, meanwhile, the unit detects pressure oscillations generated by beat- to-beat pulsatile, which is used to determine the systolic and diastolic pressure, and also pulse rate.
The device compares the longest and the shortest time in- tervals of detected pulse waves to mean time interval then calculates standard deviation. Along with the blood pres- sure reading, the device also shows a symbol (arrhythmia icon) if the heart beat is irregular. (> 25 %)

consulting a physician. Neither the information con- tained herein nor this product may be used to diag- nose or treat health problems, or to prescribe drugs. If you have or suspect that you have a medical problem, please seek immediate advice from your physician. • Please ask your physician before using the device
on pregnant women, patients with implanted, electri- cal devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mas- tectomy. Please consult your physician prior to using the unit if you suffer from any of these illnesses. • Wait 30 to 45 minutes, before taking a measurement if
you have just had a caffeinated drink or a cigarette. • Relax for at least 5 to 10 minutes before taking a meas-
urement. • Do not take any therapeutic measures on the basis of
a self-measurement. Never alter the dose of a medi- cine prescribed by a physician. Consult your physician if you have any questions about your blood pressure. • Please keep the unit out of reach of infants, children
or pets, since inhalation or swallowing of small parts is dangerous or even fatal. • This device is intended for non-invasive measuring
and monitoring of arterial blood pressure. It is not intended for use on extremities other than arm or for functions other than obtaining a blood pressure meas- urement. • If you experience discomfort during a measurement,
such as pain in the arm or other complaints, press the “START” button to release the air immediately from the
cuff. Loosen the cuff and remove it from your arm. • On the rare occasion of a fault causing the cuff to re-
main fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure • >300 mmHg or constant pressure >15 mmHg for more
than 3 minutes) applied to the arm may lead to an ec- chymosis. • Too frequent and consecutive measurements could
cause disturbances in blood circulation and injuries.

devices are attached, as this could lead to temporary functional failure of the medical devices being used. • Avoid any kinks in the air hose during use, as this
could lead to a continuous increase in the cuff pres- sure and cause injury to the patient. • This unit is not suitable for continuous monitoring dur-
ing medical emergencies or surgery. This unit cannot be used simultaneously with high-frequency surgical devices. • This unit is not intended for the transportation of pa-
tients outside medical establishments. • In order to avoid inaccurate measurements, do not
expose the blood pressure monitor to strong electro- magnetic fields. • The material of the cuff has been tested and meets the
polyester or synthetic materials. • In order to avoid accidental strangulation, keep the de-
vice away from children and do not lay the tube around the neck. • Do not connect the air hose to any other medical de-
vices, as this can lead to high pressure or conduct air into intravascular systems, which may lead to injury. • Before using the unit for the first time, ensure that there
is no visible damage to the unit. • Please operate the unit under the environmental condi-
tions described in the instructions. Otherwise, the per- formance and service life of the blood pressure monitor will be adversely affected. • Please use only accessories that have been approved
by the manufacturer. Otherwise, damage may be caused to the unit, the user may suffer injury or inac- curate measurements may occur. • Dispose of the unit, accessories and components in
accordance with local regulations.

unit yourself. Only permit repairs to be carried out by authorised technicians. • In the event of unexpected error messages, please
contact your dealer. • For cleaning, please use a soft cloth and a solvent-free
cleaning agent.
The warranty for this blood pressure monitor is valid for
any error on the part of the manufacturer under normal use for 2 years from the date of purchase. If your unit does not function properly due to defective parts or as- sembly, we will repair it free of charge. With the exception of the battery and cuff, all parts of the unit are subject to this warranty. Damage caused by im- proper handling of your unit is not guaranteed.

What is blood pressure? A force is created by the heart as the ventricle forcibly ejects blood into the blood vessels and through the vas- cular system. Another force is created by the arteries as they resist the blood flow. Blood pressure is the result of these two forces
Is my blood pressure normal? See the following blood pressure classification chart re- leased by the WHO (World Health Organization) for eva- luation of your blood pressure level.
110 Severe Hypertension (Grade 3)
Moderate Hypertension (Grade 2)
Mild Hypertension (Grade 1)
Systolic mmHg 130 180 170 160 150 140 120
Classification of blood pressure monitor
Optimal < 120 < 80 green
Normal 120 - 129 80 - 84 green
High - Normal 130 - 139 85 - 89 green
Grade 1 Hypertension 140 - 159 90 - 99 yellow
Grade 2 Hypertension 160 - 179 100 - 109 orange
Grade 3 Hypertension >= 180 >= 110 red

What are systolic and diastolic pressures? Systolic pressure is the highest pressure at the height of the heart’s contraction. Diastolic pressure is the lowest pressure when the heart is resting.
What about low blood pressure? In general, a lower blood pressure reading is better un- less it causes unpleasant symptoms such as fainting and/ or lightheadedness.
Fluctuation and variation in blood pressure The following chart shows possible blood pressure fluctu-
ations during a 24-hour period.
The following factors will influence blood pressure mea-
surement results and cause variations:
mmHg · Systolic · Diastolic

An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, this equipment records the heartbeat intervals and works out the standard deviation. If major pulse irregularities occur, the arrhythmia icon appears on the display along with the blood pressure readings.
pulse irregularity consistent with an irregular heart-beat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.

Information about the Device

Content: 1. Blood pressure monitor (GT-1115) 2. Cuff (Type BF applied part) 22 cm - 42 cm 3. User manual 4. USB cable (5 V = 1 A) 5. Storage bag

The cuff is designed to fit arm sizes between

Explanation of the Display

Average value The average value of the latest 3
Low battery Battery is low and need to be charged
User 1 Start measurement and save the measuring results for User 1
User 2 Start measurement and save the measuring results for User 2

Power Supply and Charge Power


Input: AC 100 - 240 V ~50/60 Hz 0,4 A Max Output: 5 V 1000 mA
Power needs to be charges under the following circumstances:
Method 1 · Method 2

If you switch on the blood pressure monitor and the LCD display flashes “ “. This means that the battery charge is low. Please recharge it. You can, however, still carry out the measurement.
If the LCD display “ “ appears, this means that the battery charge is too low. The blood pressure monitor will switch off automatically and you must recharge it imme- diately.
During recharging, the LCD display flashes and shows the following symbols:

If it is not charged at all, the following symbols appear on the LCD display during charging:
If charge level 1 has been reached, the following symbols appear on the LCD display during charging:
If charge level 2 has been reached, the following symbol ap- pears on the LCD display during charging:
If charge level 3 has been reached, the following symbol ap- pears on the LCD display during charging:

rechargeable lithium-ion battery. Please do not try to disassemble or force open the blood pressure monitor yourself or permit unauthorised maintenance person- nel to do so.
times. If the battery fails to recharge or the unit cannot be used normally, please contact your dealer. If you measure your blood pressure three times a day, the unit can be used for up to 25 days without recharging.
recharging process. When the unit is fully charged, please pull out the plug in good time.
pressure monitor. The battery is built-in and is not re- placeable.
tremely high or low ambient temperatures.
charging. Always remove the charging unit before cleaning.

Activate your blood pressure monitor
Your blood pressure monitor is activated when the date and time are set.
Setting Date and Time
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (time format: 24 H)

mode for year setting.

The user ID will blink.
Select User

Attaching the Pressure Cuff
2,5 cm · Posture during measurement

LCD display 0 – value is displayed


Display and save the measurement


Both User 1 and User 2 have 60 memory slots.

No. 1
The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record is dropped from the list.
press the “MEM” button and re- lease it.



Deleting readings


Do not modify or dis- assemble the unit or the arm cuff.
Use a cloth moistened with water or neutral detergent to clean the body of the unit and then wipe it dry.


Keep the unit in a suitable place. Avoid high tempe- rature, direct sunlight, high moisture, and dust.
Do not press the “START” button if the cuff has not been properly wrapped around the arm.


Press and hold down “MEM“ button to activate the unit
The cuff is not properly fastened or pressure in the cuff is too high
Rest for a short time, re-fasten the cuff and measure the blood pressure again.
The unit has de- tected movement or pulse rate is too low.
Rest for a short time and retake the meas- urement.
Rest for a short time and retake the meas- urement.
Start the measuring process again. If this problem re-occurs, contact your dealer.
Relax for a moment. Re-fasten the cuff and measure again. If the problem re-occurs, consult your physician.

Model-No. GT-1115
Model-No. 3,7 V 1000 mAH built-in rechargeable lithium-ion battery
Display mode Digital LCD Display V.A. 44 mm x 64 mm
Measurement mode Oscillographic testing mode
Measurement pressure: SYS: 60 mmHg ~ 230 mmHg (8,0 kPa ~ 30,7 kPa) DIA: 40 mmHg ~ 130 mmHg (5,3 kPa ~ 17,3 kPa)
+5 °C to +40 °C Relative humidity: ≤85 % RH Air pressure: 80 kPa to 106 kPa
Storage and transportation condition
-20 °C bis +60 °C Relative humidity: 10 % RH to 93 % RH
Cuff size 22 cm ~ 42 cm
Net Weight approx. 550 g excluding cuff
External dimen- sions approx. 131 mm x 78 mm x 42 mm
Attachment USB cable, user manual, storage bag
Measurement range · Normal working condition

Degree of protection IP20
Subject to changes in the interests
Device classifi- cation · of technical progress

Geratherm® is certified in accordance with Council Direc- tive 93/42/EEC and EN ISO 13485 and is entitled to affix the CE-mark (Notified Body: TÜV Rheinland LGA Products GmbH).
The blood pressure monitor conforms to
ISO/EN 14971 Medical devices — Application of risk management to medical devices
Labelling EN 980 symbols for the labelling of medi- cal devices
EN 1041 Medical equipment manufac-turers to provide information
EN 60601-1 +A1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
General Require- ments for Safety
IEC/EN 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC/EN 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests.
EN ISO 81060-1 Non-invasive sphygmomanometers – Part 1: Require- ments and test methods for non-automated measurement type (ISO 81060-1) EN 1060-3 +A2 Non-invasive blood pressure monitor - Part 3: Supplementary requirements for elec- tromechanical blood pressure measuring system


DIN EN ISO 81060-2 Non-invasive sphygmomanometers – Part 2: Clinical investigation of automated measurement type (ISO 81060-2)
IEC/EN 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability IEC/EN 62366 Medical devices - Application of usability engineering to medical devices
IEC/EN 62304 +AC Medical device software - Software life cycle processes

Store at relative humid- ity levels between 10 % and 93 % R.H.
The device must not be disposed of with household waste

Information on electromagnetic compatibility (EMC)
Electronic devices such as PCs and mobile phones can lead to the exposure of medical devices in operation to electromagnetic interference from other devices. This can lead to malfunction of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with any other devices.
The EN 60601-1-2 standard regulates the requirements for EMC (electromagnetic compatibility) and defines the levels of immunity to electromagnetic interference and the maximum electromagnetic emission levels for medi- cal devices.
This blood pressure monitor, which is manufactured by Geratherm Medical AG, complies with the EN 60601-1-2 standard in relation to both immunity and emissions.
However, special precautions should be observed:
Please only use the device indoors and not in the vici- nity of mobile phones or microwave ovens. In the case of devices whose maximum power exceeds 2 W, the minimum distance from your blood pressure monitor should be 3.3 metres.

Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only beused in such environments:
Emission test Compatibility Notes concerning the EM environment
RF energy is used only to maintain device’s operation. Therefore, its RF emissions are so low that it’s not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B The device is suitable for use in all establish- ments, including domesticestablishments, and those directly connected to the public low-voltage power supply networkthat supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3 complies
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only beused in such environments:
Immunity test IEC 60601 – test level
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrostatic discharge (ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
The power frequency magnetic field should be measured at the intended place of installation in order to ensure that it is low enough.
3 A/m 3 A/m
Recommended separation distances between portable and mobile RF communication
The device is intended for use in an electromagnetic environment where radiated RF disturbancesare under control. User can help prevent electromagnetic interference by keeping the device at a minimum distance from portable and mobile RF communications equipment (transmitters). Below table details the maximum output power of transmitter:
Rated maximum output power of transmitters in Watt
Seperation distance according to frequency of transmitter / m
150 kHz to 80 MHz d = 1.2 √P
80 MHz to 800 MHz d = 1.2 √P
800 MHz to 2.5 GHz d = 2.3 √P
0.01 0.1 1 10 100
0.12 0.38 1.2 3.8 12
0.12 0.38 1.2 3.8 12
0.23 0.73 2.3 7.3 23
For transmitters rated at a maximum output power not listed above, the recommended separa- tion distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Compliance level

Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments:
Immunity test IEC 60601 – test level
Compliance level Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz 3 Vrms d= 1.2 √P
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz 3 V/m d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recom- mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity ofequipment marked with the following symbol:
Guidance and manufacturer´s declaration – electromagnetic immunity
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
This equipment needs to be installed and put into service in accordance with the informa- tion provided in the user manual; Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equip- ment and should be kept at least a distance d = 3,3 m away from the equipment. (Note: As indicated in Table 6 of IEC 60601-1-2 for ME EQUIPMENT, a typical cell phone with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 3 V/m)
The current version of the standards is valid.
1) · 2)

Geratherm Medical AG Fahrenheitstraße 1 98716 Geschwenda Germany www.geratherm.com




NOTICE D’UTILISATION
GT-1115


Pagina's waarnaar wordt verwezen in deze handleiding — fabrikant, ondersteuning en downloads.
Wat de op het display getoonde foutcodes betekenen en hoe u ze wist.
| Code | Betekenis | Oplossing | |
|---|---|---|---|
| E 3 | The cuff is not properly fastened or pressure in the cuff is too high | Rest for a short time, re-fasten the cuff and measure the blood pressure again. | p.21 |
| E10 or E11 | The unit has de- tected movement or pulse rate is too low. | Rest for a short time and retake the meas- urement. | p.21 |
| E20 | No pulse signal is detected | Loosen clothing around the arm and measure again. | p.21 |
| E21 | The measuring process was not successful. | Rest for a short time and retake the meas- urement. | p.21 |